DACIS
-
Upload
hainguyenbkvhv -
Category
Documents
-
view
221 -
download
0
Transcript of DACIS
-
7/25/2019 DACIS
1/85
Close Interval Potential Surveys (CIPS or CIS) using
MCMs Integrated Pipeline Survey Test Equipmentand Database Management Package
CIS Training ManualDA Meter Version
M.C. Miller Co., Inc.
11640 U.S. Highway 1, Sebastian, FL 32958 U.S.A.
Telephone: 772 794 9448; Website: www.mcmiller.com
-
7/25/2019 DACIS
2/85
MANUAL CONTENTS
Page
SECTION I: INTRODUCTION AND BACKGROUND
I. 1
I. 2I. 3
I. 4
Introduction .
Background on ECDA Methodology and Role of CIS ...CIS Overview ..
Key Components of a CIS Measurement System
4
4
5
6
SECTION II: MCMs REFERENCE ELECTRODE FOR CIS
II. 1 Background ... 8
II. 2 How to Prepare Reference Electrode for Use 10
II. 3 How to Use Reference Electrode in CIS Measurements ............... 11
II. 4 How to Use Matched-Pairs of Reference Electrodes (Canes) .. 12
SECTION III: MCMs DA METER (DATA-LOGGER)
III. 1 Introduction ... 14
III. 2 How to Set Up the DA Meter for CIS Applications .. 15
III. 3 How to Make Changes to Your Set Up Selections 46
SECTION IV: TEST EQUIPMENT HOOK-UPS FOR CIS
IV. 1 How to Make Cable Hook-Ups for CIS Measurements . 47
IV. 2 How to Attach Cables and Accessories to the DA Meter 48
Continued Over/
2
-
7/25/2019 DACIS
3/85
SECTION V: HOW TO PERFORM CLOSE INTERVAL SURVEYS
V.1 How to Carry Test Equipment During a CIS .. 50V.2 How to Examine and Record a High/Low Voltage Waveform 51V.3 How to Record the Voltages at a Starting Test Station 53V.4 How to Perform a CIS .. 54
V.5 How to Use the GPS Buttons, if Active, on the Survey Screen . 65
SECTION VI: HOW TO COPY SURVEY FILES FROM THE
DA METER TO YOUR PC
VI. 1 Introduction 67
VI. 2 The Manual Approach ... 67
VI. 3 Using the Driver in the ProActive Program ... 68
APPENDIX 1:
APPENDIX 2:
How to Connect the Chainer and Counter Unit
for CIS Applications .. 77
How to Delete Survey Files from the DA Meter ...... 85
3
-
7/25/2019 DACIS
4/85
SECTION I: INTRODUCTION AND BACKGROUND
I. 1 Introduction
This manual is designed to provide individuals who will be performing
Close Interval Potential Surveys (CIPS or CIS) with a practical knowledge
of how to select appropriate test equipment, how to set up and properlyconnect the test apparatus, how to actually perform close interval potential
surveys, and, ultimately, how to transfer (upload) stored survey data from
data-loggers to a database management system, after performing such surveys. The equipment and software described here comprise MCMs
Integrated Pipeline Survey Test Equipment and Database Management
Package.
I. 2 Background on the ECDA methodology and the role of CIS
By way of background, the National Association of Corrosion Engineers
(NACE), under a directive from the U.S. Government, recently developed a
methodology for assessing and reducing the impact of external corrosion on
the integrity of onshore buried pipelines (primarily ferrous pipelines). The
methodology is termed an External Corrosion Direct Assessment (ECDA).ECDA is a continuous improvement process designed to not only identify
areas where external corrosion is underway, but to also predict potentialfuture corrosion areas, which will assist greatly in future corrosion
prevention.
The ECDA methodology is a 4 step process and each step must be
performed to fully-satisfy the integrated requirements of an ECDAevaluation. The 4 major steps in the process are summarized below:
Step 1: Pre-Assessment.This step requires the integration of historical, construction, operations andmaintenance records for each ECDA Region.
Step 2: Indirect Inspection.
The objective of this step is to estimate areas of active corrosion by
performing above-the-ground measurements.
4
-
7/25/2019 DACIS
5/85
The results obtained in Step 2 are compared with the information compiled
in Step 1 and areas that need to receive Direct Examination are identified.
Step 3: Direct Examination.
The objective of this step is to excavate (dig) at least in those areas identifiedusing the complementary steps above (steps 1 and 2) and, by directinspection, to measure the extent of corrosion in those areas. Typically, wallthickness is measured and other corrosion-related data are collected in this
step.
Step 4: Post-Assessment
The objective of this step is to prioritize the repair schedule and to decide on
the mitigation approaches to be used based on the results obtained in Step 3.The period of time before the next integrity evaluation is to be undertaken is
also decided in this step for each ECDA Region.
The subject of this Training Manual is a measurement process that is applied
in Step 2 of the ECDA evaluation process, ie, in the Indirect Inspection
Step. The measurement process is known as a close interval potential
survey (CIPS or CIS), the details of which are discussed below.
I. 3 CIS Overview
With regard to Indirect Inspection of a buried pipeline, one major
technique that is employed is a close interval potential survey (CIS), which
involves performing above-the-ground measurements of pipe-to-soil
potential (voltage) along the entire length of the pipeline. In a CIS, it iscritical that measurements be made in a highly consistent fashion (from
individual technician to individual technician) and data (pipe-to-soilpotentials) be recorded and analyzed in a consistent fashion in order that
such data can be considered an integral part of an ECDA evaluation.
Close interval potential surveys are performed in order to obtain a
continuous potential profile along the pipeline.
5
-
7/25/2019 DACIS
6/85
In the Cathodic Protection industry, it is well known that pipe-to-soil
potential (voltage) measurements at test stations, which are typically spaceda considerable distance apart, are insufficient to judge the overall condition
of a pipeline and to judge whether or not there is complete protection. As a
result, close interval potential surveys involving the measurement ofpotentials at short intervals along the entire length of a pipeline have becomethe industry standard. In fact, with regard to the ECDA protocol, pipe-to-soil potential readings are typically recorded at 2.5 feet intervals between
test stations. (Test stations are insulated electrodes that are in permanent
electrical contact with the pipeline and that can be contacted above ground.)
I. 4 Key Components of a CIS Measurement System
The diagram below (Figure 1) illustrates the essential components of a close
interval potential survey measurement apparatus.
Figure 1: Key Components of a CIS Test Equipment System
6
-
7/25/2019 DACIS
7/85
The key components of the CIS apparatus are the reference electrode,
connecting the negative terminal of the data-logger to the soil, and the datalogger, the positive terminal of which is connected to the test station (and
therefore to the pipe). The data-logger is a sophisticated digital
voltmeter/data storage unit. Consequently, with this apparatus, the potentialdifference (voltage difference) between the pipe and the soil (at the referenceelectrode location) can be measured, and this data point (voltage at a specificlocation along the length of the pipe with respect to the reference electrode)
can be stored for processing by means of the digital voltmeter/data storage
unit (data logger). Similarly, data (of pipe-to-soil potential with respect to
the reference electrode) can be recorded at intervals of, typically, 2.5 to 5.0 feet along the length of the pipeline.
The ultimate goal of a CIS is to identify locations (if any) along the length of
a buried pipeline that are not registering a sufficient potential differencebetween pipe and soil, which would be indicative of locations that might be
experiencing external corrosion. As discussed below in Section II, there isan industry standard (0.85 Volts or 850mV) which is applied in the Cathodic
Protection industry and which represents the minimum potential difference(voltage) recorded between the pipe and the soil (with respect to a particular
reference electrode (see below in Section II)), that signifies sufficientcathodic protection. Since, on a pipeline that is under cathodic protection(impressed current CP), the pipe is held at negative potential (negative side
of the rectifier is connected to the pipe), the minimum potential differencebetween pipe and soil would be -0.85 Volts. Any more-positive (less-
negative) voltages, for example, -0.7 Volts, would suggest insufficientcathodic protection and would indicate a location where external corrosion
might be taking place. Actually, a potential difference of greater than -0.85Volts (for example, -1.0 Volt or higher) might be required to be in the safe
area with regard to a pipeline being fully cathodically protected undercurrent flow conditions, particularly if the amount of (ionic) current flowing
in the soil from the anode to the pipe is large. This situation would result in
a significant, so-called, IR drop (voltage drop) due to the current flow in the
soil which must be added to the minimum 0.85 Volt potential difference toensure sufficient cathodic protection. As is discussed in Section III, it is
possible to determine the magnitude of this IR drop voltage during the
performance of a CIS by conducting the CIS measurements in the High-Low, or current-interrupted, mode, where the pipe-to-soil potential is
sampled as the current is switched ON and OFF in a cyclic fashion (see
7
-
7/25/2019 DACIS
8/85
Section III). The critical pipe-to-soil potential (with regard to ensuring
sufficient cathodic protection) would be the potential measured during thecurrent OFF part of the cycle, since in this case the IR drop would be
eliminated.
