1VF TIMS Module
302 Enzo Drive San Jose, CA 95138
Tel: 1-408-363-8000 Fax: 1-408-363-8313
MAN-22448-001 Rev B00
VF TIMS Module
Part of the MTT and xDSLFamily of Products
Users Manual
SSMTT-06BM
SUNRISE TELECOI N C O R P O R A T E D
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
2/44
2 SSMTT-6B
WARNING
Using the supplied equipment in a manner not specified by Sunrise
Telecom may impair the protection provided by the equipment.
CAUTIONS!
Do not remove or insert the module while the test set is on. Insertingor removing a module with the power on may damage the module.
Do not remove or insert the software cartridge while the test set is
on. Otherwise, damage could occur to the cartridge.
End of Life Recycling and Disposal Information
DO NOT dispose of Waste Electrical and Electronic Equipment (WEEE)as unsorted municipal waste. For proper disposal return the product to
Sunrise Telecom. Please contact our local offices or service centers for
information on how to arrange the return and recycling of any of ourproducts.
The Waste Electrical and Electronic Equipment Directiveaims to minimize the impact of the disposal of electrical and
electronic equipment on the environment. It encourages and
sets criteria for the collection, treatment, recycling, recovery,
and disposal of waste electrical and electronic equipment.
EC Directive on Waste Electrical and Electronic Equipment (WEEE)
2010 Sunrise Telecom Incorporated. All rights reserved.
Disclaimer: Contents subject to change without notice.
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
3/44
3VF TIMS Module
VF TIMS Module
1 VF TIMS Module ................................................ 51.1 LEDs ................................................................................. 5
1.2 Module Connector Panel .................................................. 5
2 VF TIMS Main Menu .......................................... 72.1 Configuration .................................................................... 8
2.2 Place Call ......................................................................... 92.2.1 Receive a Call ............................................................. 12
2.3 Transmit Frequency ........................................................ 13
2.4 Receive Frequency and Measurements ......................... 172.4.1 Single Frequency ......................................................... 172.4.2 Frequency Sweep ........................................................ 18
2.4.3 3-Tone Slope ............................................................... 20
2.5 Noise and Impairments .................................................. 222.5.1 Noise/SNR ................................................................... 22
2.5.2 Impulse Noise .............................................................. 25
2.6 Loop Current and Voltage ............................................... 272.7 Call Analysis ................................................................... 28
2.8 Other Setup .................................................................... 302.9 View/Store/Print .............................................................. 362.9.1 Saving a Test ............................................................... 36
2.9.2 Viewing a Stored Test.................................................. 362.9.3 Printing a Stored Test .................................................. 37
2.9.4 Deleting a Stored Test ................................................. 372.9.5 Locking & Unlocking a Stored Test .............................. 37
2.9.6 Renaming a Stored Test .............................................. 37
3 Applications .................................................... 393.1 Express Limited Warranty ............................................... 40
Index ..................................................................... 41
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
4/44
4 SSMTT-6B
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
5/44
5VF TIMS Module
1 VF TIMS Module
The SSMTT-6B module is a VF TIMS (Voice Frequency - Trans-
mission Impairment Measurement Set) designed for testing,installation, and maintenance of 2 and 4 wire voice circuits. Itsupports the voiceband frequency range from 20 Hz to 20 kHz.Basic TIMS measurements include:
Level/frequency
Noise with tone measurement 3-tone slope or programmable frequency sweep generation
3-level impulse noise measurement
In addition to basic TIMS measurements, the module also sup-ports 2-wire signaling for loop-start, DID (Direct Inward Dial), andGround Start. In addition DID + GS has PBX/CO with DP, DTMF,
MF dialing, and talk/listen capabilities.
For more application details, see Section 3of this Users Man-ual.
This module can be used with the full family of MTT and xDSL
test sets, however there may be minor screen differences betweenthe color and b & w models. This Users Manual uses the color
screens for illustrative purposes.
