Icd basic 042411(dr chang)
-
Upload
taiwan-heart-rhythm-society -
Category
Technology
-
view
2.374 -
download
1
Transcript of Icd basic 042411(dr chang)
Basic ICD Treatment
張坤正 醫師
中國醫藥大學 內科副教授
中國醫藥大學附設醫院 心臟內科主任
Advance EP Training
April/24/2011
ICD Concepts
ICD Automated Functions
Troubleshooting
ICD ConceptsOverview
•Evolution of ICDs•The ICD System•Components
– Device• Battery, Capacitors & Voltage• Circuitry• Connector Blocks
– Leads• Transvenous & Epicardial• Electrical Design• Connectors: IS-1 & DF-1
– Programmer
The Evolution of ICDs
1970
• Patent granted for first totally implantable defibrillator
• System used an intracardiac catheter and SQ patch with detection via RV pressure transducer
ICD Evolution
Michael Mirowski (1924-1990)
ICD Evolution
•1947 First human internal defibrillation •1956 First human external defibrillation•1969 First external canine prototype tested•1970 First implantable prototype (895 g)•1975 First implantable Defib in canines (250 g)•1980 First human implant @ Johns Hopkins•1985 ICD market released (350 units)
•1991 Non thoracotomy lead systems•1995 Pectoral ICD systems•1997 ICD & DDD•1998 ICD & DR•1999 ICD & Atrial Defibrillation•2001 ICD & Resynchronization Therapy
ICD Evolution
THE ICD SYSTEM
How it Works
The ICD System
*Animation
The ICD SystemHow it Works
Atrium & Ventricle• Bradycardia sensing • Bradycardia pacing• Antitachycardia
pacing
Ventricle• VT prevention• Antitachycardia pacing• Cardioversion• Defibrillation
Device Components
ICD
ICDDevice Components
c
c
c
Battery – Provides low voltage energy
High Voltage Shock – Delivered on demand
Capacitor – Stores high energy for use
Transformer – Multiplies Voltage
ICDHow High Voltage Is Created
How High Voltage Is Created
VoltageMultiplier
andTransformer High
VoltageCapacitor
High EnergyShocks(Up to 800 Volts)
Battery
Pacing Voltages
For Pacing
For Defibrillation
3.2 Volts
ICD
Battery Depletion
ICD
Battery DOWN
Charge Time UP
Both can Indicate Elective Replacement
(ERI)
Battery & Charge Time Relationship
Battery & Charge Time Relationship
Battery Voltage & Charge Time in an optimized ICD battery
Battery & Charge Time Relationship
The Importance of Charge Time
10 Seconds of VF
The longer the charge time, the longer this rhythm continues.
ICDConnector Block
ICDConnector Block
ICD Connector Blocks
Components
23
New DF4 DR Header (2 set screws)
Today’s Current® DR RF Header (6 set screws)
COMPONENTS
High Voltage Leads
High Voltage Leads
Transvenous
Introduced into the heart through a vein.
Epicardial
Placed directly on the epicardium of the heart.