In any case, once a critical pipe-to-soil potential difference has been
established for sufficient cathodic protection of a particular pipeline (takingthe IR drop into account), a CIS can be performed to monitor the condition
of the pipe, by comparing the pipeline potential profile recorded with theideal case scenario, which would be a uniform (constant) potential along the
length of the pipe.
The following two sections, Sections II and III, describe the setup
procedures and operation of the reference electrodes and data-loggers
employed in MCMs Integrated CIS Test Equipment Package.
SECTION II: MCMs REFERENCE ELECTRODE FOR CIS
II. 1 Background
As illustrated in Figure 1 in the above section (Section I), the reference
electrode makes physical contact with the ground (soil) and is connectedelectrically to the negative terminal of the data-logger (digital voltmeter/data
storage unit). Since the reference electrode is an integral part of theelectrical circuit that is allowing the data-logger to measure the potential
difference between the pipe and the soil, the reference electrode must make good and consistent electrical contact with the soil at each measurement
point during a CIS.
The industry standard reference electrode for pipe-to-soil measurements is
the copper-copper sulphate half-cell or, as its more commonly referred to
as, the copper sulphate reference electrode. A schematic illustration of anMCM copper sulphate reference electrode is shown in Figure 2. Also shownin the figure is the reference electrode attached to a push-button cane
assembly for use in CIS measurements (see Section III. 3, step 6).
8
-
7/25/2019 DACIS
9/85
Figure 2: MCMs Reference Electrode
As can be seen in Figure 2, the reference electrode consists of a pure copper
rod that is immersed in a saturated copper sulphate solution within the body
of the electrode. The copper rod connects electrically to the negativeterminal of the data-logger. During operation, the porous plug is saturated
with the copper sulphate solution and it is the solution-saturated plug thatmakes actual electrical contact with the soil (not the copper rod itself).
It should be noted that the copper/saturated copper sulphate solution
combination within the body of the electrode, is actually a half-cell with abuilt-in half-cell potential (voltage).
9
-
7/25/2019 DACIS
10/85
Consequently, when the data-logger is connected between the reference
electrode and the pipe, as shown in Figure 1 (Section I), the observedvoltage reading is actually a combination of two voltages, the copper-copper
sulphate half-cell voltage (a constant voltage) and the pipe-to-soil voltage
(variable, depending on corrosion conditions).
Therefore, if it can be assumed that the reference electrode half-cell voltage
is fixed (a constant value), any variation in the voltage measurements
recorded during a CIS can be assumed to be due to variations in pipe-to-soilpotential and, therefore, due to the external corrosive condition of the pipe in
the various measurement areas. The industry standard voltage value of-0.85 Volts discussed in Section I. 3, which represents a minimum pipe-to-
soil potential difference for adequate protection against corrosion, is actuallya combination (sum) of the reference electrode half-cell voltage and the
pipe-to-soil voltage.
Consequently, since so much emphasis is placed on this critical
voltage value, it is imperative that the reference electrode be
set up, operated and maintained appropriately so that its half-
cell voltage value will, in fact, be fixed.
II. 2 How to Prepare Reference Electrode for Use
The reference electrode will arrive from M.C. Miller Co., Inc. (MCM)
assembled, but without liquid in the plastic tube. There will, however, be a
concentration of blue copper sulphate crystals inside of the transparentplastic tube. To prepare the reference electrode for use, the transparent
plastic tube should be unscrewed from the threaded cap/copper rod assembly
(by turning the plastic tube counter-clockwise with respect to the threadedcap/copper rod assembly) and the tube (with the copper sulphate crystals at
the bottom) should be filled up to about inch from the bottom of the
threads with distilled water (or MCMs Electrode Anti-Freeze solution).The plastic tube should then be screwed back on to the threaded cap/copperrod assembly and tightened firmly (do not over-torque) to effectively seal
the tube via the compressed O-ring. At this point, the electrode assembly
should be shaken a few times to make sure that a saturated solution of
copper sulphate is formed inside the tube. Remember that the half-cell is
10
-
7/25/2019 DACIS
11/85
established, as discussed in (Section II. 1), between the copper rod and a
saturated copper sulphate solution. Once the tube is shaken, the liquid willbe blue in color and there should always be some excess copper sulphate
crystals remaining out of solution at the bottom of the tube. Before using the
reference electrode for the first time, a period of at least 5 minutes should beallowed, after filling the tube and establishing the saturated solution, in orderthat the porous plug on the bottom of the electrode tube can become moist.The protective cap supplied with the electrode assembly should be kept on
the porous plug end when the electrode is not being used.
For more details on the preparation and maintenance of copper sulphate
reference electrodes, including important safety precautions, please visit
the M. C. Miller Co., Inc. website at www.mcmiller.com. On the home page, click on manuals, data sheets and then click on Electrodes How
to use and maintain copper sulphate electrodes.
II. 3 How to Use Reference Electrode in CIS Measurements
In order to proceed with CIS measurements, remove the protective cap from
the porous end of a properly setup (prepared) MCM reference electrode [see
Section II. 2 for preparation instructions].
Place the reference electrode (porous plug side down) on the ground over thepipe at the first measurement location (typically at a test station electrode).
The porous plug of the reference electrode (suitably moist) should be in firmcontact with moist soil.
This may require digging in where the earths surface is dry. If the soil is
particularly dry, it may be necessary to moisten the soil around the electrodeusing fresh water to obtain good electrical contact.
The top terminal of the reference electrode is connected to the negative side
of the data-loggers voltmeter and the positive side of the voltmeter isconnected to the test station electrode (Note: The specifics on cable hook-
ups are provided in Section IV).
11
-
7/25/2019 DACIS
12/85
For CIS work, the reference electrode plastic tube assembly is attached to an
extension rod (commonly know as a cane) for upright handheld operation.The cane handle houses a push-button switch (for triggering voltage
recordings and other events) and the terminal on the cane for electrical
connection to the data-logger is a 5-pin connector. As discussed in SectionIV, the receiving terminal for the reference electrode on the data-logger isalso a 5-pin connector. All electrical cables are provided for theseconnections in the MCM Integrated CIS Test Equipment Package. A
schematic illustration of a reference electrode cane assembly is shown above
in Figure 2.
If a matched-pair of reference electrodes (canes) is to be used (see Section
II. 4), the top terminals of the two canes are connected to an MCM adapter(dual-probe adapter) and the adapter is connected (effectively) to the
negative terminal of the voltmeter (see Section IV for cable hook-ups).
At this point, a voltage reading can be made using the data-logger,
establishing the pipe-to-soil potential difference (with respect to the copper
sulphate electrode) at the first measurement location. The operatingprocedures for the data-logger are described in Section III. All
subsequent CIS measurements can be made in the same fashion by movingdown the length of the pipeline between test stations and placing thereference electrode (or electrodes, if using a matched-pair) in good electrical
contact with the soil at intervals of 2.5 feet, or some other distance interval.Since the distance between test stations can be significant, the electrical wire
(trail wire) connecting the data-logger to a test station needs to have
considerable length.
II. 4 How to Use Matched-Pairs of Reference Electrodes (Canes)
In CIS measurements, it is convenient to use a pair of matched reference
electrodes, rather than a single reference electrode. A pair of electrodes is
referred to as either a set of poles or a set of canes. As is discussed in
Section III, MCMs data-loggers can be set up to accept readings (voltage
readings) during the performance of a CIS from two reference electrodes(canes). In this case, the canes would be connected to the data-logger via thedual-probe adapter mentioned above.
12
-
7/25/2019 DACIS
13/85
Since each cane in the pair has a push-button switch located on the handle
section on top of the cane, the individual performing the CIS can determinewhich of the two canes (reference electrodes) is to be involved in a voltage
recording at any given time. As discussed in Section III. 3, these push
buttons can also be designated to have other functions, such as registering asurvey flag location, for example.
When performing CIS measurements using a pair of canes, one of the canes
is held in the operators right hand (green-colored cane) while the secondcane is held in the operators left hand (red-colored cane). When, for
example, the right-hand cane is placed in electrical contact with the soilabove the pipe and the push-button switch is activated on the cane, the DA
Meter (see Section III. 3) will record the pipe-to-soil potential with respectto the right-hand cane (reference electrode). Similarly, when the left-hand
cane is placed in electrical contact with the soil above the pipe and the push-button switch is activated on the cane, the DA Meter will record the pipe-to-
soil potential with respect to the left-hand cane (reference electrode),
assuming that both canes have been set up to read voltages (see Section
III. 3, Step 7).
There are several advantages to using a pair of canes (reference electrodes),
rather than a single cane (a single reference electrode) for CIS. For example,both canes can be used to record pipe-to-soil voltages in an alternating
fashion (one after the other) as the operator walks down the length of thepipeline. In this fashion, the operator can establish a rhythm, very much like
a cross-country skier, contacting the soil above the pipe every 2.5 feet bytriggering the left-hand cane, then the right-hand cane, then the left-hand
cane etc., etc.
