TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY
description
Transcript of TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY
![Page 1: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/1.jpg)
TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO
MEDIANTE IL COUNTING TEST FACILITY
Davide Franco
13/10/2003
![Page 2: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/2.jpg)
Outline• Introduzione:
• Descrizione dell’apparato Borexino e del prototipo in piccola scala (CTF)
• Analisi dati CTF3:
• Determinazione della scala energetica
• Radon, 226Ra (238U), 85Kr
• Analisi energetica
• Analisi radiale
• Purificazione dell’azoto da 39Ar e 85Kr
• Misure di purezza di campioni di azoto
• Misure sull’efficienze di purificazione di carboni attivi
• Misura di 11C prodotto da muoni cosmici
![Page 3: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/3.jpg)
Borexino è un esperimento progettato per la misura diretta del 7Be mediante la reazione:
+ e + eRivelazione attraverso spalla Compton a 667 keV
Paradosso 7Be/8B
Cosa possiamo imparare da una misura diretta del 7Be econ una precisionedel 10%
Bahcall – Pena-Garay hep-ph/0305159:
• Testare modelli solari
• Vincolo sul flusso totale
• Diversi regimi di oscillazione sotto 1 MeV?
![Page 4: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/4.jpg)
L’Esperimento Borexino
Laboratori Nazionali del Gran Sasso (profondità di 3800 mwe)
Il rivelatore e’ strutturato in shell
Caratteristiche del rivelatore (dal centro):• Scintillatore: PC + PPO
(300 ton, 100 ton di massa fiduciale)
• Sfera in nylon (d = 8.5 m)
• Liquido di buffer: PC + DMP
(1040 ton)
• 2200 fototubi
• Sfera in acciaio (d = 13.7 m)
• Buffer esterno di acqua ultrapura
• Serbatoio d’acciaio (h e dbase = 18 m )
![Page 5: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/5.jpg)
Richieste di radiopurezza
![Page 6: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/6.jpg)
Radiopurezza• Scelta dei materiali
• Struttura concentrica
• Procedure di installazione e montaggio
• Water Extraction • rimozione di impurezze metalliche (U, Th e K)
• SilicaGel • rimozione di impurezze ioniche
• Distillazione• rimozione di impurezze chimiche
• Stripping • rimozione di gas disciolti(< 5 ppm)
• Filtraggio• rimozione di particolato > 0.05 μ
Sistemi di purificazione “on-line”
![Page 7: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/7.jpg)
Counting Test Facilities
Prototipo su piccola scala:
• 5 ton di scintillatore
• 100 PMT
• buffer di acqua
• 238U = (3.5 ±1.3) 10-16 g/g
• 232Th = 4.4 10-16 g/g
• 14C/12C = (1.94 ± 0.09) 10-18
1.51.2
![Page 8: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/8.jpg)
Fasi di purificazione e test in CTF-3
Test del 14C
Water ExtractionColumn (Silica Gel) Shroud Adj.
PMT Off
![Page 9: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/9.jpg)
Informazioni dal rivelatore
• carica (adc) trasformazione in energia
• problemi di non linearità in scala: quenching α, β e “γ”
• tempo (tdc) misura del profilo temporale (α/β )
ricostruzione della posizione
• problema della riflessione totale
• coincidenze ritardate identificazione di segmenti di catena
• acquisizione su una seconda parallela catena elettronica (gruppo 2)
Il segnale dei PMT, processato dalla catena elettronica, fornisce:
![Page 10: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/10.jpg)
Analisi dati
• Fit dello spettro energetico identificazione dei singoli contributi
• Fit dello spettro radiale identificazione dei contributi “interni”, “superficiali” ed eventualmente “esterni”
• Analisi del profilo temporale discriminazione della tipologia α o β del contributo
• Decadimenti veloci identificazione del fondo
![Page 11: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/11.jpg)
Scala energetica: light yield attraverso il 14C
(( )) ()e eNC E g E W W dW
N(We)dWe= peWe(W0-We)2F(Z,We)C(We)dWe
• pe , We = momento ed energia dell’ e emesso• W0 = endpoint• C(We) = fattore di forma• F(Z,We) = funzione di Fermi
Forma teorica:
N e g dipendono dal numero di p.e./MeV !