High Voltage Leads
Passive
Active
No Fixation
High Voltage LeadsTypes of Transvenous Leads
Transvenous
Endocardial Lead Systems
Passive Fixation (Tines)
Active Fixation (Screw)
Transvenous LeadsPlacement
Single Coil Dual Coil
Distal Coil
Proximal Coil
ConnectorPins
Typical Transvenous Lead
High Voltage Leads
Steroid Tip
Transvenous Lead
*Steroid not available in all leads
Benefits of Steroid:• Reduces acute Trauma and
Inflammation
• Rapid Tissue Damage Repair
• Reduced Fibrosis
Steroid Tip
Transvenous Lead
*Steroid not available in all leads
Lead Design
Optimal Tissue Contact
Transvenous Lead
*Steroid not available in all leads
Electrical Design
Lead Electrical Design
• High Voltage Leads can have 2 Circuits:
PACING & SENSING
SHOCKING
Circuits
Lead Electrical Design
Bipolar
Circuits
PACING & SENSING
Lead Electrical Design
•The electrical circuit for pacing and sensing includes:•Cathode – negative electrode (Tip)
•Anode: positive electrode (Ring or Coil)
Bipolar Configuration
Pacing/Sensing Circuit
Lead Electrical Design
True vs. Integrated
Circuits
Bipolar
PACING & SENSING
Polarity
Bipolaruses dedicated ring can as anode
Integrated Bipolaruses RV Coil can as anode
Pacing/Sensing Circuit
PolarityPacing/Sensing Circuit
Lead Electrical Design
High Voltage Coils
Circuits
True vs. IntegratedBipolar
PACING & SENSING
SHOCKING
Polarity
Single Coil• 1 high voltage coil per lead:
- RV (Right Ventricle)- Other extra HV Coils/Patches
• 1 or 2 Connector Pins - (HV only or P/S + HV)
Dual Coil• 2 High Voltage coils per lead:
- RV + SVC
• 3 Connector Pins (P/S + 2 HV)
Shocking Circuit
High Voltage Leads
Pace / Sense Connection = IS-1(International Standard)
High Voltage Defibrillation Connections = DF-1(Defibrillation Standard)
Connector Pins
Shocking Circuit
•The pathway in which high energy is delivered•Determined by electrodes used:
– HVA = active device– HVB = RV Coil– HVX = optional HV electrode (such as an SVC Coil)
Example: A > B
Vectors
A(Active Can)
B(RV HV Coil)
Shocking Circuit
Example: A > B(Active Can > RV HV Coil)
Vectors
Shocking Circuit
Example of Dual Coil: AX > B(Active Can + SVC HV Coil > RV HV Coil)
Vectors
Shocking Circuit
A > B AX > B
Vectors
COMPONENTS
Programmer
ICD Automated Functions
ICD Automated FunctionsOverview
• Sensing–The EGM Signal–The Sensing Circuit–Marker Channels
• Detection–Suspension (Magnet)–Rate & Duration (NID)–Detection Zones–SVT Discrimination
• Single Chamber• Dual Chamber
–Redetection
• Therapy –Tachyarrhythmia
• Anti-tachycardia Pacing (ATP)
• Cardioversion (Non-committed)
• Defibrillation (Committed)
–Bradyarrhythmia• Pacing Modes• Parameters• Other Brady Therapies
Low Power
ICD Automated Functions
Sensing
Sensing
• Sensing - what the device “sees”
• Electrical Activity - what the device is looking for
• Lead – contains the ‘eyeball’ of the device
SensingTrue vs. Integrated Bipolar
•Range of ‘eyeball’ is determined by polarityTrue Bipolar Sensing
Tip-to-ringSmaller surface areaMore “localized” sensing
Integrated Bipolar Sensing
Tip-to-coilLarger surface areaBroader sensing area
Sensing
•The ‘eye’ of the device– Constantly watches for electrical activity
– Sees or Senses all electrical signals that pass
– Signal produced by a passing electrical wave is called an IntracardiacElectrogram (EGM)
SENSING
The Intracardiac Electrogram EGM
Sensing
• Sensing is:
–The process of identifying cardiac depolarizations from an intracardiac electrogram
SensingThe EGM Signal
• The signal from a depolarization wave passing between sensing electrodes
Depolarization WaveProcessed by
Device
The EGM Signal
• Measured by:– Amplitude
• Peak-to-peak measurement (height) of deflection
• Measured in Millivolts (mV)
– Slew Rate• Speed of deflection change over time• Measured in volts per second (V/s)
The EGM Signal
• Measured Peak-to-Peak
Typical Ventricular EGM
• >5 mV for optimal sensing
Amplitude
The EGM Signal
• Measures Peak-to-Peak change in voltage over time
• R >.75 volts/sec for optimal sensing
Typical Ventricular EGM
Slew Rate = [V/s]
dVdt
Slew Rate
The EGM Signal
• Nearfield– Electrodes are close in range
• EGM Source = Tip-to-Ring / Tip-to-Coil
– Commonly a narrow signal (less myocardium in range)
– Used for arrhythmia detection • Devices are hardwired to nearfield for detection
