[IEEE 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference (2012 NSS/MIC) - Anaheim,...
Click here to load reader
Transcript of [IEEE 2012 IEEE Nuclear Science Symposium and Medical Imaging Conference (2012 NSS/MIC) - Anaheim,...
2012 IEEE Nuclear Science Symposiwn and Medical Imaging Conference Record (NSS/MIC) MI0-61
Prototype integrated system of DOI- PET and the RF
coil specialized for simultaneous PET-MRI
measurements
Fumihiko Nishikido, Takayuki Obata, Naoko Inadama, Eiji Yoshida, Hideaki Tashima, Mikio Suga, Hideo
Murayama, Hiroshi Ito and Taiga Yamaya, Member, IEEE
Abstract-We are developing a PET system integrated with a
birdcage RF coil for PET-MRI. The integrated system is intended
to realize a highly sensitive PET system for simultaneous
measurements. In the proposed PET-MRI system, PET detectors
which consist of scintillator crystal blocks, photo sensors and
front-end circuits with four-layer DOl encoding capability are
placed close to the objective. Therefore, the proposed system can
achieve high sensitivity without degradation of spatial resolution
at the edge of the FOV due to parallax error. The photo sensors
and front-end circuits should be shielded to minimize noises from
the MRI and noise influence on the MRI imaging. Elements of
the RF coil are inserted between the crystal blocks inside of the
shielding material so as not to interfere with the RF-pulse.
At the last MIC conference, we demonstrated the possibility of
realizing the proposed PET-MRI system with a prototype PET
detector and a commercial RF-coil. The prototype PET detector
consisted of a LGSO crystal block and a 4 x 4 MPPC array. The
crystals were arranged in a 4 x 4 x 4 layer with four-layer DOl
capability. The detector and electrical circuit were packaged in
an aluminum shielding box. Since they were located inside the
MRI magnetic field, most circuit elements were made of
nonmagnetic materials.
As a next step, we constructed a new RF-coil system which can
be mounted on PET detectors between each coil element. We
carried out experiments with the prototype RF-coil and the four
layer DOl detector. The prototype RF-coil has eight RF-coil
elements and the PET detector can be mounted on gaps between
them. Only one detector was mounted in the gap positioned at
one side of the prototype RF-coil. We evaluated the performances
from the 2D flood histogram, energy resolution and MRI images.
We observed no degradations of the performance of the PET
detector and MRI image in simultaneous measurements.
I. INTRODUCTION
WE are developing a PET system integrated with a birdcage RF coil for PET-MRI. A conceptual scheme of the
proposed PET-MRI system is shown in Fig. 1; it is intended to realize a highly sensitive PET system for simultaneous measurements.
Manuscript received November 19,2012. This work was supported in part by the U. S. Department of Commerce under Grant No. BSI23456 (sponsor acknowledgment goes here).
F. Nishikido, T. Obata, N. Inadama, E. Yoshida, H. Tashima, H. Murayama, H. Ito and T. Yamaya are with the National Institute of Radiological Sciences, Chiba 263-8555, Japan, (telephone: +81-43-206-3187, e-mail: funis@nirs. go.jp. [email protected]. inadama@nirs. go.jp. rush@nirs. go.jp, mur@nirs. go.jp, [email protected], taiga@nirs. go.jp).
M. Suga is with Chiba University, Chiba 263-8522, Japan (telephone: +81-43-290-3688, e-mail: [email protected]).
PET detectors which consist of a scintillator block, photo sensors and front-end circuits with four-layer DOl encoding capability are placed close to the objective. Therefore, the proposed system can achieve high sensitivity without degradation spatial resolution at the edge of the field of view due to parallax error. The photo sensors and front-end circuits should be shielded to minimize noises from the MRI and noise influence on MRI imaging. If the shielding material is inside the RF coils, the RF pulse is blocked by the shielding material and then complete images cannot be obtained. Therefore, each RF coil element is inserted between the scintillator crystal blocks and then medially located in the shielding materials. Previously, we demonstrated the possibility of realizing the
proposed PET-MRI system with a prototype PET detector and a commercial RF-coil [1]. As a next step, we constructed a new RF-coil system on which PET detectors could be mounted in gaps between each coil element. We carried out experiments with the prototype RF-coil and a four-layer DOl detector. We evaluated the PET detector performance measured with a 3.0T MRI and image quality of the prototype RF-coil in simultaneous measurements.
Gradient coil
RF coil elem ent
Scintillator
Shielding box (MPPC and circuits)
Fig. I. A conceptual scheme of the proposed PET-MRI system
II. MATERIAL AND METHODS
A. Four-layer DOl detector for PET-MRl
The prototype PET detector consisted of an LGSO crystal
block and a 4 x 4 mUlti-pixel photon counter (MPPC) array
(SI1064-050P). The size of each crystal element was 2.9 nun
x 2.9 mm x 5.0 mm. The crystal elements were arranged in a 4
x 4 x 4 layer with reflectors. Details of the reflector
978-1-4673-2030-6/12/$3l.00 ©20 12 IEEE 2750
arrangement and the four-layer DOl encoding method were
described in ref. [2]. Readout pixels of the MPPC were 3 x 3
mm3 and they consisted of 50 �m cells. The detector and
electrical circuits were packaged in an aluminum shielding
box. Since they were located inside the MRI magnetic field,
most circuit elements were made of nonmagnetic materials.