Also, one of the canes, for example, the left-hand cane could be used to
register the location of survey flags, which are typically located every
100 feet down the length of a pipeline. With this arrangement, the right-hand cane would be used to record pipe-to soil voltages every 2.5 feet while
the left-hand cane would register survey flags. This methodology has theadvantage of periodic checks on the operators location, with the DAMeters software being able to re-establish starting points every 100 feet
(survey flag separation).
13
-
7/25/2019 DACIS
14/85
A critical requirement with regard to using reference electrode pairs is that
the two electrodes (canes) be essentially identical. As discussed in SectionII. 1, a copper/copper sulphate reference electrode is a half-cell with a built-
in voltage and the half-cell is an integral part of the electrical circuit of
Figure 1 (Section I). Consequently, if two reference electrodes (canes) areused to measure pipe-to-soil potential in an alternating fashion, their half-cell voltages need to be essentially identical if any measured pipe-to-soil
potential differences are to be attributed to the corrosive condition of the
pipe, rather than to any variation in the half-cell voltages of the two
reference electrodes.
A pair of canes can be tested for compatibility in the following manner: The
two electrodes can be placed end-to-end (with the moist porous plug ends infirm contact) and the voltage difference between them measured using a
high-input-impedance voltmeter connected between their top terminals. If avoltage difference (difference in half-cell voltages) of less than 10mV is
observed, the two reference electrodes (canes) are matched. The tworeference electrodes could also be placed in a plastic bucket containing fresh
water and the voltage difference measured again between their top terminals.If this voltage is less than 10mV, the two reference electrodes are matched
for use as a cane pair in CIS measurements.
SECTION III: MCMs DA METER (DATA-LOGGER)
III. 1 Introduction
MCM data-loggers are sophisticated high-input-impedance digital
voltmeter/data storage units that are used in conjunction with the reference
electrodes described above in Section II to perform CIS measurements.
When properly connected with reference electrodes (as described in Section
IV), a data-logger reads and records pipe-to-soil voltages (with respect to a
reference electrode) at each of the measurement locations along the length of
the pipeline (typically every 2.5 feet for CIS work).
14
-
7/25/2019 DACIS
15/85
The particular data-logger described here is the, so-called, DA Meter. This
data-logger is a combination of a high-input-impedance digital voltmeter anda handheld Personal Computer (PC). The operating system of the PC is
Windows for Handheld PC 2000 and MCM has designed proprietary
application software programs in order that the DA Meter can be employedto run both Scripted Site Surveys and Pipeline Surveys (such as CIS,DCVG and Surface Potential Surveys).
Your DA Meter data-logger will arrive from MCM with the Windows
operating system and our survey application programs installed. The survey
application programs will remain installed unless the main battery pack, andits backup, are allowed to fully-discharge, in which case the application
programs will have to be re-installed (see your DA Meter Users Manual).However, your survey data will always be safe, regardless of battery status,
as these data are stored in a non-volatile memory (CompactFlash Card).
The particular application software program required to run Close Interval
Survey (CIS) applications is called the DA Meter PLS (Pipe Line Survey)
program and the use of this program is described below in detail in Section
III. 2.
III. 2 How to Set Up the DA Meter for CIS Applications
The following section outlines the steps required to setup the DA Meter toparticipate in CIS measurement applications. The setup process establishes
the conditions of the particular survey about to be performed and identifiesthe section of pipeline that is about to be examined by the CIS application.
The setup process also establishes a file in which the voltage recordings(survey data) will be stored. At the completion of the survey, CIS data can
then be retrieved by a PC that is in communication with the DA Meter, byaccessing the file in which the survey data are stored. The process of
transferring CIS data to a PC following completion of a survey is discussedin Section VI.
Step 1:Switch on the DA Meter by pressing the power button. Assuming that the
battery pack is charged, the screen will light up and will display the
Windows Powered (Handheld PC 2000) desktop screen as shown below(assuming that the operating system running your DA Meter is WindowsHandheld PC 2000).
15
-
7/25/2019 DACIS
16/85
Step 2:
Double tap (using the special pen (stylus) provided with the DA Meter) onthe DA Meter PLS icon. This will open the software program that will
run the DA Meter for Pipeline Surveys. The window shown below willappear.
Note: It may take a few seconds for this window to appear.
This is the Main Survey Window. At this point, a survey has not been set up
(the setup process is what we are going through here) and so no information is currently displayed in this window.
Step 3:Tap once on the Survey button situated in the Menu Bar in the top left-hand corner of the window. The window shown below will appear.
16
-
7/25/2019 DACIS
17/85
Under Survey there are several options. If this is a new survey (not a
continuation of a previous survey) tap once on New Survey. The windowshown below will appear.
Step 4:Enter a filename for the Survey using the keyboard.
Note: This is an important step as the filename is used to identify the survey
and, also, recorded data (voltages) will be stored in this named file for futureretrieval. It is highly recommended that a protocol be established for
selecting Survey Filenames. Critical information should be included in thefilename, such as pipeline companysname, city or state in which the
pipeline is located, pipeline number and section of pipeline number undersurvey. The protocol developed should be applied consistently for each
survey.
17
-
7/25/2019 DACIS
18/85
For example, lets assume that pipeline company XYZ has a pipeline located
in Texas and that the pipeline is identified as pipeline 12 and a survey isbeing performed on section 085 of this pipeline. A good filename for this
survey would be:
XYZ TX 12 085 CIS
When this data file is later accessed, with this filename we know the name of
the pipeline owner, we know the state in which the pipeline is located, weknow the pipeline number, we know the section number of the pipeline that
was surveyed and we know that it was a CIS.
Note: You will not be permitted to use invalid characters, such as
slashes( / or \ ), as part of a filename. You will be alerted if you try to
use any invalid characters.
Step 5:Tap once on the OK button. The window shown below will appear. This is the first of 5, so-called, setup windows.
Step 6:
Select Survey Type.For CIS measurements there are 2 options for type of survey: Trigger CIS
and Continuous CIS. Other survey types are available as can be seen bytapping once on the pull-down list arrow button in the Survey Type field,as shown below.
18
-
7/25/2019 DACIS
19/85
As can be seen from the above screen, DCVG, combined DCVG/CIS and
Surface Potential Surveys are available for selection, however, for CIS-Only
surveys, either Continuous CIS or Trigger CIS should be selected by tappingonce on the selection.
In Trigger CIS mode, voltage recordings are triggered by the operator
pressing on the push-button switch located on top of a reference electrode
cane (data probe) while making electrical contact with the soil above the
pipe with the copper/copper sulphate half-cell end of the cane.
Consequently, in this survey mode, the operator is in control of the timing
of when voltage recordings are made as a CIS is conducted.
Note: The Trigger CIS mode should also be selected when using thetrigger cable (part # SIN024) for automatic triggering of readings inconjunction with the wire chainer/counter unit.
In contrast, with the Continuous CIS mode, the timing of when voltage
recordings will be made is preset. For instance, if you select 1 second
intervals, voltage recordings will be made every second. Consequently, the
operator must synchronize his cane repositioning pace to match the preset voltage recording timing.
Step 7:Select the function (operating mode) of the cane buttons associated with the
two reference electrodes (canes) to be used in the CIS. Selections are made
by tapping choices in the D.C. P. and Survey fields shown in the Step 5
screen above.
19
-
7/25/2019 DACIS
20/85
The function of the cane buttons during regular survey measurements is
determined by the selection made in the survey field while the selectionmade in the D.C.P. field determines the function of the cane buttons when
measurements are made at specially-identified locations (devices).
Cane Button FunctionsDuring Regular Survey (Between Devices)
The functionality available for the cane buttons depends on the survey mode
selected in Step 6 [Trigger mode or Continuous (automatic) mode].
If the Trigger CIS mode is selected, the screen will appear as shown below.
For the above screen, the pull-down-list arrow button in the Survey field was
tapped to reveal the 4 choices for the push buttons on the pair of canes to beused in the Trigger CIS (the left cane, held in the left hand with the left
thumb operating the left cane button and the right cane, held in the righthand with the right thumb operating the right cane button) .
Manual Triggering of Readings:
The left and right cane buttons can both be selected to read voltages
(read read) meaning that as the operator walks down the pipeline
alternately contacting the soil with the left and right canes, a voltagerecording will be made when either cane is triggered. Alternatively, one of
20
-
7/25/2019 DACIS
21/85
the canes can be selected to read voltages while the other is selected to
register the location of a survey flag (survey flags are typically located every100 feet down the length of a pipeline). This would be the (flag read) or
(read flag) choice. In this case, the push button on the cane selected to
register flags would only be pressed at each survey flag location. Finally,the (flag flag) choice can be selected and, in this case, both canes wouldonly be used to register survey flags. Pressing the push-button on the readdesignated cane or pressing the cane button on the flag designated cane
has the same affect as tapping the read button (and saving the reading) or the
flag button, respectively.
