0624 안양대 수시모집요강 · 2019-06-25 · 등 중·고교와 대학 연계사업을확대하고, 전공체험·대입전형안내 등 학부모와학생들에게실질적도움이되는
LOGO Medical Instrument Dynamic Characteristics & Biopotential 김제선 김준범 김현준...
-
Upload
shawn-shepherd -
Category
Documents
-
view
215 -
download
2
Transcript of LOGO Medical Instrument Dynamic Characteristics & Biopotential 김제선 김준범 김현준...
LOGO
Medical InstrumentDynamic Characteristics & Biopotential
김제선김준범김현준김한울 (1 등 )노재선
Group 3 :
Medical Instrumentation
Contents
Transient response 1
Steady state response2
Distortionless Mearsurment3
Biopotential4
Medical Instrumentation
Dynamic Characteristic
Time dependency
Most medical instruments must process signals that are functions of time. It is this time-varying property of medical signals that requires us to consider dynamic instrument characteristics.
Medical Instrumentation
Transient response
x(t)
y(t)
H
CASE1
Step Response
The transfer function for a linear instrument or system expresses the relationship Between the input signal and the output signal mathematically.
Medical Instrumentation
Transient response
System behavior Dependence of the system behavior on the value of the damping ratio ζ, for under-damped, critically-damped ,over-damped, and undamped cases, for zero-velocity initial condition. The behavior of the system depends on the relative values of the two fundamental parameters, the natural requency ω0 and the damping ratio ζ. I
n particular, the qualitative behavior of the system depends crucially on whether the quadratic equation for γ has one real solution, two real solutions, or two complex conjugate solutions.
Medical Instrumentation
Dynamic Characteristic
Critical damping (ζ = 1)
When ζ = 1, there is a double root γ (defined above), which is real. The system is said to be critically damped. A critically damped system converges to zero faster than any other, and without oscillating.
Medical Instrumentation
Transient response
Over-damping (ζ > 1)
When ζ > 1, the system is over-damped and there are two different real roots. An over-damped door-closer will take longer to close than a critically damped door would.
Medical Instrumentation
Transient response
Under-damping (0 ≤ ζ < 1)
Finally, when 0 ≤ ζ < 1, γ is complex, and the system is under-damped. In this situation, the system will oscillate at the natural damped frequency ωd, which is a function of the natural frequency and the damping ratio.
Medical Instrumentation
Transient response
Under-DampingUnder-Damping
Critical-DampingCritical-Damping
Over-DampingOver-Damping
ζ = 1
ζ > 1
ζ < 1
Medical Instrumentation
Dynamic Characteristic
First-order system
ExponentialTime constant
System
Second-order system
Natural frequencyUnder-dampingCritical-dampingOver-damping
Medical Instrumentation
Dynamic Characteristic
3 차x y
x y1 차 2 차
*인수분해에 의해서 3 차는 1 차 , 2 차로 표현 가능
Medical Instrumentation
Steady state response
System
CASE2
Sinusoidal Steady State Frequency Response
Medical Instrumentation
Steady state response
Linear system(Principal of superposition)
LinearSystem
Linear combination
Basis of X =
Set of all x(t) is X,
x(t) is a linear combination-dependent frequency
Medical Instrumentation
Steady state response
Fourier Transform
Impulse response
Convolution - LTI(Linear time invariant)
Frequency transfer function
Amplitude response
Phase response
Medical Instrumentation
Steady state response
Review
Euler’s law
Medical Instrumentation
Steady state response
Example
Medical Instrumentation
Steady state response
H(jw)
각각의 주파수에 대한 출력을 알고 있으므로 입력의 합에 대한 결과 역시 알 수 있다 .(Superposition)
|H|
ω
ω
Medical Instrumentation
Steady state response
주파수는 같고 크기와 위상만 달라짐
Input
output
[ 주파수에 따라 출력의 모양이 달라진다 . ]
Medical Instrumentation
Distortionless Measurment
System
Time delay
Instrument elements that give an output that is exactly the same as the input, Except that is delayed in time by , are defined as time-delay elements.
Medical Instrumentation
Distortionless Measurment
transposition
|H|
ω
ωA
Flat amplitude response
Linear phase response
Medical Instrumentation
Distortionless Measurment
H*(jw)
Example
X(t) Y(t)
“Phase = Frequency X Time”
Medical Instrumentation
Example of distortion(amplitude)
X(t)
+
Medical Instrumentation
Example of distortion(phase)
Medical Instrumentation
Example of distortion입력의 주파수 범위가 w1 에서 w2 일때 무왜곡을 측정하기 위한 H(jw)?
ω
ω
Non-causal
존재할 수 없다빛의 속도보다 빠르면 존재
ω1 ω2
ω1 ω2
Medical Instrumentation
Biopotential
Insulating Membrane
Insulating membrane
Na+
Cl-
Na+
Cl-
Neutral(1%) Neutral(10%)
V
Voltage is zero.
*Half-cell potential is zero.
Medical Instrumentation
Biopotential
Permeable Membrane
Na+ Na+
Cl- Cl- Membrane is permeable to both Na+ & Cl-
V
Voltage is zero.
Neutral(1%) Neutral(10%)*Half-cell potential is zero.
Medical Instrumentation
Biopotential
Semi-permeable Membrane
Na+
Cl-
Na+
Cl- Semi-permeable Membrane(Only to Na+)
V
*Half-cell potential is zero.
Diffusion Repulsive
Neutral(1%) Neutral(10%)
Medical Instrumentation
Biopotential
Na+
Cl-
Na+
Cl-
V
Diffusion Repulsive
Neutral(1%) Neutral(10%)
Electronic Coulomb force < Diffusion force -> Na+ permeate.
Electronic Coulomb force = Diffusion force -> Na+ stop permeating
Dynamic Equilibrium
LOGO
Medical Instrumentaiton