Figure 2 shows a photograph of the prototype detector.
Specified temperature control and correction of variance of the
MPPC gains were not applied.
Fig. 2. Photograph of the prototype four-layer DOL detector for PET MR!.
B. Setup
We constructed a prototype RF-coil dedicated to the
proposed PET-MRI system. The diameter of the RF-coil was
26cm. There were eight RF-coil elements and the PET
detector could be mounted on gaps between the RF -coil
elements. Only the one detector as used in the performance
evaluation experiment was mounted in the gap at the right side
of the prototype RF-coil as shown in fig. 3.
Fig. 3. Experimental setup for simultaneous measurements of PET and MRI.
Figure 4 shows a block diagram of data acquisition system
for experiments. The MPPC output signal of each MPPC
were divided into two signals. The first signals were fed into
simple non-inverting amplifier circuits and individually
recorded by 16ch analog-to-digital converters (C009H, Hoshin
Electronics Co., LTD.) as list-mode data. The other signals fed
into sum modules and used as the trigger and gate signal.
In all measurement experiments, only the detector and read
out cable were placed in the MRI room. The other modules,
such as a power supply, shaping amplifiers and ADCs, were
placed in the operator room as shown fig.4.
MRl room op erator room
Fig. 4. Experimental setup for simultaneous measurement experiments.
C. Experiment
First, we evaluated performance of the prototype four-layer
DOl detector in simultaneous measurements with the 3.0T
MRI (Siemense, verio). In the measurements, a 22Na point-like
source was positioned in front of the DOl-PET detector. We
compared between position histograms with/without MRI
measurement.
Second, we evaluated the influence of the PET
measurement on the MRI images. We obtained MRI images
(a) without the PET detector, (b) with the PET detector turned
off and (c) with the PET detector turned on (simultaneous
measurement). The gradient echo method was applied as the
MRI protocol.
III. RESULTS
Figures 5 shows position histograms for uniform irradiation
of the 22Na point source with and without MRI measurements.
Only 511 ke V photo-peak events were indicated on these
position maps. Each spot represented interacting events in
certain crystal elements. The crystals in all four layers could
be identified in both position histograms. Comparison of both
position histograms showed no degradation of crystal
identification performance by MRI measurement. The energy
resolutions of 16.9% and 17.4% were obtained for a single
crystal element with and without MRI measurements,
respectively.
(a)Without MRI measurements (b) Simultaneous measurements Fig. 5. Position histograms for unifonn irradiation of 511 keY gamma rays with and without MRl measurements
Figures 6 show magnitude images (left) and phase images
(right) measured for an MRI phantom by the gradient echo
method. The top, middle and bottom figures were obtained in
the experiments without the PET detector, with the PET
detector turned off and with simultaneous measurements,
2751
respectively. Signal to noise ratios for each of the experiments
were (a)109.0, (b)88.9 and (c)87.1. These results showed that
the influence of the PET measurements was negligible in
simultaneous measurements. In addition, the influence of
presence of the PET detector on the MRI images was larger
than that of noise from the electric current of the PET detector.
(a) Without PET detector
Fig. 6. MRI magnitude images (left) and phase images (right) measured by the gradient echo method. Red boxes show the position of the prototype PET detector.
IV. DISCUSSION AND CONCLUSION
We are developing a new system with integrated PET
detectors and an RF coil for PET-MRI. This system will
realize high spatial resolution and sensitivity for the PET
scanner by using the four-layer DOl detectors. We constructed
the RF-coil system for PET-MRI on which PET detectors can
be mounted between each coil element. We carried out
experiments with the prototype RF-coil and the four-layer
DOl detector. The PET performance with the 3.0T MRI and
the image quality of the prototype RF-coil in simultaneous
measurement experiments were evaluated. As a result,
sufficient performance of the PET detector and MRI imaging
function were observed in simultaneous measurements.
In the future, we will increase the number of PET detectors
and evaluate the perfonnance in other simultaneous
measurement experiments.
REFERENCES
[I] T. Tsuda, H. Murayama, K. Kitamura, et aI., "A Four-Layer Depth of Interaction Detector Block for Small Animal PET", IEEE Trans. Nucl. Sci., vol. 51, No. 5, pp. 2537-2542, Oct., 2004
[2] F. Nishikido, A. Tachibana, T. Obata, "Feasibility study for a PET detector integrated with an RF coil for PET-MRI", 2011 IEEE NSS& M1C conf., MICI3-7, 2011, Spain
2752