Risoluzione del rivelatore:
2
2
(x)
2
ge
x2
( )
![Page 12: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/12.jpg)
Scala energetica: light yield attraverso il 14C
f (E) = funzione approssimata di quenching da studi MC
C’(E) = C( f (E) )
Light Yield
parametro libero del fit
![Page 13: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/13.jpg)
Scala energetica: andamento temporale del light yield
Tutte le analisi qui presentate sono normalizzate per il light yield e per 100 PMT
![Page 14: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/14.jpg)
Possibili contributi
Contributi studiati in CTF-3:
• 222Rn - 218Po - 214Pb - 214Bi - 214Po
• 210Pb - 210Po - 210Bi
• 85Kr
• 40K
• 39Ar
![Page 15: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/15.jpg)
Determinazione del rate di 238U dal 222Rn
Per determinare il rate dell’238U
coincidenze 214Bi-214Po (~ 200 s):
222Rn in equilibrio secolare con la catena dell’238U decade con la vita media del 238U (6.49 109 y)
si osserva un plateau
Complicazione:
222Rn introdotto nello scintillatore (in seguito a manipolazioni) decade con la vita media del 222Rn (5.48 d)
![Page 16: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/16.jpg)
Radon dalla coincidenza 214Bi-214Po
A oggi, 1.1 c/d.Assumendo l’equilibrio secolare238U = (2.6 ± 0.2)x10-16 g/g
![Page 17: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/17.jpg)
Identificazione del 85Kr attraverso le coincidenze ritardate
Decadimento del 85Kr:
• β- 99.563 % BR 85Rb (Emax = 687.4 KeV)
• EC 0.434 % BR 85mRb (Emax = 173.4
KeV)
Il 85mRb decade nello stato fondamentale con vita media 1.46 µs emettendo γ a 514 keV
DC:• sequenza di decadimenti del 85Kr (via EC) e del 85mRb nella finestra temporale
• Svantaggi: il BR del decadimento via EC e’ molto piccolo(si richiede alta statistica)
Analisi energetica:• spettro β del decadimento del 85Kr (via β- )
• Svantaggi: lo spettro del 85Kr si confonde con lo spettro del 39Ar
![Page 18: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/18.jpg)
Data Set Days 85Kr-85mRb (num of events r<2m)
85Kr activity
(c/d/mIV)
85Kr-85mRb
(num of events r<0.8m)
85Kr activity
(c/d/miv)
1: 2056-2073 24.1 20 ± 4 409 ± 82 10 ± 3 358 ± 108
2: 2074-2093 21.1 19 ± 4 443 ± 84 7 ± 3 287 ± 123
3: 2094-2122 33.0 38 ± 6 510 ± 81 19 ± 4 498 ± 105
4: 2125-2149 26.3 7 ± 3 118 ± 50 0 < 91 (90%)
5: 2154-2161 12.5 3 ± 2 106 ± 71 0 <191 (90%)
6: 2162-2183 24.1 5 ± 2 92 ± 71 0 <99 (90%)
7: 2188-2281 94.6 31 ±6 145 ± 28 2 ± 1 18 ±9
8: 2282-2345 99.1 22 ± 5 98 ± 22 7 ± 3 61 ±26
9: 2346-2434 165.8 36 ± 6 96 ± 16 9 ± 3 47 ±16
![Page 19: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/19.jpg)
![Page 20: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/20.jpg)
Analisi Radiale: formalizzazione
ft cr
n
tf = tempo di volo del fotone
c = velocità della luce
n = indice di rifrazione
• Contributo interno:
Convoluzione di r2 con la risoluzione del rivelatore (distribuzione di Rayleigh)
parametri liberi:
• Posizione del vessel (*)
• Risoluzione efficace
• Rate
• Contributo superficiale:
Distribuzione di Rayleigh
parametri liberi:
• Posizione del vessel (*)
• Risoluzione
• Rate
• Contributo esterno:
Trascurato
![Page 21: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/21.jpg)
Problema con la sottrazione del radon
![Page 22: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/22.jpg)
Parametrizzazione dello spettro energetico
Ogni contributo e’ trattato indipendentemente come convoluzione della forma spettrale con la risoluzione del rivelatore
α + γ
Eα-quenched + Eγ
β + γ
Eβ + EγEγ210Po
Radon
210Bi
85Kr
Total
![Page 23: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/23.jpg)
Fit dello spettro energetico
![Page 24: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/24.jpg)
