LMIC Leica SP5 BASIC use guide (Version 2)

6
10/2010 LMIC Leica SP5 BASIC use guide (Version 2) Start-up Uncover microscope, tilt illumination arm back, move objectives all the way down (objectives will be down unless you moved them up). Turn on the Hg‐lamp if you will need it. Turn on the laser power switch (switch on the right) and turn on the laser key. Turn on Scan box (middle switch). Turn on computer/microscope. Log‐on, form should come up automatically. If computer is already on click on the icon for “shortcut to iCenter application” to bring up the sign‐in sheet. o add your IU username o the account number to be charged o and the last name of the lab PI. Start software (LAS AF icon on desktop). o Be sure configuration is on “machine”. After stage initialization, move illumination arm back to upright position. You are now ready to look at something.

Transcript of LMIC Leica SP5 BASIC use guide (Version 2)

Page 1: LMIC Leica SP5 BASIC use guide (Version 2)

10/2010

LMICLeicaSP5BASICuseguide(Version2)

Start­up

‐ Uncovermicroscope,tiltilluminationarmback,moveobjectivesallthewaydown(objectiveswillbedownunlessyoumovedthemup).

‐ TurnontheHg‐lampifyouwillneedit.‐ Turnonthelaserpowerswitch(switchontheright)andturnonthelaserkey.‐ TurnonScanbox(middleswitch).‐ Turnoncomputer/microscope.‐ Log‐on,formshouldcomeupautomatically.Ifcomputerisalreadyonclickontheiconfor

“shortcuttoiCenterapplication”tobringupthesign‐insheet.o addyourIUusernameo theaccountnumber

tobechargedo andthelastnameof

thelabPI.

‐ Startsoftware(LASAFiconondesktop).o Besureconfigurationison“machine”.

‐ Afterstageinitialization,moveilluminationarmbacktouprightposition.‐ Youarenowreadytolookatsomething.

Page 2: LMIC Leica SP5 BASIC use guide (Version 2)

Microscope

‐ Itonlyworkswellwhenitisclean!Pleasedoyourparttokeepitclean.‐ Insertslideinslotinz‐galvostageinsert.(Samplepointsdowntowardobjective).Ifyouwillneed

oil,putasmalldroponthecoverslipbeforeyouloadyourslide.o Avoidgettingoilonthenon‐oilobjectives–ifyoudo–askforhelp,wewillcleanthe

objective.‐ Findyoursample–Istartwiththe10Xobjective(ithasalongworkingdistanceandwon’ttouch

theoilwhenitisintherightfocalplane).

ScopeOperation–lotsoflittlebuttons.Softwarecontrolsmostthings,manyareautomatic.

LeftSide:

1. ChangeTL–changestransmittedlightfrombrightfieldtoDIC.

2. ChangestoFluorescencelightpath–Canalsodothisbyjustselectingafiltercubeonfrontpanel.

3. Changesfiltercubes4. ChangeCS–changestoconfocalmode5. Changesaperturediaphragm

(automaticallyadjusted)6. Lightintensityadjustment7. Fielddiaphragm–howmuchoffieldof

viewyousee.8. Brightfieldlightswitch

Center:

1. Alllighttoeyepiece.2. Cameraportselection.Willhappen

automatically.3. Magnificationselection–wedon’thave

thisfeature–leavebuttonalone.4. Ifyoupress#3,pressthisbuttontoget

back.5. Fluorescenceshutter(makesnonoise).6. GreenFluorescencefiltercube7. RedFluorescencefiltercube

8. Analyzercube9. Emptyslotforimaging10. DAPIFluorescencefiltercube11. Emptyslot

Page 3: LMIC Leica SP5 BASIC use guide (Version 2)

RightSide:

1. Pressandholdtomoveobjectiveuptosetposition.

2. Setbutton.Tosettopposition,pressandholdsetbuttonwhilepressingtopbuttononcetoclear,againtoset.

3. Pressandholdtomoveobjectivedown.Pleasedon’tsetbottomposition.

4. Changeobjective(toleft)5. Changeobjective(toright)6. Switchfromdrytoimmersion

objective.