Note: The standard Sunrise Telecom AC adaptor may generatelow frequency noise which affects the frequency testing results.For this reason we recommend that during VF TIMS measure-
ment the battery or the Sunrise Telecom AC adapter, SS138Ebe used.
1.1 LEDsThe module does not use any LEDs for reporting on circuit condi-tions. It does however indicate its use by the MODULE LED.
1.2 Module Connector Panel
The module connector panel is shown in the following gure:
HANDSET 2W/4W
Figure 1 Module Connector Panel
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
6/44
6 SSMTT-6B
The module has the following ports:
HANDSET
Connect a standard handset using an RJ-9 connector. It is used
for talk/listen functions.
2W/4W
The 2W/4W port is a 6-pin RJ-11 connector. The center two pins (3& 4) are used in 2-wire mode. In a 4-wire mode running in normal
operation, the center pins (3 & 4) are used for transmitting andpins 2 & 5 are used for receiving. Pin 6 is ground.
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
7/44
7VF TIMS Module
2 VF TIMS Main Menu
The module main menu is shown in the following gure.
VF TIMS MAIN MENU
CALL ANALYSIS 1, 2
LOOP CURRENT 12.6
RECEIVE FREQ & MEAS
2.4
TRANSMIT FREQUENCY
2.3
PLACE CALL 12.2
CONFIGURATION
2.1
NOISE & IMPAIRMENTS
2.5
OTHER SETUP 1
VIEW/STORE/PRINT
2.9
2.8
2.7
RECEIVE FREQUENCY & MEAS
3-TONE SLOPE
FREQUENCY SWEEP
SINGLE FREQUENCY
2.4.1
2.4.2
2.4.3
NOISE & IMPAIRMENTS
IMPULSE NOISE
NOISE/SNR
2.5.1
2.5.2
MODULEKey
1 not in 4-wire configuration2not in DID CO, DID PBX,
GS CO, or GS PBX signaling
Figure 2 Module Menu Tree
While in VF TIMS 2-wire menu, the test set displays the status of
the POTS interface and test set status. The status is shown, as
in Figure 3, below the time of day line after the carrot >.
11:50:45
> RINGING 3638000 600ohm CALLING 600ohm IDLE TX RX 600ohm 600ohm IDLE 600ohm IDLE 600ohm CALLING 3601916 600ohm RMT
DID-CO STATISTICS
PRE-WINK TIME : 1394 msWINK DURATION : 306 ms
WINK OFFHOOK DUR:
11:50:45
> CALLING 3601916 600ohm RMT PLACE CALL
SIGNALING : DID CO
DIAL TYPE : WINK-ST DIAL MODE : DTMF
DIAL NUMBER:
3601916
RECEIVED WINK FROM PBX
XMT DIGITS :
Figure 6 Placed Call Screens
In the screen on the left, XMT DIGITS displays any digits dialedafter the call has been placed. In this example, none have been
dialed. In both screens, another icon is present under CALLING.RMT indicates the status of the remote device, here it is on hook.
The following F-keys are available:
STAT(F1): This is available only when SIGNALLING is DID COand DIAL TYPE is WINK ST. The following is reported in the right
screen of Figure 6:
PRE WINK TIME: The time the call is initiated until the winkwas received in milli-seconds.
WINK DURATION: The length of the wink in milli-seconds.
WINK OFFHOOK DUR: Displays the time between the winkand the remote side went off hook. A gure is displayed only
when the RMT (remote) side goes offhook.
When nished, press ESC to return to the screen shown onthe left of Figure 6 and the rest of the F-keys:
FLASH (F3): Use to momentarily go on-hook/off-hook.
ONHOOK(F4): Use to end the call with the screen returning tothe state shown in Figure 5.
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
12/44
12 SSMTT-6B
2.2.1 Receive a Call
These screens are only displayed when a call is received.