I Batch 14C Test I SG
I WE BL II WE
II SG 8 Months 16 Months
![Page 25: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/25.jpg)
Equilibrio BiPo-210 raggiunto
Sistematica dovuta all’incertezza sul raggiodel vessel:5 cm d’incertezza 16% di sitematica
![Page 26: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/26.jpg)
• Sono state utilizzate tutte le informazioni disponibili dai dati di CTF
• Studio sulla scala energetica e sul quenching (work in progress)
• Test della distillazione• Finalità dell’analisi:
– valutazione dei contributi – stabilità del rivelatore
• I metodi di analisi utilizzati verranno direttamente impiegati in Borexino
Riassumendo:
![Page 27: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/27.jpg)
Purificazione dell’azoto da 39Ar e 85Kr
• 39Ar: T1/2 = 269 y; - (end-point 565 keV); 1.4 Bq/m3 Ar
• 85Kr: T1/2 = 10.8 y; , - (end-point 687 keV); 1 MBq/m3 Kr
• Requisiti per Borexino:
~100 nBq/m3 (1 ev/day nel volume fiduciale) per 39Ar e 85Kr
• Solubilità:
Ar: N2/PC = 4.4 (SDS) (N2/H2O = 37)
Kr: N2/PC = 1.4 (H. Simgen) (N2/H2O = 21)
• => requisiti in N2:
0.4 Bq 39Ar/m3 N2 0.31 ppm Ar in N2
0.14 Bq 85Kr/m3 N2 0.14 ppt Kr in N2
Obiettivi:
• individuare N2 prodotto direttamente a bassa contaminazione
• studiare il sistema di purificazione più efficiente
![Page 28: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/28.jpg)
Misura dei campioni di azoto
Spettrometro di massa
gas
liquid
Pipes backed out and flushed with nitrogen for some days
Pipettes
Dewar (180 L) con azoto liquido
N2
6.0
Samplepurification
Limite del rivelatore (per 1 cm3 di N2): 1 ppb per Ar e 0.1 ppt for Kr
![Page 29: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/29.jpg)
Adsorbimento con zeoliti e carboni attivi (modello di Langmuir)
legge di Henry: nads = H p
![Page 30: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/30.jpg)
Purificazione dell’azoto da Kr e Ar per adsorbimento con carboni attivi
gas
liquid
600-L dewar with Kr-enriched (100 - 400 ppt)
liquid nitrogen (Westfalen AG)
Mass spectr.
N2
6.0
Samplepurification
LN2, LArCryocool
100/300-cm3 column filled with adsorber
bubbler
VRet H ()VP
C0
CN
![Page 31: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/31.jpg)
Cromatografia: L=N x H.E.T.P.
![Page 32: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/32.jpg)
Purificazione da 85Kr con diversi assorbitori
Zeoliti non adattiall’adsorbimento del 85Kr
LN2 Temperature
LAr Temperature
12 kg di Carbosieve III purificano 2000 m3 di azoto
corrispondente al fabbisogno di Borexino per 4 giorni
![Page 33: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/33.jpg)
LN2 temperature
LAr temperature
![Page 34: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/34.jpg)
Risultati: concentrazione Ar/Kr per diversi campioni di azoto
Description Ar [ppm] Kr [ppt]
HD RPN2 (4.0) 200 1 680
GS RPN2 (4.0) 10 40
GS HPN2 10 30
Linde Worms (7.0) 0.05 7
SOL Mantova (6.0)
0.2
0.005
8
0.04
Westfalen AG Hörstel (6.0+ Kr)
Westfalen AG Hörstel (6.0+ Kr)
Westfalen AG Hörstel (6.0+ Kr)
Westfalen AG Hörstel (6.0)
Westfalen AG Hörstel (6.0)
0.008
0.004
0.002
0.005
0.0005
170
250
430
5
0.06
Goal 0.31 0.14
![Page 35: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/35.jpg)
The 11C problem in the CNO-pep NW
Expected -rate in Borexino in 100 ton (BP2004 + LUNA + LMA) in the energy range [0.8 – 1.4] MeV:
• pep-: 0.9 c/d
• CNO-: 0.6 c/d
Internal background in [0.8 – 1.4] MeV: 0.6 ev/day @ 10-17 g/g U, ThIn situ production muon-induced 11C:
• 7.5 c/d in the range [0.8 – 1.4] MeV
required reduction factor > 10
Goal: tagging and removing 11C event by event!!!
![Page 36: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/36.jpg)
-induced 11C: the net reaction
1112C C nm+ ® +
01 11eCC e n+® + +
~ 30 min!!!