Automatic Triggering of Readings:
When the trigger cable (part # SIN024) is used in conjunction with the wirechainer/counter unit, the trigger cable signal simulates the action of the left-
hand (red-handled) data probe push button, while the right-hand (green-handled) data probe push button is disabled (has no effect). Consequently,
in this case, it is important that the left-hand data-probe push button bedesignated as a read button, while the right-hand push button can be
designated as either read or flag (again, the right-hand data probe pushbutton is not a factor). The appropriate selections would be read read, or,
read flag. In this case, survey flags can be designated by tapping the Flagbutton on the survey screen.
If Continuous CIS mode is selected, the screen will be as shown below.
21
-
7/25/2019 DACIS
22/85
The 4 choices for the set of two canes in the Continuous CIS mode are
revealed in the Survey field by tapping once on the pull-down-list arrow.The 4 choices are shown in the screen above. In this case, one of the cane
buttons can be selected to effect a pause in the continuous (automatic every
so many seconds as prescribed) voltage recordings while the other canebutton is selected to register survey flag locations [(flag pause) or (pauseflag)]. Alternatively, both cane buttons can be selected to register flags or to affect pauses. The pause cane would also act as the start or restart cane.
Cane Button FunctionsAt Device Locations
Devices are defined as specially-identifiable entities that exist along thelength of the pipeline. Examples of Devices include, rectifiers, test
stations, valves, etc.
For both CIS modes (Trigger and Continuous), a cane button function
selection (selection will apply to both left and right canes) has to be made
with regard to Data Collection Points (see D.C.P. field in above screen). AData Collection Point is a Device location point.
In this case there are 3 choices: None, Accept and Accept and Save.
Your selection is made by tapping once in the circle beside your choice.
If none is selected: Pressing a cane button will do nothing. In this case,
you would have to tap the accept and then save buttons on the screen,having previously tapped the device button, if you wished to read and
record the voltage data at the Device location.
If accept is selected: Pressing a cane button has the same effect as
tapping the accept button on the Device Readings screen.
If accept and save is selected: Pressing a cane button has the same effect
as tapping the accept button and then tapping the save button on the DeviceReadings screen.
For the most part, selecting the accept and save choice will be preferable,
since voltage recordings at special data collection points (such as rectifiers,
test stations, valves, etc.) are important and should be saved as part of theoverall CIS and the most convenient way to do this is to use the push-button
on one of the canes, having tapped on the Device button on the screen
when a Device is encountered.
22
-
7/25/2019 DACIS
23/85
Step 8a:Make Survey Walking Direction selection
The Survey Walking Direction field is shown in the screen above as part
of the Setup 1 of 5 screen. There are only 2 selection choices: Increasing
and Decreasing.
This selection establishes the walking direction during the survey. If the
survey is going to entail walking from a low-numbered test station to ahigher-numbered test station, for example, select Increasing by tapping
once on the circle beside Increasing. Select Decreasing if the opposite
is true.
Step 8b:Make selection of Metric units if required.
By checking off the box labeled Metric, the reading interval (distance
between voltage recordings) and the flag internal (flag spacing) will be
displayed on the Setup 3 of 5 screen in meters, as opposed to feet.
Step 8c:Check-off DCVG Sidedrains in CIS Survey if you plan to log DCVG
sidedrain readings at previously located anomalies during the course of aCIS.
By doing so, a special Device selection , named Mark DCVG will be
made available in the Device list, which has a sidedrain prompt associatedwith it, as indicated in the screens shown below.
23
-
7/25/2019 DACIS
24/85
The Mark DCVG device will not appear in the Device list if this box is not
checked, which would be the case if you were conducting a regular CIS.
If the following selections have been made so far, the Setup 1 of 5 screen
will appear as shown below:
Survey Type: Trigger CIS
Survey: Flag Read
D.C.P.: Accept Save
Survey Walking Direction: Increasing; Metric: NOT selected
Step 9:Make Rectifier Mode selection.
24
-
7/25/2019 DACIS
25/85
In the case where a pipeline is cathodically-protected via an impressed
current (current flow to the pipe from a sacrificial anode driven by an external DC voltage source), the rectifier (DC voltage source) is typically
always ON, so that the line is continuously protected. However, with
current flowing in the soil to the pipe, the pipe-to-soil voltages beingmeasured during a CIS will include any voltage drop (IR drop) occurring inthe soil itself.
If a large amount of current is flowing in the soil and/or the soil has a high-
resistivity, the magnitude of the voltage drop (IR drop) could be significant.
Consequently, it is important to measure the actual pipe-to-soil potential(voltage) with no current flowing in the soil, with a view to establishing that
the pipe is at a sufficiently negative potential (at least -0.85 Volts withrespect to copper sulphate reference electrodes). The problem is that the
rectifiers (DC voltage sources) cannot be switched OFF for the duration ofCIS measurements, since the pipeline would lose its cathodic protection
while the CIS measurements were being performed.
The solution to this problem is to only interrupt the current flow for very
short periods of time (typically less than a second) and to be able to recordthe pipe-to-soil voltages (with respect to the reference electrode) during the
current OFF periods. If pipe-to-soil voltages are also recorded during thecurrent ON periods, the magnitude of the IR drop canbe determined,
which would just be the difference between the current ON and currentOFF voltages.
In terms of CIS measurements, a convenient way to sample or capture
current ON and current OFF pipe-to-soil voltages is to employ aCurrent Interrupter which is a current-switching device that is connected
to a Rectifier. The Current Interrupter can be programmed to switch therectifier current ON and OFF in a predetermined cyclic fashion.
In this case, when a cane is in contact with the soil and the push button onthe cane is pressed, in the process of a CIS, the DA Meter will record the
High and Low voltages taken over at least one current ON/OFF cycle.
Such a Current Interrupted method is known as a High/Low CIS mode.
25
-
7/25/2019 DACIS
26/85
It should be noted that all rectifiers that might have an affect on a pipeline
under survey (ie, might generate current in the line) need to simultaneouslybe switched ON and OFF in an identical cyclic fashion and, in addition, all
current-interrupter waveforms (ON/OFF) cycles need to be synchronized
with each other. In other words, all rectifiers need to be ON at the sametime and for the same length of time, and all rectifiers need to OFF at thesame time and for the same length of time. One way to accomplish suchsynchronization, is to have GPS receivers integrated with current-interrupter
switches and to employ a GPS satellites timing system to control the current
switching timing on all participating rectifiers simultaneously.
Such GPS Receiver/Current Interrupter Switches are available through
MCM. This type of current-interrupted method is known as a GPS
Satellite-Synchronized High/Low CIS mode.
Figure 2 shows an illustration of a CIS being performed using GPS satellite-
synchronization of rectifier current-interrupter switches. Again, theON/OFF current switching cycle would be synchronized for all rectifiers
having an affect on the pipeline section being surveyed.
26
-
7/25/2019 DACIS
27/85
Figure 2: A GPS satellite-synchronized high/low CIS mode
illustration
Of course, a CIS can be performed with rectifiers permanently ON (current
permanently flowing in the soil during measurements) and this method is
known as a Non-Interrupted CIS mode, since, in this case, the rectifiercurrent is non-interrupted.
Finally, a CIS can be performed with rectifiers permanently OFF, which
could be the case, for example, if a potential profile is desired of a depolarized pipeline, ie, a pipeline whose polarization potential has had achance to fully dissipate. Under such conditions, however, the pipeline
would not be cathodically-protected.
27
-
7/25/2019 DACIS
28/85
Rectifier Mode Selection:
By tapping once on the pull-down list arrow button in the Rectifier Modefield on the Setup 1 of 5 screen, the screen shown below will appear.
As can be seen on the above screen, there are 3 choices for RectifierMode: Cycle ON/OFF, Always ON and Always OFF.
If Always ON is selected, it will be assumed that the rectifiers affecting the
line will be permanently ON, which would be the case, for example, for a
non-interrupted CIS.
If Always OFF is selected, it will be assumed that all rectifiers having anaffect on the line will be permanently OFF (at least for the duration of the
CIS).
If Cycle ON/OFF is selected, it will be assumed that all rectifiers having an
affect on the line have current-interrupters installed and that their switchingwaveforms will be synchronized. This would be the case for a High/Low
CIS.
Step 10:
Tap once on the Next button.
The screen shown below will appear (Setup 2 of 5).
28
-
7/25/2019 DACIS
29/85
Step 11:Make Selection of Maximum Acceptable % Difference between Far and
Near Ground Readings.