11:50:45
RECEIVE CALL> RINGING 4083601259 600ohm RINGING 4083601259 600ohm IDLE 600ohm IDLE RX 600ohm< SNR MEASUREMENT
NOISE FILTER : C-MESG
RX FREQUENCY : 1004 HzRX LEVEL : -20.2 dBmNOISE : 10.3 dBrnSNR : 10.6 dBm
STORE
Press
ENTER
11:50:45
NOISE SNR
> IDLE 600ohm< NOISE MEASUREMENT
MEASURE TYPE :PINOISE FILTER : C-MESGNOTCH FILTER : DISABLE
(PRESS ENTER TO START)
PI
11:50:45
> IDLE RX600ohm< POWER INFLUENCE
NOISE FILTER : C-MESG
POWER INFLUENCE :
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
23/44
23VF TIMS Module
NOISE: Measure the background noise on a cable pair. To
perform this measurement, there must be a quiet terminationor open at the end of the cable pair.
SNR: Signal-to-Noise Ratio is a ratio of the quality of the signalwith noise compared to the noise. This measurement must beperformed with a test tone, typically 1004 Hz.
PI: Power Inuence: Noise to ground.
NOISE FILTER
Options: C-MESG (F1), 3.4K (F2), D (F3)
Determine the type of filter used for the measurement.
C-MESG: C-Message weighs frequencies between 600 Hzand 3 kHz most heavily. It, therefore, measures noise in the
frequency spectrum most noticeable to the human ear. Thismeasurement is called metallic noise.
3.4K: This weighs frequencies above 3.4 kHz most heavily.
D: Also known as the CCITT0.41 Channel Bank Filter, weightsnoise or impulse noise as it relates to VF data or fax trans-
mission. The 3 dB bandwidth extends from 300 Hz to 3,400Hz approximating the transmission spectrum of a 28.8/V.34
modem. This filter is composed of a 300 Hz high pass filtercascaded with a 3.4 kHz low-pass lter. Both the analog anddigital realizations are shown in the following gure:
100Hz 200 300 500 700 1K 2K 3K 4K 5K 7K 10KHz
0 dB
5
10
15
20
25
30
35
40
45 dB
Frequency
Loss
Digital
Analog
Figure 15 D Filter
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
24/44
24 SSMTT-6B
NOTCH FILTERThis line indicates if the filter is enabled or disabled. Its state is
determined by the MEASURE TYPE choice.
When ready, press ENTER. The following is reported:
RX FREQUENCY: Measured frequency in Hz. Only in NOISEand SNR measurements.
RX LEVEL: Measured in terms of dBm. A dBm unit measures thedecibels relative to 1 mW of power (across the selected imped-
ance). Only in NOISE and SNR measurements.
NOISE: Measured in terms of dBrn (decibels relative to noise). 0dBrn=-90 dBm. Only in NOISE and SNR measurements.
SNR: Signal-to-Noise Ratio is measured in terms of pure decibels(dB). Only in SNR measurements.
POWER INFLUENCE: Measured only in PI measurements. This is
a noise measurement taken with respect to ground to determinenoise from sources such as power lines.
The following F-key is available:
STORE(F4): Press to save the results. See Section 2.9.
When finished, press ESC.
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
25/44
25VF TIMS Module
2.5.2 Impulse Noise
This feature detects impulse noise spikes on the signal andkeeps a running count of the number of impulse events over time.
Impulse noise is dened as a random pulse whose amplitude issignificantly higher than that of background noise. IEEE defines
impulse noise as any burst of noise that produces a voltage ex-ceeding the rms value of the background or quantizing noise by
more than 12 dB [IEEE 743-1995].
Press
ENTER
11:50:45
CLEAR
> IDLE RX 600ohm< IMPULSE NOISE
ET- 000:01:32 HIGH: 40 dBrnRT- CONT MID : 35 dBrnFILTER: C-MESG LOW : 30 dBrnNOTCH : OFF
HIGH: 0
MID : 0
LOW : 0
11:50:45
C-MESG 3.4K
> IDLE 600ohm< IMPULSE NOISE SETUP
NOISE FILTER :C-MESGDEADTIME : 100 msTHRESHOLD(LOW) : 30 dBrnTHRESHOLD SPREAD : 5 dBTEST TIME : CONTNOTCH FILTER : OFF
D STORE
Figure 16 Impulse Noise Screens
Congure the following:
NOISE FILTER
Options: C-MESG (F1), 3.4K (F2), D (F3)
Determine the type of filter used for the measurement.