Stage/Focuscontroller(salt&peppershaker):

• Top–stagemovementinX• Bottom–stagemovementinY• Back–Focus• Buttonsonleft–togglebetweenfastand

precisestagemovement.• Buttonsonright–togglebetweenfineand

coursefocus.

Page 4: LMIC Leica SP5 BASIC use guide (Version 2)

LASAFSoftware

‐ Configurationo Turnonlasersyouneed.o SetlaserpowerforArgonlaser(20%isagoodplacetostart).o Lookatobjectivelist–listedforeachisthemax.resolution.Keepthisnumberinmind

whenzooming,etc.(ItisOKtooversampleabout40%).‐ Acquisition

‐ o Establishlightpath.

TheeasiestwayistothisistoselectpreloadedLeicasettings(redarrowabove)thataresimilartoyourneeds.Thiswilldomanythings.Itturnsonlasercontroller,laseroutput.PMTs,andsetswindows.

ThenclinkonfluorophoreselectionbarforeachPMTandpickdyeyouactuallyused.

Adjustcollectionwindowsasnecessary.Makesureyouarenotcollectinglightoverlaseroutput!Windowshouldbeatleast10nmawayfromlaserwavelength.

Page 5: LMIC Leica SP5 BASIC use guide (Version 2)

o Selectobjective(redcircleabove).Payattentiontowhatimmersionmediaeachobjectiverequires.

o PickAcquisitionMode(xyz=“normal”).Theothersdosomenicethings.Askifyouneedhelp.

o Format–pickimagesize.Keepinmindmax.resolutionofyourobjective–andremember,smallerisfaster.

o Speed–defaultis400Hzandisagoodplacetostart.600Hzisasfastasitcanscanwithoutzoomingin.Slowerspeedsequalsbetterqualityandmorebleaching.

o Seewhatyouhave–hitlivebutton–youshouldgetanimage.Taketimetooptimizeimage–maybeonaspotyoudon’treallycareabout.

Focus–Usethez‐controlknoboncontrolpanel–thisistheonlyfocusthecomputerunderstands.

Ingeneral,highgainandlowlaserpowerisgood.Iftoograiny,turnlaserup.Startwithgainaround1000.

Checkforbleed‐through(turneachlaserdown). Turnonquick‐lookuptable(rainbowicononrightscreen).Bluepixelsaremaxed

out(white),greenpixelsareblack.Forbestimagestryafewbluepixelsandafewmoregreen(=mostdynamicrange).

ClickZoom,drawboxonimagewhereyouwanttozoom(remembermax.resolution).

Pinholeisautomaticallysetatoptimum–canopenitifnecessary(morelight,lessconfocaleffect).

Ifdesired,setZ‐series,findtopandbottom. HitCaptureforsingleimage,orStartforseries.

‐ ClickExperimenttab.HitSaveAll(Untilyouhitsaveall–yourimageswillbelostifthecomputercrashes)

ShutDown

‐ Firstturnoffkeyswitchforlasers.Leavelaserpowerswitchon!Thisturnsthelasersoffbutleavesthecoolingfansontoavoiddamaginglasers.

‐ Cleanobjectives!‐ Transferdatatoserverandsign‐out.

o Savealldatatotransfertoonefolder(youcanonlytransferonefoldertoserverperlog‐in)

o click“selectdatadirectory”buttonandnavigatetoyourfolder.

o Fillinatleast“ProjectTitle”soyoucanfindyourdownloadontheserver.

o Hit“Sign‐Out”

Pleasedeleteimageswhenyouaresuretheyaresafelytransferredtotheserver.Youshouldneverhaveimagesonthescopecomputerformorethan1­2sessionsofimaging

‐ Shutdowncomputer(ifyouneedtoworkonthecomputer,youmayleaveiton).

Page 6: LMIC Leica SP5 BASIC use guide (Version 2)

‐ Turnoffscanhead.‐ TurnoffHglamp.‐ Coverscope(don’tcoverHglamp).‐ Cleanuptables,etc.‐ Whencomputerscreensaredarkandthelaserfancable/ventstopshaking(lasercoolingfanhas

stopped)itissafetoturnoffpowerswitchestolasersandcomputer.