E [1.02 – 1.98] MeV{
n p d g+ ® + ~ 200 s
E = 2.2 MeV{<E> = 320 4stat. + 11 syst. GeV @ LNGS (3600 mwe)
n
11C (+)
![Page 37: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/37.jpg)
Isotopes c/d(*) in the full energy range in CTF
11C 0.54 0.067Be (1.3 0.1) x 10-2
11Be < 10-3
10C (7.3 0.8) x 10-2
8Li (2.5 0.6) x 10-3
6He (3.7 0.1) x 10-2
8B (4.0 0.7) x 10-3
9C (2.8 0.9) x 10-3
9Li + 8He (1.3 0.2) x 10-3
(*) T. Hagner et al., Astr. Part. Phys. 14 (2000) 33-47
Muon-induced contaminants in CTF
Scaling to CTF, 11C represents 80% of all the muon-induced contaminants and more than 99% in the NW-2
Experiment @ CERN(*): cross sections measured with -beams of 100 and 190 GeV on 12C target
![Page 38: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/38.jpg)
Strategy in CTF
1. Tagging cosmic muons with the muon veto:
<320 GeV>, crossing the scintillator, fires the detector
• cut on the number of photoelectrons detected by the muon-veto ( ~ 1)
2. Tagging 2.2MeV from the neutron capture:
• for each tagged <320 GeV>, a temporal gate of [1 s - 2ms] is opened
• cut: E > 1.8 MeV
• and neutron (spill-in/spill-out effect) have been evaluated via Monte
Carlo
3. Collecting data sample including 11C events:
• after each 2.2MeV a temporal gate of 300 min (10 x ) in the [1.10 – 1.65]
MeV energy window (energy = 0.82 )
• Radial cut: r has been evaluated via Monte Carlo
n
11C (+)
![Page 39: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/39.jpg)
rate ~ 1.7 c/d in 3.7 tons
Eve
nts
/320
day
s/2.
7 to
ns/
45 k
eV
Energy(MeV)
Neutron capturen + p d + 2.2 MeV
capture time ~ 200 s
2.2MeV peak
= 211 14 s
CoincidenceTime (102 ns)
Eve
nts
/320
day
s/3.
7 to
ns/
80 s
Capture time
n
11C (+)
![Page 40: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/40.jpg)
Coincidence Time (min)
Detection of in situ produced 11C rate
/( ) tNF t e Bt
t-= +
• Effective data taking: 320 days
• Radial cuts arount the center of the detector (r [0.7 – 0.8] m) for avoiding optical effects at different medium interface
• Assuming a constant background B
Fit function:
From the fit: = 29 ± 13 min
n
11C (+)
(others analyses provide a lower uncertainty on )
![Page 41: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/41.jpg)
• Fixing = 29.4 min
• r = 0.7 m (mr = 1.4 ton)
• T = 320 days
• Energy window [1.1 – 1.65] MeV: en = 0.82
(0.17 0.04)( 0.7 ) /
energy r n r n
NR r m c d
T g ge e e e e e e±
< = =× × × × × ×
Result: 11C rate in CTF
Note: less than 1 event every 5 days
n
11C (+)
![Page 42: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/42.jpg)
FLUKA has been already tested succesfully in muon-induced neutron production (LVD, etc)
Simulation strategy:
• FLUKA:
• detector geometry
• vertical muon beam spread over the whole detector surface
• neutron tracking in scintillator and water
• CTF tracking code:
• tracking of 2.2MeV starting from the neutron capture position
• CTF reconstruction code:
• reconstruction of the center of mass of the 2.2MeV deposited energy
Efficiencies: Monte Carlo Simulation
![Page 43: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/43.jpg)
Results from the MC (1)
Events in water and in scintillator
Neutron capture in scintillator and water
204 ± 2 s
Good agreement with real data (211 ± 14 s)!