When you are reconnecting the trailing wire from one test station to another
test station further down the pipeline, the last voltage recorded prior to
breaking the connection with the first test station should be recorded as aDevice voltage. After tapping on the Device button on the survey screen and selecting Reconnect, follow the predetermined voltage
recording procedures for a Reconnect. The specific procedures for a given Device, such as a Reconnect, will have been scripted via the software.
The voltage recorded prior to disconnecting the trailing wire will be
considered the FAR-Ground reading, since you will be FAR (perhaps a mile) from the first test station. Once the reconnect has been made at the
second test station, you will be prompted to make a new voltage recording.This new voltage recording will be considered the Near-Ground reading,
since you are standing very NEAR to the second test station.
By tapping once inside the field marked Max. acceptable Far and Near
Ground Reading Difference %, the number indicated (in the above case the
number shown is 5) can be changed to any % value desired by pressing the
delete (DEL) key on the keyboard and typing in a new number.
29
-
7/25/2019 DACIS
30/85
If, for example, the number (5) is entered in this field, it means that the
maximum allowable difference between FAR and NEAR voltage recordingshas been set at 5%, otherwise an error window will be displayed on the
screen. If the FAR and NEAR voltage recordings are within 5% (in this
case) of each other, there will be no error message. If, on the other hand, therecordings differ by more than 5%, an error window will appear on thescreen. The error window will indicate 2 options:
Option 1:
Option 2:
Retake Near Reading
Accept This Error
By selecting Option 1, you have the chance to make another recording at the
second test station which would be compared to the stored FAR-Ground
recording. If this voltage difference is
-
7/25/2019 DACIS
31/85
This selection pertains to a GPS receiver that can be connected to the DA
Meter. The GPS receiver can be called upon to log the precise location of
Devices or Geographical Features that are encountered during a CIS.
As can be seen from the screen, there are 4 choices for GPS Type:
None:
MCM:
NMEA:
Manual:
This means that a GPS receiver is not being used
This means that an MCM GPS receiver is being used
This means that a GPS receiver (other than the MCM receiver)
that is capable of outputting NMEA data is being used
This means that location data will be entered manually when
the GPS button is pressed on the survey screen during a CIS.
Select the appropriate choice by tapping once on your selection.
Step 14:Select GPS Options.
On the far right-hand side of the Setup 2 of 5 screen as shown above, there is
a group of boxes associated with a GPS accessory. If a GPS accessory hasbeen selected for use with the DA Meter for a particular CIS, all, or some ofthe functions available can be enabled (boxes ticked). A box can be ticked
or unticked by tapping once on the box.
The options available are as follows:
31
-
7/25/2019 DACIS
32/85
Differential GPS Required:
This box should be ticked if you only want differentially-corrected (real-
time corrected) GPS data to be logged by the DA Meter. If this box is
ticked, only GPS data that has been derived using a correction message
signal (in real-time) will be logged by the DA Meter. If this box is leftunticked, it means that you will allow the DA Meter to log either standardGPS data or differentially-corrected GPS data, depending upon what is
being output by your receiver at any given time. Please note that if this box
is checked (ticked), GPS data will not be logged unless a real-time
correction signal has been received by the correction message receiver
component of your GPS unit. In most instances, logging standard (ie,uncorrected) GPS data would be preferable to not logging any GPS data.
Use GPS Altitude:
If this box is ticked, altitude data will be included with the position datawhenever GPS data is logged. (Note: Altitude data on some GPS units is not
particularly accurate in CIS applications).
Log GPS at Flags:
If this box is ticked, GPS location data will be logged automatically at flagswhen either the flag button is tapped (directly on the survey screen) or when
the push-button on the designated flag cane is pressed.
Log GPS at DCP/Feature:If this box is ticked, GPS position data will be logged automatically at
Devices or Geo-Features when either the Device button is tapped onthe survey screen and a Device reading is logged or when the Geo-Feat.
button is tapped on the survey screen and a geo-feature is noted.
Auto Log GPS:
By tapping on the drop down menu button in the Auto Log GPS field, the
selections available will be displayed as indicated below.
32
-
7/25/2019 DACIS
33/85
By selecting one of these options, you can elect to have the GPS position
data logged automatically at every survey reading, at every second reading,at every fifth reading, at every tenth reading, or not at all (never) at survey
readings.
Step 15:Tap once on the Next button
The specific screen that will appear (Setup 3 of 5 screen) will depend on if
you selected Trigger CIS mode or Continuous CIS mode back at
Step 6 and whether or not you selected Metric units.
If Trigger CIS mode was selected, for example, the screens shown below
will appear for the non-metric and the metric cases.
33
-
7/25/2019 DACIS
34/85
Step 16:
Select Voltage Reading Interval (Distance Between Recordings)
By tapping once in the field in the Setup 3 of 5 screen labeled Distance Per
Reading you can typein the voltage reading interval (distance in feet (or
meters) expected between recordings) for the CIS. Typically, in CIS work
this expected interval distance is 2.5 feet (or 1 meter).
Step 17:Select Survey Flag Interval (Distance Between Survey Flags)
By tapping once in the field in the Setup 3 of 5 screen labeled Distance
Between Flags, you can type in the survey flag interval (distance betweensurvey flags) for the section of pipeline being measured. Typically, survey
flags are located at 100 feet intervals (or around 30 meter intervals).
Step 18:Select the maximum permissible error between the actual number of
recordings made between 2 survey flags and the expected number ofrecordings.
34
-
7/25/2019 DACIS
35/85
By tapping in the field labeled Flag Dist Error Limit % you can type in the
maximum permissible error. For example, the maximum permissible erroris indicated as 20% on the above screens. If the recording interval is
expected to be 2.5 feet and the survey flag separation is 100 feet, it means
that 40 recordings are expected. If, however, only 30 recordings areactually made between survey flags, an error window will appear on the screen, since there is a 25% difference between the expected and actualnumber of recordings made. No error window will appear if the difference
is less than 20% for this example, ie, you could have a minimum of 32
recordings and a maximum of 48 recordings between survey flags to stay
within the 20% (max.) error allowance.
Step 19:
Select whether or not you would like the recordings to be uniformly spacedbetween survey flags in cases where the actual number of recordings madebetween flags deviates from the projected number (flag separation
recording interval).
By tapping in the box labeled Auto Pacing Mode, and inserting a tick in
the box, you will enable the DA Meter to automatically adjust the actual
recordings and space them evenly over a 100 feet span (distance between
survey flags), regardless of the actual number of recordings made. This
selection is highly recommended.
Step 20:Select the number of reference electrodes (canes) that will be used in the CIS
The number of reference electrodes (also known as data probes) that will be
used in the CIS can be selected by tapping on either the one or two
circle in the box labeled Number of Data Probes.
Step 21:
Provide the Name of the Pipeline:
By tapping once in the field labeled Name of P/L, you can type in the
pipeline name. Note: This is not the same as the filename for the CIS thatwas selected back in Step 4. This is the actual name of the pipeline.
35
-
7/25/2019 DACIS
36/85
Step 22:Provide the Valve Segment Identification Number (or Name)
By tapping once in the field labeled Valve Segment, you can type in the
valve segment number, or name (if known).
Step 23:Provide the Starting Location for the Survey
Tap the pull-down list arrow button in the Starting Location field, which
will reveal the options available regarding how you would like locationinformation to be displayed. The Setup 3 of 5 screen will be as shown
below for the non-metric case.
You can select to have location information displayed on the survey screen
as station number, feet or milepost (station number, meters or kilometers forthe metric case).
Whichever selection you make here will determine how you enter your
starting location information.
For example, if your pipeline locations are represented by station numbers,you would select Station Number from the drop down list and you would
enter a starting location for the survey in the form of a station number. [Ifyou do not know the station number where youre beginning your survey,enter 0+0.0].
36
-
7/25/2019 DACIS
37/85
As an example, if you are working on pipeline ABC within valve segment
45 and you are about to begin a CIS at station number 12+0.0, your Setup 3of 5 screen would be as shown below.
Step 24:Tap once on the Next button on Setup 3 of 5 screen
The screen shown below will appear (Setup 4 of 5 screen).
Step 25:Provide the Work Order Number for the CIS
By tapping once in the field labeled Work Order #, you can type in the
work order number for the CIS if appropriate.
37
-
7/25/2019 DACIS
38/85
Step 26:Provide Your Name
By tapping once in the field labeled Technician Name you can type in
your name.
Step 27:Provide Comments
By tapping once in the field labeled Comments/Description, you can enter
any comments you might have regarding the survey (perhaps weatherconditions, soil conditions etc.).
Also shown on the Setup 4 of 5 screen are the File Name and the date/time.
Note: Do not attempt to change the File Name indicated here as this
identification will be required by the ProActive software to transfer
your CIS data to your PC or server.
Step 28:Make Voltmeter Read Mode Selection
Tap once on the Next button on the Setup 4 of 5 screen
The Setup 5 of 5 screen will appear as shown below.
