Types of X-ray Detectors

Post on 31-Dec-2016

214 views 1 download

Transcript of Types of X-ray Detectors

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation counters

Solid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Types of X-ray Detectors

Gas detectors

• Ionization chamber

• Proportional counter

• Geiger-Muller tube

Scintillation countersSolid state detectors

• Intrinsic semiconductor

• P-I-N junction

• Silicon drift

Charge coupled device detectors

• Indirect

• Direct coupled

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 1 / 10

Gas Detector Curve

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 2 / 10

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/Io = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-holepair (depending on the gas)makes this useful forquantitative measurements.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 3 / 10

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/Io = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-holepair (depending on the gas)makes this useful forquantitative measurements.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 3 / 10

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/Io = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-holepair (depending on the gas)makes this useful forquantitative measurements.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 3 / 10

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/Io = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-holepair (depending on the gas)makes this useful forquantitative measurements.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 3 / 10

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/Io = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-holepair (depending on the gas)makes this useful forquantitative measurements.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 3 / 10

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/Io = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-holepair (depending on the gas)makes this useful forquantitative measurements.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 3 / 10

Ionization Chamber

Useful for beam monitoring, flux measurement, fluorescence measurement,spectroscopy.

• Closed (or sealed) chamber oflength L with gas mixtureµ =

∑ρiµi

• High voltage applied to plates

• Calculate fraction of beamabsorbed I/Io = e−µL

• When x-ray interacts with gasatom, photoionized electronsswept rapidly to positiveelectrode and current (nanoAmperes) is measured.

• Count rates up to 1011

photons/s/cm3

• 22-41 eV per electron-holepair (depending on the gas)makes this useful forquantitative measurements.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 3 / 10

Scintillation Counter

Useful for photon counting experiments

• NaI(Tl), Yttrium Aluminum Perovskite (YAP) or plastic which,absorb x-rays and fluoresce in the visible spectrum.

• Light strikes a thin photocathode which emits electrons into thevacuum portion of a photomultiplier tube.

• Photoelectrons are accelerated in steps, striking dynodes andbecoming amplified.

• Output voltage pulse is proportional to initial x-ray energy.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 4 / 10

Scintillation Counter

Useful for photon counting experiments

• NaI(Tl), Yttrium Aluminum Perovskite (YAP) or plastic which,absorb x-rays and fluoresce in the visible spectrum.

• Light strikes a thin photocathode which emits electrons into thevacuum portion of a photomultiplier tube.

• Photoelectrons are accelerated in steps, striking dynodes andbecoming amplified.

• Output voltage pulse is proportional to initial x-ray energy.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 4 / 10

Scintillation Counter

Useful for photon counting experiments

• NaI(Tl), Yttrium Aluminum Perovskite (YAP) or plastic which,absorb x-rays and fluoresce in the visible spectrum.

• Light strikes a thin photocathode which emits electrons into thevacuum portion of a photomultiplier tube.

• Photoelectrons are accelerated in steps, striking dynodes andbecoming amplified.

• Output voltage pulse is proportional to initial x-ray energy.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 4 / 10

Scintillation Counter

Useful for photon counting experiments

• NaI(Tl), Yttrium Aluminum Perovskite (YAP) or plastic which,absorb x-rays and fluoresce in the visible spectrum.

• Light strikes a thin photocathode which emits electrons into thevacuum portion of a photomultiplier tube.

• Photoelectrons are accelerated in steps, striking dynodes andbecoming amplified.

• Output voltage pulse is proportional to initial x-ray energy.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 4 / 10

Scintillation Counter

Useful for photon counting experiments

• NaI(Tl), Yttrium Aluminum Perovskite (YAP) or plastic which,absorb x-rays and fluoresce in the visible spectrum.

• Light strikes a thin photocathode which emits electrons into thevacuum portion of a photomultiplier tube.

• Photoelectrons are accelerated in steps, striking dynodes andbecoming amplified.

• Output voltage pulse is proportional to initial x-ray energy.

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 4 / 10

Solid State Detectors

Open circuit p-n junction has a natural depletion region

When reverse biased, the depletion region growscreating a higher electric field near the junction

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 5 / 10

Solid State Detectors

Open circuit p-n junction has a natural depletion region

When reverse biased, the depletion region grows

creating a higher electric field near the junction

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 5 / 10

Solid State Detectors

Open circuit p-n junction has a natural depletion region

When reverse biased, the depletion region growscreating a higher electric field near the junction

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 5 / 10

Ge Detector Operation

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 6 / 10

Silicon Drift Detector

Same principle as intrinsic or p-i-n detector but much more compact andoperates at higher temperatures

Relatively low stopping power is a drawback

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 7 / 10

CCD Detectors - Direct

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 8 / 10

CCD Detectors - Indirect

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 9 / 10

CCD Detectors - Lightpipe Taper

C. Segre (IIT) PHYS 570 - Fall 2010 September 21, 2010 10 / 10