Fields
The EGM Signal
EGM Source = Tip-to-Ring / Tip-to-Coil
Tip-to-Ring
(Bipolar)
Tip-to-Coil
(Integrated Bipolar)
Nearfield
The EGM SignalNearfield
Morphology Comparison
SINUS RHYTHM VT
EGM Source = Tip-to-Ring / Tip-to-Coil
The EGM Signal
•Nearfield– Electrodes are close in range
•EGM Source = Tip-to-Ring / Tip-to-Coil– Narrow signal – less myocardium in range– Used for arrhythmia detection
•Farfield–Electrodes are further in range
•EGM Source = Variable–More myocardium covered in range–Can resemble surface EKG
Fields
The EGM Signal
EGM Source = Variable
Can-to-RV Coil(HVA to HVB)
RV Coil-to-SVC Coil
(HVB to HVX)
Farfield
The EGM SignalFarfield
Morphology Comparison
EGM Source = Variable
SINUS RHYTHM VT
SENSING
The Sensing Circuit
The Sensing Circuit
• Upon being ‘seen,’ a signal is transported through the sensing circuit
The Sensing CircuitThe Amplifier
The Sensing CircuitThe Bandpass Filter
The Sensing CircuitRectifier
The Sensing CircuitLevel Detector
The Sensing CircuitLevel Detector
The Sensing Circuit
The Sensing Circuit
Sensitivity
•Minimum amplitude of electrical signal that registers as a sensed event
– Set in Millivolts (mV)– Programmable Setting
Sensitivity
If set at .15 mV
Programmed
SensitivityProgrammed
If set at .3 mVT-wave Not Sensed
SensitivityAuto-Adjusting
Sensitivity
• Allows sensing of fine VF waves
• Prevents sensing of T-waves, cross-chamberevents, and pacing artifacts
Auto-Adjusting
Auto-Adjusting SensitivityDecay Constant
Sensed Refractory
Max R-wave amplitude measured at 4 mV
Max R- wave amplitude measured at 7 mV
Threshold Start set to 50% of measured R-wave or 3 mV
2 mV 3 mV
R- wave is Sensed
Sensed Refractory
Maximum Sensitivity
Ventricular
Threshold Start
Decay Delay
• The Decay Delay holds the sensitivity threshold at the starting value for a programmable amount of time
60 ms
0 ms
R-wave
T-wave
Question?
What are the consequences of undersensing?
• Failure to sense VF, therefore, no therapy initiated
• Failure to treat VT that can accelerate to VF
• Can be fatal
Question?
What are the consequences of oversensing?
• Inappropriate therapies
• Potential for inducing fatal arrhythmias
• Patient suffering due to inappropriate therapy
SENSING
Marker Channels™
•Display:– Pace/Sense annotations
– ICD functional annotations:
• Sensing
• Detection
• Therapy
• Other information
Marker Channel™
* in Medtronic devices
ICD Automated Functions
Detection
Question?
What does “Detection” mean?
A device defined arrhythmia is present based on what is sensed.
Detection
• Confirms a sensed rhythm as an arrhythmia based on:– Rate – Duration
Detection
• Measured in:–Beat-to-beat intervals (milliseconds), or–Beats-per-minute (BPM)
Detection Rate
• Classifies rhythm by detection zone:–VT = Ventricular Tachycardia–VF = Ventricular Fibrillation
• Programmable in ranges of ratesExample: VT = 162 bpm – 188 bpm
VF = 188 bpm and faster
Detection
• Measured in:– Number of intervals to detect (NID), or– Length of time to detect
• Programmable by:– Beat or interval counters
• Consecutiveex: 16 beats within the detect zone
• Probabilistic (percentage or fraction) ex: 12 out of 16 beats within the detect zone
– Time in seconds
Detect Duration
Detection
Used for detection of VT
Consecutive Counter
Detection
NID = 12/16
Probabilistic Counter
Detection
Non-Sustained: Duration not met / No Detection
Sustained: Duration met / Detection Occurs
Detect Duration
Question?
What if the fast, sustained ventricular rhythm is the result of an SVT?
SVT Discriminators
•SVT Discriminators:– Prevent detection of tachyarrhythmias
caused by the presence of an SVT
– Prevent inappropriate, unnecessary therapy due to rapid SVT conduction
– Are present in some form in most ICDs
SVT Discriminators
•Discriminate based on:– Waveform morphology
• EGM Width (single chamber)
• Wavelet– Onset of arrhythmia– Stability of arrhythmia– Relationship between P- and R-waves
(dual chamber / requires an atrial lead)
SVT Discriminators
•Measures and stores the QRS characteristics of a normal sinus beat
•Identifies SVT vs. VT based on the QRS changes that occur in most VTs
SINUS RHYTHM
VT
Waveform Morphology
Electrical wavefronts generated by focus approach electrodes from different directions depending on the focus location.