38
-
7/25/2019 DACIS
39/85
The above screen is used to establish the mode in which the digital voltmeter
will be operating during the CIS measurements. Remember, the data beinglogged by the DA Meter are actually voltages that are measured by a
voltmeter.
The 5 choices available for the voltmeters reading mode are observed by
tapping once on the pull-down-list arrow button in the Read Mode field.The screen will appear as shown below.
The 5 choices for voltmeter mode are as follows:
Note: Selection of voltmeter mode should be made with reference to
your selection of Rectifier Mode back at Step 9 and your selection ofGPS Accessory back at Step 13.
Single Read:
This voltmeter mode would be selected if you were performing a
non-interrupted CIS, with either the rectifiers ON continuously [AlwaysON] or OFF continuously [depolarized pipeline].
On/Off Pairs (DSP mode):
This voltmeter mode would be selected if you were performing a
High/Low CIS (current-interrupted CIS). In this mode, the voltmeter usesdigital signal processing to determine the voltage during the ON portion of
the rectifier-current cycle and the voltage during the OFF portion of thecurrent cycle. A measurable difference between the High and the Low
voltage readings is required for this mode, ie, a measurable IR drop is
required.
39
-
7/25/2019 DACIS
40/85
With this voltmeter reading mode selected, after tapping on the OK
button, the screen shown below will appear.
The specific On and Off times for the particular current-interruption
waveform cycle that will be employed during your CIS should be entered inthe Cycle (ms) fields. The On and the Off times should be entered inmilli-seconds.
Note: Ideally, the interruption waveform period (which is the On time plus
the Off time) should be less than the time between triggered voltage
recordings on a CIS. For short waveform periods (< 1 second, for example)
this is not a concern, as a typical walking pace would translate to a time
between triggered voltage recordings of greater than 1 second.
However, care should be taken if longer waveform periods are in effect (for
example, several seconds long waveform periods), in which case a slowerthan normal walking (and triggering) pace would have to be employed to
avoid recording several same value readings until such times as a newwaveform cycle has occurred.
On/Off Pairs (Min/Max mode):
This voltmeter mode would be selected if you were performing a High/Low
CIS but the DA Meter was having trouble distinguishing between the Highand the Low voltage readings. This would be the case, for example, if the
IR drop was very small (remember that the difference between High and
Low voltage readings in a current-interrupted CIS is just the magnitude of
the IR drop). In this case, the voltmeter would determine the average of the
maximum and minimum voltages read during each rectifier-current cycle.
40
-
7/25/2019 DACIS
41/85
With this voltmeter reading mode selected, the screen shown below will
appear.
Again, you would make your selection of ON and OFF cycle times. In this
case, you would also provide a number for the Moving Average Samples.In the above case, since the number 4 has been entered, 4 sample readingswould be used to compute an average value for the Minimum (Min) voltage
reading during each OFF portion of the cyclic waveform and, also, 4 sample
readings would be used to compute an average value for the Maximum
(Max) voltage reading during each ON portion of the waveform.
On/Off Pairs (GPS Sync):
This voltmeter mode can be selected if you are using the MCM GPS unit onyour DA Meter and current-interrupter switches equipped with GPS unitswere being employed on your rectifiers. In this case, voltage readings
(High/Low readings) are synchronized with the current-interrupter switchingtiming.
With this voltmeter reading mode selected, the screen shown below will
appear.
41
-
7/25/2019 DACIS
42/85
Again, you would make your selection of ON and OFF cycle times.
Also, in this case, you would tap on the GPS Settings button, which would
bring up the screen shown below.
You are being asked here to make several selections.
First, select the type of GPS unit you will be using (currently, only the MCM
GPS unit can be used in the GPS Sync voltage reading mode).
42
-
7/25/2019 DACIS
43/85
Next, enter your On Delay and your Off Delay selections by tapping in
the appropriate field and typing in the requested delay time in milliseconds.These delay times are employed so that any spiking in the voltage waveform
that might occur as the rectifier-current is switched from ON to OFF and
from OFF to ON does not become a factor in the voltmeters determinationof the true ON and OFF voltage values for each CIS measurement. Forexample, if 150 ms was selected for the Off Delay, the data logger wouldrecord the voltage value sampled 150 ms after the rectifier-current was
switched from the ON to the OFF state. Also, if 150 ms was selected for the
On Delay the data logger would record the voltage value sampled 150 ms
after the rectifier-current was switched from the OFF to the ON state.
Finally, select your current-interrupters Downbeat frequency. As
indicated by tapping on the pull-down-list arrow button in the Downbeat
field, there are 3 selection choices; Each Minute, Each Hour and Midnight.If Each Minute is an option for your current-interrupter, we suggest
making this selection. Also, if your interruption cycle starts with the current
in the ON state (the first transition is from ON to OFF), place a tick in the
Start Cycle box (remove the tick if the opposite is true).
Single Read (GPS Timing):
This voltmeter mode is similar to the Single Read mode except that in thiscase the data logger uses the GPS clock signal to associate voltage
recordings with the actual time they were made.
With this voltmeter reading mode selected, the screen shown below will
appear.
43
-
7/25/2019 DACIS
44/85
The only additional information that needs to be supplied here is the type of
GPS unit you will be using, which you can select by tapping on the GPSSettings button. Currently, only the MCM GPS unit can be used in this
voltmeter reading mode application.
Step 29:Select Voltmeter Range and Input Impedance
By tapping on the pull-down list arrow button in the Range field, you can
select the voltmeter range and input impedance settings for your application.
The recommended settings for the voltmeter for CIS are 5.7 Volts DC
(full-scale) and an input impedance value of 400MOhm, These settings
provide a relatively-fast response time (~80 ms), which is important inrectifier-current-interrupted applications. Also, a 400MOhm input
impedance setting eliminates reading errors associated with any high
source resistance measurements.
Note: Caution has to be exercised if any of the ranges beginning with the
number 4 (for example, 4V, 400mV or 40mV) are selected for CIS
applications, since these represent relatively-slow response settings. Suchsettings provide a higher level of AC filtering, but at the expense of speed of
response. Rather than the response time of 80ms for the 5.7V range, these
settings have a response time greater than 1 second. Consequently, suchsettings are not appropriate for fast current-interruption cycles, ie,
interruption On and Off times shorter than 1 second, as true On and Off
potentials would not be recorded.
Step 30:Pull Up Active Survey Screen.
Tap once on the OK button on the Voltmeter Setting screen (or the GPS
Settings screen), which will pull up the Survey screen shown below.
44
-
7/25/2019 DACIS
45/85
As voltages are recorded by the DA Meter (by either tapping on the Readbutton on the Survey screen or by pressing the appropriate cane push button,
the Distance From Start (total distance from the start of the survey)
parameter will increase in increments of 2.5 feet, or whatever the Distance
Per Reading value was that was entered back at Step 16. (Distances will bein meters, if the metric option was selected).
Also, the Distance From Know Station parameter will increase in the same
increments as voltages are recorded. The difference in this case, however,
will be that when each known station is registered, this distance parameter
will begin again at zero. In other words, this will show the distance you are
assumed by the DA Meter to have traveled from the last known station thatyou encountered (and registered).
You are now ready to perform a close interval survey using the
DA Meter.
45
-
7/25/2019 DACIS
46/85
III. 3 How to Make Changes to Your Set Up Selections
Once youve setup the DA Meter for a particular Close Interval Survey, as
described in the above Section, you can make changes to your setup
selections with the exception of your Survey Type selection and the
currently indicated Location. If you have selected a Trigger Mode CIS,for example, you cannot change to a Continuous Mode CIS, without
setting up a new survey.
You can, however, make other selection changes. For example, if you
examine the pipe-to-soil voltage waveform prior to beginning a survey(highly recommended) and discover that the IR drop magnitude is quite
small, or there is considerable noise on the waveform, you might decide to
switch the voltmeter reading mode from On/Off Pairs (DSP mode) to On/OffPairs (Min/Max mode). Or, in another case, you might decide to change theDistance Per Reading interval from 2.5 feet to 5.0 feet, for example.
To make setup selection changes, tap on Survey Options at the top of the
screen. The screen will appear as shown below.
To make (allowed) changes, tap on Settings on the above menu, tap on Change Global Settings and proceed to make any desired changes.To simply view your Set Up selections, you can tap on View Settings on
the above menu (after tapping on Survey Options).
46
-
7/25/2019 DACIS
47/85
SECTION IV: TEST EQUIPMENT HOOK-UPS FOR CIS
IV. 1 How to make Cable Hook-Ups for CIS Measurements
The connections for CIS measurements employing MCM test equipment are
illustrated in Figure 4.