C-MESG: C-Message weighs frequencies between 600 Hzand 3 kHz most heavily. It, therefore, measures noise in the
frequency spectrum most noticeable to the human ear. 3.4K: This weighs frequencies above 3.4 kHz most heavily.
D: Also known as the CCITTO.41 Channel Bank Filter,weights Noise or Impulse Noise as it relates to VF data or faxtransmission. The 3 dB bandwidth extends from 300 Hz to
3,400 Hz so as to approximate the transmission spectrum ofa 28.8/V.34 modem. This filter is composed of a 300 Hz high
pass lter cascaded with a 3.4 kHz low-pass lter. Both theanalog and digital realizations are shown in Figure 15.
DEAD TIMERange: 10 ms to 255 ms
This is the measurement delay after the test set detects the initial
impulse. Dead Time begins as soon as the test set detects theinitial impulse. It resumes measuring events after the dead timehas elapsed. This prevents the test set from measuring the same
impulse noise spike multiple times.
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
26/44
26 SSMTT-6B
THRESHOLD (LOW)Range: 30 dBrn to 90 dBrn
Dene the lower threshold value for an impulse noise event. When
the test set detects noise above this threshold, the test set recordsit as an impulse noise event.
THRESHOLD SPREADRange: 2 to 6 dB
Set the difference between the Low, Middle, and High event
counts. In the top left screen in Figure 16, THRESHOLD (LOW)is set for 30 dBrn; this is the low count. The spread value is setfor 5. Therefore, the low count refers to any noise between 30 and
35; the middle count refers to any noise between 35 and 40. Highrefers to anything above 40 dBrn.
TEST TIME
CONT (Continuous) is the only option.NOTCH FILTEROptions: ON (F1), OFF (F2)
Choose whether to use this lter during the measurement.
When ready, press ENTER to start and refer to the right screenshown in Figure 16.
ET: Elapsed Time since ENTER was pressed.
RT: Remaining Time displays CONT, since this is the only option
for the TEST TIME setting.
The noise for HIGH, MID, and LOW is reported in dBrn. The fol-lowing counters provide the impulse noise results:
HIGH: Number of impulse noise spikes detected whose level falls
within the high value (above HIGH).
MID: Number of impulse noise spikes detected whose level fallswithin the middle value (in between MID and HIGH).
LOW: Number of impulse noise spikes detected whose level fallswithin the low value (in between LOW and MID).
The following F-keys are available:
CLEAR (F1): Resets all counters to zero. The test continues
running.
STORE(F4): Press to save the results. See Section 2.9.
When finished, press ESC.
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
27/44
27VF TIMS Module
2.6 Loop Current and Voltage
11:50:45
> IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE RX600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm IDLE 600ohm< VIEW/STORE/PRINT Free space: 21475 kbyte NAME TYPE LOCK
1. 302 ENZO DR DSLAM2. 333W57ST DSLAM3. MULBERRY-3/11 DSLAM4. TARRYT-3/12 DSLAM5.6.7.8.9.
10.VIEW PRINT more
Figure 26 View/Store/Print Screen
The following F-keys are available.
VIEW (F1): Allows viewing of a selected le, see Section 2.9.2.
PRINT (F3): Allows printing of a selected le, see Section 2.9.3.
RENAME(more, F1): Allows renaming a selected le, see Sec-
tion 2.9.6.
UN/LOCK (more, F2): Allows locking and unlocking a le, see
Section 2.9.5.
DELETE (more, F3): Allows deleting a le, unless locked, seeSection 2.9.4.
2.9.1 Saving a Test
1. From any screen with a SAVE F-key, press it and the informa-tion is saved with a generic lename.
2.9.2 Viewing a Stored Test
1. From the module main menu, select VIEW/STORE/PRINT.2. Select the desired file using .3. Press VIEW (F1) and the les screen(s) will be displayed.