Sinus
Supra-Ventricular
Ventricular
Classification: MD Morphology Discrimination
SVT Discriminators
•Match Threshold Programming– Lower %
• More likely to withhold appropriately; less likely to detect true VT
– Higher %• Less likely to appropriately withhold; more likely to detect true VT
Wavelet
SVT Discriminators
•Based on the premise that most VTs are characterized by a sudden onset
•Evaluates the acceleration of the ventricular rate
•Discriminates between:–Gradual rate increase–Abrupt rate increase
•Determines VT present if rateincrease is abrupt
Onset
SVT Discriminators
• When On:– Averages 4 beats and compares with an
average of previous 4 (multiplied by programmed Onset Percentage)
• Onset met if recent average is less than previous
Onset
SVT Discriminators
• Onset Percentage = 81%
530ms X 81% = 430ms
430ms ≠ 460ms = Onset Not Met
Onset
* in Medtronic devices
SVT Discriminators
•Based on the premise that AF conducts irregularly to the ventricles (and VT is a stable, regular rhythm)
•Discriminates regular from irregular intervals within a detect zone
Stability
SVT Discriminators
UnstableVaries >50 ms from previous 3
Stability = 50 msStability
SVT Discriminators
• Considers P and R relationship to discriminate SVT from VT
• Dual Chamber (requires an atrial and ventricular lead)
• Can be used in conjunction with other discriminators
Dual Chamber
Stability40 ms
V > AV = A
Compare atrial & ventricular frequency
V < A
Ventricular frequency > VT detection
Morphology45%
5 out of 8 AVA80 ms
Any
SVTOnVTOff
Bigeminal Avoidance
Morphology45%, 5/8ATU On
OnsetPassive
Optimized DR Discriminators
SVT Discriminators
•What it looks for:–Atrial Fibrillation / Flutter–Sinus Tachycardia–Other 1:1 SVTs (such as AVNRT)
•How it works:–Analyzes:
• Pattern (P:R wave relationship)• AV Association• Rate• Regularity (R waves)
PR Logic™
PR Logic™
•Distinguishes SVTs by analyzing P and R-wave:– Pattern: number and position of atrial events relative to ventricular events
Pattern
PR Logic™Example: Sinus Tachycardia
PR Logic™Example: Atrial Fibrillation
PR Logic™Example: AVNRT (Other 1:1 SVTs)
Junctional Zone
Question?
What about far-field R-wave sensing in the atrial channel?
PR Logic™
•Algorithm identifies and withholds therapy for:–Consistent, short-long pattern
Far-field R-wave Sensing
Question?
What happens when a magnet is applied over an ICD?
Tachy Detection
• ICDs of different manufacturers respond to magnets differently
•Use when:– EMI is present (surgery, TENS, etc.)– Temporary suspension is preferred over permanent programming
– Therapy is temporarily not desired
Magnet Mode
ICD Automated Functions
Therapy
Question?
Can you name some therapies delivered by an ICD?
ICD Therapies
• ICD Therapy– Low Power (Pacing Therapies)
• Anti-tachycardia Pacing (ATP)• Bradyarrhythmia Pacing
– High Power (Shock Therapies)• Cardioversion• Defibrillation
Anti-Tachycardia Pacing
Anti-Tachycardia Pacing
Re-entry initiated
ATP delivered at a rate faster than tachyarrhythmia.Wavefronts collide.