Figure 4: Cable Connections for Hook-Up of MCMs CIS
Test Equipment
47
-
7/25/2019 DACIS
48/85
As can be seen from Figure 4, a pair of canes (reference electrodessee
Section II. 4) is illustrated, the left-hand cane (red cane) and the right-handcane (green cane). These canes, which will be placed on the soil above the
pipeline in an alternating fashion, have push buttons on top of the handles so
that the operator can trigger voltage recordings on his command at each ofthe CIS measurement locations as well as at Devices. See Section III. 2for a discussion on cane button functionality (Step 7 of the Set Up Process).
The canes (data probes) are connected as shown to the input terminals of
the dual-probe adapter and the output terminal of the adapter is connected
to a 5-pin Data Probe Connector socket on the top side of the DA Meter.This effectively connects the canes (reference electrodes) to the negative
terminal of the data-logger (DA Meter). For CIS, the cables connecting thedata probes to the dual-probe adapter are both black band cables.
In order that the DA Meter will record pipe-to-soil voltages when the
canes are triggered, the positive (red) banana plug terminal on the back ofthe DA Meter is connected as shown to the pipe via a spool of wire
contained in a Hip Pack. The connection to the pipe is made by hooking acable from the red terminal on the DA Meter (banana plug terminal) to the
banana plug terminal on the underside of the Hip Pack and connecting the
trail wire coming out of the topside of the Hip Pack to the Test StationElectrode (and therefore to the pipe).
A GPS receiver unit is also illustrated in Figure 4. Such a unit can be
connected to the DA Meter in order that the position of items such as flags,devices and geo-features can be recorded during the performance of a CIS,
either manually by tapping on the log GPS button on the survey screen at
each critical location or automatically by pre-programming the DA Meter as
described in Section III. 2 (Steps 13 and 14).
IV. 2 How to Attach Cables and Accessories to the DA Meter
The terminals on the back of the DA Meter for the various connections
described above are illustrated in Figure 5.
48
-
7/25/2019 DACIS
49/85
Figure 5: Connection terminals on rear of the DA Meter
The 5-pin Data Probe Connector is shown above as Terminal 1. As
discussed above, the reference electrode canes are connected to this terminal
via the dual probe adapter. Again, the reference electrodes are effectively
connected to the negative terminal of the voltmeter with this connection.
(Consequently, the black banana plug terminal (Terminal 4) is not usedwhen push button canes are employed)
The red (positive) banana plug terminal (for connection to the hip pack) is
shown as Terminal 3. This will establish connection to the positive
terminal of the voltmeter.
Terminal 5 is a GPS Accessory terminal. A GPS receiver capable of
outputting NMEA data (such as a Trimble Pathfinder unit or the MCM unit)
would be connected to Terminal 5 (9-pin terminal). Terminal 2 is currently
earmarked for future use.
Finally, Terminal 5 is also used (after performing a CIS and removing theNMEA GPS receiver, if used) to connect a serial data communication cableto a PC in order that the survey data recorded by the DA Meter can be
uploaded to a PC. See Section VI for a discussion on how to transfer(upload) CIS data to a PC for data management.
49
-
7/25/2019 DACIS
50/85
SECTION V: HOW TO PERFORM CLOSE INTERVAL
SURVEYS
V. 1 How to Carry the Test Equipment During a CIS
With the equipment connected as shown in Figure 4, and the DA Meter set
up as described in Section III. 2, you are now ready to perform a Close
Interval Survey.
To make a pipeline survey more manageable, MCM has developed a special
harness which allows the DA Meter and the trail-wire Hip Pack to be carriedaround the waist area in a hands-free fashion, allowing the individual to be
able to position the reference electrodes (canes) over the pipeline (every 2.5
feet, or so, in an alternating manner) and to be able to trigger the push button canes when appropriate to do so.
With the harness assembly, the DA Meter sits on a tray at waist level
allowing the operator to view the screen at all times and to make any
selections required by tapping on the screen. Also, the dual-probe adaptershown in Figure 4, is attached to the underside of the tray, allowing
convenient (5 pin cable) connection of the adapters output to the DAMeter. [Again, this will effectively connect both data-probes (reference
electrode canes) to the negative terminal of the voltmeter].
The Hip Pack, which contains a spool of copper wire (trail wire) [typically 1
mile or 3 miles in length], is attached to the waist belt of the harness and, as
described in Section IV. 1, a cable connects the banana plug terminal on theunderside of the Hip Pack to the red (positive) banana plug terminal on theDA Meter. Typically, the Hip Pack sits on the operators left-hand side at
waist level.
In addition, a GPS receiver unit (together with its battery pack and antenna)
can also be attached to the waist band of the harness, typically on theoperators back at waist level, with the antenna rising to above-head height.
50
-
7/25/2019 DACIS
51/85
V. 2 How to Examine and Record a High/Low Voltage Waveform
When you are performing a High/Low (current-interrupted) CIS, it is
recommended that you examine the pipe-to-soil voltage waveform at your
starting location (starting test station) and it is suggested that you make arecording of this voltage waveform using your DA Meter. The nature of thevoltage waveform that you are examining will reflect the nature of therectifier-current waveform that is currently in-effect on your pipeline.
With at least one of the two reference electrodes making good electrical
contact with the soil above the pipe at the first test station and a test cableconnected from the test station to the red banana plug terminal on the DA
Meter (the cable connection to the Hip Pack will be temporarilydisconnected), you will see voltage readings displayed for the ON and the
OFF portions of the current-interrupted waveform cycle for each cyclesee screen shown below for a snap-shot.
To examine the ON/OFF pipe-to-soil voltage waveform, tap once on the
WAVE button on the menu bar at the top of the Survey screen.
Alternatively, tap on the Survey Options button and then tap on Wave.
The screen shown below will appear.
By tapping once on the Read button on the above screen, you can view the
actual ON/OFF voltage waveform. A typical waveform would be as shownin the following screen.
51
-
7/25/2019 DACIS
52/85
As can be seen from the above screen, the voltage reading is High for 0.7
sec. and Low for 0.3 sec. which was the prescribed cycle for the rectifier-
current in our example.
It can also be noted by examining the above pipe-to-soil voltage waveform
that the transitions from High-to-Low and from Low-to-High readings arequite sharp (as opposed to being step-like). This means that all rectifiershaving an affect on this section of pipe are being switched ON and OFF in a
synchronized fashion, which is the ideal case.
To record this waveform, tap once on the save button on the above screen.
The Save As window shown below will appear on the screen.
52
-
7/25/2019 DACIS
53/85
You are being asked here to establish a file in which the waveform data will
be stored (saved). The name of this file will be the name you type into theName box on the above screen. For ease of future retrieval when you
have uploaded your survey data to the Data Management software package
(ProActive) [see Section VI], it is recommended that you use the same nameas your survey filename with the addition of something like Wave TS 1 tothe file name. This would represent the waveform data recorded at TestStation 1 at the beginning of your survey.
You should then tap once on the OK button. This process will save the
waveform data in the DA Meters CompactFlash memory in the DA
Meter_DATA folder for future reference.
Tap once on the Exit button on the waveform screen and this will bring
you back to the survey screen.
V. 3 How to Record the Voltages at a Starting Test Station:
At this point, it is recommended that you record the ON/OFF voltages at the
starting test station. This will represent the first Device on this survey.
To do so, tap once on the Device button on the survey screen. The
Device Readings screen shown below will appear.
Note: An additional device, named Mark DCVG will be available for
selection if the box labeled, DCVG Sidedrains in CIS Survey waschecked-off during initial survey set up.
53
-
7/25/2019 DACIS
54/85
As can be seen from the above screen, you can choose from a number of
Devices. In this case, you would tap once on Single Test Station tohighlight this selection. You should then tap once on the Next button.
The screen shown below will appear.
As shown above, the High and Low voltage readings measured at the
starting test station will be displayed on this screen. You have the option at this point to tap once on the Accept button and then once on the Save
button to record the data at the first test station. As an alternative, you couldpress the trigger button on either your right-hand or left-hand cane (reference
electrode) to record the data, assuming that you selected the accept save option for the canes at D.C.P.s (see Section III. 3, Step 7).
Saving the first test station voltages will bring you back again to the Survey
screen. You are now ready to proceed with the CIS.
V. 4 How to Perform a CIS
In order to illustrate some of the important processes associated with
performing a CIS, we have designed a fictitious section of pipeline and the
steps outlined below with regard to conducting a CIS pertain to this fictitioussection. Our 350 feet long section of pipeline is shown below in Figure 6.
54
-
7/25/2019 DACIS
55/85
Figure 6: Fictitious Section of Pipeline
The important features on this section of pipe are as follows:
TS1: Test Station 1. Coincides with station 0+00.0 (reference zero station)
Flags: Located at 1+00.0 (100 feet from reference zero station), 2+00.0
(200 feet from reference zero station) and 3+00.0 (300 feet fromreference zero station).