2.9.3 Printing a Stored Test
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
37/44
37VF TIMS Module
1. Connect a SunSet printer to the serial port of the test set.
For other types of printers or for more information, refer to the
Storing and Printing chapter in the test set Users Manual.
2. From the module main menu, select VIEW/STORE/PRINT.3. Select the desired file using .4. Press PRINT (F3) and the le will begin printing.
5. When finished, press ESC to return to VIEW/STORE/PRINT.
2.9.4 Deleting a Stored Test
1. From the module main menu, select VIEW/STORE/PRINT.
2. Select the desired file using .3. Press DELETE (more, F3) and the file is deleted if the file is
unlocked.
2.9.5 Locking & Unlocking a Stored Test
1. From the module main menu, select VIEW/STORE/PRINT.2. Select the desired file using .
3. Press UN/LOCK (more, F2) and the file is locked or unlockedas indicated to the right of the file name. Refer to the lock icon
shown in Figure 26.
2.9.6 Renaming a Stored Test
1. From the module main menu, select VIEW/STORE/PRINT.2. Press to select the desired file.
3. Press RENAME (F1) to display a character entry screen.4. Press INPUT (F3). Note that the A character is highlighted
and the INPUT F-key has changed to STOP.5. Press to move the cursor to a character and pressENTER to place that character in the FILENAME line. Repeat
until finished. Enter up to 16 characters. If a mistake is made:
A. Press STOP (F3) and move the FILENAME cursor to theincorrect character.
B. Press DELETE (F2) to delete the character or press INSERT(F1) to insert a character.
C. Press INPUT (F3) to select a character. Press ENTER toinsert the new character to the left of the cursor.
6. Press SAVE (F4) to save and return to the VIEW/STORE/PRINT screen shown in Figure 26.
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
38/44
38 SSMTT-6B
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
39/44
39VF TIMS Module
3 Applications
Responder
or RemoteTest Server
PSTN
2W/4W
Receive
Calls
OR
OR
2W/4W
Figure 27 Place/Receive Calls and Send Digits
The VF TIMS module is capable of the following functions:
Place calls to normal POTS phones, another test set, or a
Responder/Remote Test Server. Receive a call (detect ringing voltage).
Send digits, which allows you to interact with systems likevoice mail as well as Responders and Remote Test Servers
for setting up tests. Act as a responder device. Transmit and receive tones simultaneously while performing
measurements.
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
40/44
40 SSMTT-6B
3.1 Express Limited Warranty
This Sunrise Telecom product is warranted against defects inmaterials and workmanship during its warranty period. The war-
ranty period for this product is contained in the warranty page onhttp://www.sunrisetelecom.com.
Sunrise Telecom agrees to repair or replace any assembly or
component found to be defective under normal use during thisperiod. The obligation under this warranty is limited solely to re-pairing or replacing the product that proves to be defective within
the scope of the warranty when returned to the factory. This war-ranty does not apply under certain conditions, as set forth on the
warranty page on http://www.sunrisetelecom.com.
Please refer to the website for specic details.THIS IS A LIMITED WARRANTY AND THE ONLY WARRANTY
MADE BY SUNRISE TELECOM. SUNRISE TELECOM MAKES
NO OTHER WARRANTY, REPR SENTATION OR CONDITION,EXPRESS OR IMPLIED, AND EXPRESSLY DISCLAIMS THE
IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESSFOR A PARTICULAR PURPOSE AND NON-INFRINGEMENT
OF THIRD PARTY RIGHTS.