0 4 8 12Coincidence Time (100 s)
Num
ber
of e
vent
s
n
11C (+)
Angular Distribution(cos)
Energy Distribution(GeV)
![Page 44: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/44.jpg)
Results from the MC (2)
Distance from neutron production position to (normalized spectra):
• neutron capture
• 2.2MeV center of mass
Only internal events
Distance (m)
82% (87%) of neutrons are contained in 1 m radius sphere with center in 2.2MeV center of mass (neutron capture position)
Cumulative probability:
Distance (m)
%
n
11C (+)
![Page 45: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/45.jpg)
Efficiencies and final results
(0.17 0.04)( 0.7 ) /
r n
R r m c dge e e
±< =
× ×
Including the systematics due to:
• scintillator mass
• light yield: energy scale conversion
0.060.040.60 0.14( .) ( .) /measuredR stat syst c d+
-= ±
Rexpected = 0.54 ± 0.06 c/d
0.29r n ge e e× × =
From MC:
![Page 46: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/46.jpg)
Implication in Borexino
• Software cut– spherical volume
around the reconstructed 2.2 MeV
• Muon veto cut– cylindrical volume
around the muon track
How to reduce 11C background:
11C is removed blinding the intersection of the two volumes for 5-10 11C-lifetime
Main challenge: keeping low the total dead volume x time
Spherical cut aroundneutron capture
Cylindrical cut around-track
Neutron production
ReconstructedMeV
-track
Scintillator
![Page 47: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/47.jpg)
Summarizing
Distance (m)
Assuming:
• -signal (pep+CNO) rate: 1.5 c/d
• trace contaminants rate 0.6 c/d (238U and 232Th ~ 10-17 g/g )
• distribution of the distance covered by neutrons from FLUKA simulation
• expected 46 neutrons/day
• 100 tons of fiducial mass
• only the spherical cut
• NW-2: 0.8 – 1.4 MeV
![Page 48: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/48.jpg)
Signal-to-backround ratio ~ 1 implies a loss of volume-time detector fraction ~ 5%
Dead Volume-Time Fraction
Signal/Background Ratio
5% 1
20% 1.2
50% 1.5
Preliminary results
![Page 49: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/49.jpg)
Conclusions and perspectives
• Good agreement 11C rate measured/expected in CTF• 11C is tagged event by event• Loss of volume-time detector fraction is minimized
in Borexino • Improvements: cylindrical cut, neutron distribution
from 12C, etc.
=> Borexino has the potential to probe pep and CNO ns
![Page 50: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/50.jpg)
“Profilo” della purificazione
C0
CN
VPVRet H ()
Using „standard“ purity of LN2 from Linde/Mantova/Westfalen we need:
- Kr reduction factor 50- Continous purification for 1 week with 20 m3/h
We assume:
- Purification at LN2 temp. (77 K)- 2 kg of adsorber
N 10HT 77 [mol K/(kg Pa)]
CN =
½ C
0
ads
t
RTm
VH Re
![Page 51: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/51.jpg)
Results: H and N for different adsorbers
Description N
H T
[mol K/Pa/kg]
Goal 10 77
MS 5Å, Si/Al 1 no purification observed
MS 10Å, Si/Al 1 1 > 1.5 10-2
Zeolite, BEA-type, 5.3Å, Si/Al 200 1 3.1 10-1
Zeolite, MFI-Type, 6.6Å, Si/Al 75 1 2.3 10-1
Activated Charcoal, CarboAct 1 2.3
Activated Charcoal, Merck 1 2.3
Activated Charcoal, type C38/2 3 1.5
Carbon Cloth 10 2.3
Carbon Molecular Sieve SIII 1 10.8
Skeleton nanostructured carbon (7 Å) 2 11.6
Ad
sorp
tion
fro
m li
qu
idat
–19
6 o C
Ad
sorp
tion
fro
m g
as
Carbon Molecular Sieve SIII (-145 oC) 8 7.7
Carbon Molecular Sieve SIII – predict. for -190 oC 8 130
![Page 52: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/52.jpg)
CTF-3
ShroudVessel
Rn in acqua ~ 5 mBq/m3
![Page 53: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/53.jpg)
Andamento temporale dei contributi α di bassa energia nell’ultimo periodo di presa dati
Error bar: componente α risultante dal fit
• nella regione energetica del 210Po• nel volume fiduciale < 0.7 m• dopo la sottrazione della componente α del segmento del Rn
![Page 54: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/54.jpg)
Discriminazione α/β:rapporto “coda su totale”
Tempo d’integrazione totale: 500 ns
ns
![Page 55: TEST DEI SISTEMI DI PURIFICAZIONE DI BOREXINO MEDIANTE IL COUNTING TEST FACILITY](https://reader035.fdocument.pub/reader035/viewer/2022062222/56815193550346895dbfc8ff/html5/thumbnails/55.jpg)
Spettri energetici
Gruppo 1 + Gruppo 2
p.e.
Gruppo 1
Gruppo 2
p.e.
p.e.
Taglio di volume fiduciale (< 0.6 m)Statistica ~ 1 d