TS2: Test Station 2 located at 2+40.0 (240 feet from reference zero station)Asphalt Road: A 20 feet wide asphalt road located b/w 2+70.0 and 2+90.0
(270 feet from reference zero to 290 feet from reference zero)Valve: Located at 3+50.0 (350 feet from reference zero station)
55
-
7/25/2019 DACIS
56/85
Since, in our example, we have set up the right-hand cane to be the read
cane, voltages will be recorded each time you trigger the right-hand canespush button (by pressing down on the button) as you walk down the length
of the pipeline. Triggering a recording using the right-hand cane button, in
this case, would have the same effect as tapping on the Read button on theabove screen.
Since you recorded voltages at the starting test station (TS1 in Figure 6), you
have already established a Device location and, as shown below, thisdevice is identified in the graph on the survey screen by the letter D. In
fact, all Devices encountered as you go down the length of the surveysection will be marked on the graph in this fashion. Please note, however,
that this first D marker will disappear from the graph as soon as the firstsurvey measurement is recorded. All subsequent Device locations,
however, will be indicated permanently with a D marker on the graph inthe survey screen.
Since we indicated during the DA Meters setup process that we expected to
record voltages every 2.5 feet, each time you trigger readings using the
right-hand canes push button, the data logger will assume that you havetraveled an additional 2.5 feet down the line section. Consequently, youshould try to place the electrodes over the pipe at 2.5 feet intervals.
For instance, the screen will appear as shown below after the second
triggering event (it is assumed that your first reading (first triggering eventwill be at the starting location).
56
-
7/25/2019 DACIS
57/85
As can be seen from the above screen, the DA Meter assumes that you are
now at a distance of 2.5 feet from where you started, which was at startingstation 0+0.0 (or, 0+00.0).
The DA Meters software program will add 2.5 feet to the assumed distance
from the starting station every time the right-hand cane button is triggered.
If you recall, we also set things up so that when the left-hand canes pushbutton is pressed (triggered), we would record the location of survey flags,
which are typically placed at 100 feet intervals down the length of the pipe.In this case, triggering the left-hand cane would have the same effect as
tapping once on the Flag button on the survey screen. Consequently,when you encounter your first survey flag and trigger the left-hand cane, the
DA Meters software will assume that you have traveled a distance of 100feet from the starting test station. (A flag is considered a Known Station)
At this point, the screen will appear as shown below.
57
-
7/25/2019 DACIS
58/85
Notice that the station number is now 1+00.0 The 1 represents 100 feet,
so the DA Meter assumes that you are now 100 feet from your referencezero location. Notice also the F marker on the graph (for Flag).
If you continue on down the pipe triggering voltage recordings using the
right-hand canes push button until you encounter the next survey flag andyou then press the left-hand canes push button to register this second survey
flag, the screen will appear as shown below.
In this case, the station number is now 2+00.0 The 2 represents an
assumed 200 feet from the reference zero location. A second Flag markeralso appears on the graph.
Lets now assume that you encounter a second test station (TS2 in Figure 6)
at a distance of 40 feet from the second survey flag location and you need todo a reconnect of the trail wire at this location.
58
-
7/25/2019 DACIS
59/85
In this case, when you encounter the test station, you would tap once on the
Device button on the survey screen and then highlight reconnect. The
Device Readings screen at this point will appear as shown below.
By tapping once on the Next button, the screen shown below will appear.
You are being prompted here to accept and save the Far Ground voltages
prior to breaking the trail wire connection (see Section III. 2, Step 11 for anexplanation of Far- and Near-Ground readings). By doing so, the screen
shown below will appear.
59
-
7/25/2019 DACIS
60/85
At this point you are being prompted to take Metal IR readings at the test
station. To do so, you would connect a test lead from the black banana plugterminal on the DA Meter to the test station and disconnect the reference
electrode data-probes (you could place the ceramic plug ends of the probeson top of your shoes to provide an open circuit). At this point, the DA Meterwill be reading the potential difference between the two test stations (the far-
ground station and the near-ground station). If youd like to record theMetal IR data, you would save the readings by tapping on accept and then
save after setting up to take the readings.
The next screen to appear is as shown below.
In this case, you are being prompted to accept and save the Near Ground
voltages after connecting the trail wire to the second test station (TS2). If
you recorded the Metal IR data, you would first disconnect the test lead fromthe DA Meter and then reconnect the data-probes.
60
-
7/25/2019 DACIS
61/85
If the Far- and Near-Ground voltages are within the maximum allowable %
difference range that you established during your setup procedure (seeSection III. 2 step 11), you will be returned to the Survey screen as shown
below. If, on the other hand, the Far- and Near-Ground voltages differ by
more than the allowable range, a window will appear showing all fourreadings and asking you whether or not you will accept the readings as is, orif you would like to re-take the near-ground readings.
Now, lets assume that at an additional 30 feet down the pipe from the
reconnect location, you encounter a Geo-Feature, ie, a geographical
feature of some type. In this case, when you reach this location, you wouldtap once on the Geo-Feat. button on the survey screen. At this point, the
screen would be as shown below.
You can select the type of geo-feature that you have encountered by tappingonce on the pull-down-list arrowbutton in the Feature field.
61
-
7/25/2019 DACIS
62/85
Alternatively, you could perform a quick search for the feature by typing in
the Quick Search field the number associated with the feature or the first
few letters of the name of the feature.
If, for example, the geo-feature is an Asphalt Road, and the width of the
road that you have to cross is 20 feet, you can enter a Feature Length of 20feet (see Figure 6). You could restart voltage recordings immediately on the
other side of the road. However, if there is a survey flag an additional 10feet beyond the other side of the road, you might want to restart making
voltage recordings at the flag location. If this were the case, the survey
screen would appear as shown below.
By tapping once on the Save button, the survey screen shown below will
appear.
62
-
7/25/2019 DACIS
63/85
As can be seen from the above screen, the DA Meters software assumes
that you are 2.5 feet short of the survey flag on the other side of the asphaltroad and that your next voltages will be recorded when you trigger the
read cane when you are actually at the survey flag location. Notice that a
G marker has been added to the graph (for Geo-Feature).
By triggering the read cane (your right-hand reference electrode cane), the
survey screen shown below will appear. You can also trigger the flag
cane (your left-hand reference electrode cane) at this time, to register thesurvey flag.
Now lets assume that you encounter a Valve at an additional 50 feet
down the pipe from this survey flag (see Figure 6). Since, a valve isconsidered a Device, you would stop and tap once on the Device button
on the survey screen and then tap on Valve. The screen shown belowwould appear.
63
-
7/25/2019 DACIS
64/85
In this case, you would record the pipe-to-soil voltages at the valve location
by triggering the right-hand cane (or by tapping once on the Accept buttonand then once on the save button on the survey screen).
Having done so, the screen shown below will appear.
This screen allows you to note the condition of the valve and to make any
relevant comments about the valve.
After tapping once on the Save button, the Survey screen shown below
will appear.
Since, in our fictitious section of pipe, the valve represents the end of the
section, you would terminate the survey by tapping once on Survey in the
menu bar and selecting Finish Survey.
64
-
7/25/2019 DACIS
65/85
At this point you have completed the CIS on this section of the pipeline and
the survey data have been stored (saved) in the DA Meters memory in thefile that you named at the outset of the CIS. You are now ready to upload
the survey data to MCMs Database Management System (see Section VI).
V. 5 How to Use the GPS Buttons, if Active, on the Survey Screen
If you selected a GPS receiver type during your setup process (Section III. 2,
step 13), the two GPS button will be active on the Survey screen as shown below.
If you have a GPS receiver attached to your DA Meter (via the COM 1
Port), and the receiver has been properly-configured to output NMEA datastrings (in the GGA formatsee your DA Meter Users Manual, Appendix
1, if you are using a Trimble GPS Pathfinder receiver), you will be able toconveniently examine position data in real-time at any given location on
your survey using these GPS buttons.
By tapping once on the GPS Data button, you can view a summary of the
GPS data associated with your current location. You can also see how many
satellites (Sat. In Use) were used in the determination of the position data for
your present location. Ideally, a minimum of 5 satellites should be involved
for good position accuracy.
65
-
7/25/2019 DACIS
66/85
You can also view the NMEA data output by the GPS receiver by tapping
once on the View NMEA Data button on the above window. New
NMEA data will stream into the DA Meter every second. You can freeze
this data by tapping once on the Stop button, which will allow you toexamine the NMEA data strings.
You can also log (record) GPS data at any location on your survey, by
tapping once on the Log GPS button. Alternatively, as discussed inSection III. 2, Step 14, you can pre-program the DA Meter to automatically
log (record) GPS data at items such as Flags, Devices and Geo-Features.
With reference to Section III. 2, Step 14 (setting-up the DA Meter with
GPS choices):
If you placed a tick in the box labeled Log GPS at Flags, GPS data will be
logged automatically at survey flags when either you tapon theFlag button on the survey screen or when you trigger the flag designated
cane.If you placed a tick in the box labeled Log GPS at DCP/Feature, GPS datawill be logged automatically at Devices or Geo-Features when you tap
the Device button o