SUNRISE TELECOM INCORPORATED302 Enzo Drive San Jose, CA 95138 U.S.A.Tel: 1-800-701-5208 Fax: 1-408-363-8313
Internet: http://www.sunrisetelecom.comE-mail: [email protected]
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
41/44
41VF TIMS Module
Index
Symbols3-Tone Slope Screens
ATTENUATION; 21ATTEN REF LEVEL; 20
SIGNAL FREQ; 21SIGNAL LEVEL; 21
CCall Analysis Setup Screen
CALL TYPEDTMF, MF, or DP; 28
Configuration Screen
AUDIO DEVICEHANDSET or MIC/SPK; 8
IMPEDANCE; 8
MODE2 and 4 wire; 8
F
Figures01 Module Connector Panel; 502 Module Menu Tree; 7
03 Status Indicators; 704 Configuration Screen; 8
05 Place Call Screen; 906 Placed Call Screens; 11
07 Receive Call Screens; 1208 Transmit Frequency-Fixed Mode Screen; 1309 Transmit Frequency-Sweep Mode Screen; 14
10 Transmit Frequency-3-Tone Mode Screen; 1611 Single Frequency Screen; 17
12 Frequency Sweep Screens; 1813 3-Tone Slope Screens; 20
14 Noise/SNR Screens; 2215 D Filter; 2316 Impulse Noise Screens; 25
17 Loop Current/Voltage Screen; 2718 Call Analysis Setup Screen; 28
19 Call Analysis Test Set Connection; 2820 DTMF Call Analysis Screens; 29
21 Other Setup, Loop-Start Screens; 3022 Other Setup, DID CO-WINK-ST Screens; 3223 Other Setup, DID CO-IMMD-ST Screens; 33
24 Other Setup, DID PBX-WINK ST Screens; 3425 Other Setup, DID PBX-IMMD-ST Screens; 35
26 View/Store/Print Screen; 3627 Place/Receive Calls and Send Digits; 39
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
42/44
42 SSMTT-6B
Frequency Sweep Screens
ATTENUATION; 19ATTEN REF LEVEL; 18
SIGNAL FREQ; 19SIGNAL LEVEL; 19
I
Impulse Noise ScreensDEAD TIME; 25
ET; 26HIGH; 26
LOW; 26MID; 26NOISE FILTER
C-MESG, 3.4K, or D; 25NOTCH FILTER; 26
RT; 26
TEST TIME; 26THRESHOLD (LOW); 26THRESHOLD SPREAD; 26
LLoop Current/Voltage Screen; 27
MModule Connector Panel
2W/4W; 6HANDSET; 6
Module Menu Tree; 7
N
Noise/SNR ScreensMEASURE TYPE
NOISE, SNR, or P.I.; 22
NOISE; 24NOISE FILTER
C-MESG, 3.4K, or D; 23NOTCH FILTER; 24POWER INFLUENCE; 24
RX FREQUENCY; 24RX LEVEL; 24
SNR; 24
O
Other Setup ScreensDELAY PERIOD; 35
DIALING TYPE; 35DIAL MODE
DTMF, MF, or DP; 31
DIGIT OFF TIME; 31
8/14/2019 MAN-22448-001_B00_SSMTT-06B VF-TIMS_MMD
43/44
43VF TIMS Module
DIGIT ON TIME; 31
DP & BREAK; 31DP PPS; 31
FLASH DURATION; 30HI TONE FREQ; 31INTERDIGIT PRD; 31
LOW FREQ LEVEL; 31LO TONE FREQ; 31
SIGNALING; 30TWIST (HI - LO); 31WINK DURATION; 35
PPlaced Call Screens
PRE WINK TIME; 11WINK DURATION; 11
WINK OFFHOOK DUR; 11
Place Call ScreenDIAL MODE
DTMF, MF, or DP; 10
DIAL NUMBER; 10DIAL TYPE
WINK-ST or IMMD-ST; 9
SIGNALINGSee entry for selections; 9
SPEED DIAL; 10
R
Receive Call Screens; 12
S
Single Frequency ScreenATTENUATION; 17
ATTEN REF LEVEL; 17SIGNAL FREQ; 17SIGNAL LEVEL; 17
Status Indicators; 7
T
Transmit Frequency ScreenMODE; FIXED, SWEEP, 3-TONE, QUIET; 13
Transmit Frequency ScreensDWELL TIME; 15
FREQUENCY; 14
FREQ STEP; 15LEVEL; 14,15
SKIP FREQ; 15STOP FREQ; 15