Subsequent Pulse: Wavefronts collide closer to re-entry circuit
Subsequent Pulses: Wavefronts collide even closer to re-entry circuit
Arrhythmia terminated
Anti-Tachycardia Pacing
•Has programmable:–Sequences – the number of times ATP will be applied upon re-detection
Anti-Tachycardia Pacing
•Has programmable:
–Pulses – the number of pulses per sequence
Anti-Tachycardia Pacing
•Has programmable:– Sequences – the number of times ATP will be applied upon re-detection (max = 10 in most)
–Pulses – the number of pulses per sequence (1-15)
–Rate of pulses delivered (percent or ms)
Anti-Tachycardia Pacing
• Types:–Burst
•A series of pacing pulses delivered at equal intervals
•Interval decrement per sequence
Anti-Tachycardia PacingBurst
Anti-Tachycardia Pacing
•Types:–Burst
•A series of pacing pulses delivered at equal intervals
•Interval decrement per sequence–Ramp
• A series of pacing pulses delivered at ever decreasing intervals
• Adds a pulse per sequence
Anti-Tachycardia PacingRamp
Anti-Tachycardia PacingBurst and Ramp Comparison
Programmed Values:
Number of S1 Pulses = 4Number of Sequences =
4R- S1% = 91%Decrement* = 10 ms
* Decrement between sequences
Programmed Values:
Number of S1 Pulses = 4Number of Sequences = 4R-S1% = 91%Decrement* = 10 ms
* Decrement between pulses
* Adds a pulse per sequence
ICD Therapies
•Tachyarrhythmia Therapy
–Cardioversion (CV)•Non-committed shock therapy (must synchronize to an R-wave to be delivered)
•Designed to treat re-entrant tachycardias
–Anti-Tachycardia Pacing (ATP)•Pacing pulses delivered at a rate faster than the rhythm detected
•Can successfully terminate re-entrant tachycardias
Low Power
High Power
Cardioversion
•Delivers shock on an R-wave•Aborts if synchronization cannot be obtained due to arrhythmia termination
CardioversionProgramming
ICD Therapies
•Tachyarrhythmia Therapy
–Cardioversion (CV)•A non-committed shock (must synchronize to an R-wave to be delivered)
•Designed to treat organized tachyarrhythmias
–Anti-Tachycardia Pacing (ATP)•Pacing pulses delivered at a rate faster than the rhythm detected
•Can successfully terminate re-entrant tachycardias
Low Power
High Power
–Defibrillation Shock•A shock delivered to the heart to terminate a tachyarrhythmia
High Power
ICD Therapy
VT
FVT
VF
Benefits of Tiered Therapy
Defibrillation Programming
*Medtronic Programming Screen
Question?
When does the device call a therapy a success?
ICD Therapies
•“8 to terminate” rule–8 sinus and/or paced beats outside the slowed programmed detection zone
Termination/Redetection
* in Medtronic devices
ICD System Troubleshooting
Troubleshooting
•ICD System Issues–Lead System & Connections–Device–Other Issues (Patient, EMI)
Overview
ICD SYSTEM ISSUES
Lead Systems & Connections
Question?
In an ICD System, what component is generally known to be the source of most issues?a. device
b. lead
ICD System Issues
• Points of Failure– Acute (common suspects)
• Connector• Lead Dislodgement/Perforation
Leads
ICD System Issues
Points of Failure– Chronic (common suspects: High points of stress/pressure)
Lead fracture (commonly exhibits HIGH impedance/resistance)
Lead insulation break (commonly exhibits LOW impedance/resistance)
Leads
ICD System Issues
Points of Failure– Acute (common suspects)
Connector
What kinds of things can go wrong here?
Can you identify a problem?
Connector
Set screw is obstructing the bore
ICD System Issues
Connector
ICD System Issues
What might happen if a lead is connected?
ICD System Issues
What is going on here?
Set Screw Noise
Connector
ICD System Issues
• Points of Failure– Acute (common suspects)
• ConnectorLead dislodgement/perforation
How do we know when this occurs?
Common signs– Intermittent or loss of capture– Intermittent or loss or sensing– Inappropriate therapy during SVT
(in this example due to atrial lead dislodgement)
Lead Dislodgement
Atrial Dislodgement
ICD System Issues
Can you identify the dislodged lead(s)?
Lead Dislodgement
ICD System Issues
Lead Dislodgement
ICD System Issues
• Avoiding dislodgements– Ensure sufficient slack in lead– Use suture sleeves– Check lead tip stability during implant
ICD System Issues
• Occurs when:– Lead tip exits the heart, through the heart
Lead Perforation
ICD System Issues
•Diagnosing–Can be seen on x-ray–Definitive by Echo–Threshold and lead impedance may remain unchanged (due to possible continuous contact with tissue outside the heart)
•Possible patient symptoms–Change in pressure–Cardiac tamponade–Dyspnea
Lead Perforation
ICD System Issues
• Points of Failure– Acute (common suspects)
• Connector• Lead dislodgement/perforation
– Chronic (common suspects: High points of stress/pressure)
Lead fracture (commonly exhibits HIGH impedance)
Can you identify a problem?
1st Rib-Clavicle Crush (lead fracture)
ICD System Issues
Lead Crush
1st Rib & Clavicle Crush
Lead Fracture
ICD System Issues
Lead Fracture
ICD System Issues
Can you identify the fracture?
• Common behavior– Erratic sensing– Intermittent or loss of capture– High lead impedance
Lead Fracture
ICD System Issues
Lead Fracture
ICD System Issues
Example
ICD System Issues
• Points of Failure–Acute (common suspects)
• Connector• Lead dislodgement/perforation
– Chronic (common suspects: high points of stress/pressure)
Lead fracture (commonly exhibits HIGH impedance)
Lead insulation break (commonly exhibits LOW impedance)
Coaxial Inner Insulation Breach
Insulation Break
ICD System Issues
• Common Behavior–Lead impedance
• Low• Can be intermittent
–Capture threshold• Sudden rise or loss• Can be intermittent
–Sensing• Over/undersensing• Can be intermittent
Insulation Break
ICD System Issues
Example
ICD SYSTEM ISSUES
Device
ICD System Issues
• Common device issues:– Long charge times– Battery depletion
• Elective Replacement (ERI)• End of Life (EOL)
Device
– Inappropriate programmingAccelerationOver/UndersensingOutput
T-wave oversensing
Device
ICD System Issues
Can you identify a problem?
Device
ICD System Issues
Undersensing*Medtronic Programming Screen
Can you identify the problem?
Device
ICD System Issues
Undersensing
Can you identify the problem?
Infection/Erosion/Allergy
Allergy/Infection/Allergy
Allergy Test Kit Contents:
Polyurethane
Silicone
Titanium
Platinum Iridium
Polysulfone
EpoxyGold-sputtered devices
Infection/Erosion/AllergyAllergy Testing
Allergy Test Metals(included in kits)
Abandoned Leads
• Chatter between active and abandoned leads may cause inappropriate therapy
Electromagnetic Interference(EMI)
60 Hz (oversensing)
• The sensing of electrical signals other than those produced by the heart
Test Your Skills
What might you suspect from this EGM?
Case Study (Noise)
• Male 62 years• ICD Implant for VT/VF• Patient got 11 shocks
Fast Path Summary
Trends: Signal Amplitude
Trends: Lead Impedance
Impedance Decrease
Appropriate or Inappropriate Therapy?
Episode 52A/V lead 皆有noise
Case study (Morphology Change)
• Female 72 years old• ICD for VT episodes• Standard follow-up patient didn’t feel
anything.
FastPath Summary
Episode Directory
Episode 1: Appropriate or Inappropriate Therapy?
Morphology score low
Diagnostics
Therapy
MD Auto Update
Morphology auto update score increase to 100
Episode 3: Correct Diagnosis?
SVT Diagnosis
Diagnostics
After morphology update Correct Discreamation is SVT
CMUH
CMUH
Indications for ICD Therapy
Implantable Cardioverter-Defibrillators
ICD therapy is indicated in patients who are survivors of cardiac arrest due to ventricular fibrillation or hemodynamically unstable sustained VT after evaluation to define the cause of the event and to exclude any completely reversible causes.
ICD therapy is indicated in patients with structural heart disease and spontaneous sustained VT, whether hemodynamically stable or unstable.
ICD therapy is indicated in patients with syncope of undetermined origin with clinically relevant, hemodynamically significant sustained VT or VF induced at electrophysiological study.
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.
ICD therapy is indicated in patients with LVEF less than or equal to 35% due to prior MI who are at least 40 days post-MI and are in NYHA functional Class II or III.
ICD therapy is indicated in patients with nonischemic DCM who have an LVEF less than or equal to 35% and who are in NYHA functional Class II or III.
ICD therapy is indicated in patients with LV dysfunction due to prior MI who are at least 40 days post-MI, have an LVEF less than or equal to 30%, and are in NYHA functional Class I.
ICD therapy is indicated in patients with nonsustained VT due to prior MI, LVEF less than or equal to 40%, and inducible VF or sustained VT at electrophysiological study.
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.
Implantable Cardioverter-Defibrillators
ICD implantation is reasonable for patients with unexplained syncope, significant LV dysfunction, and nonischemic DCM.
ICD implantation is reasonable for patients with sustained VT and normal or near-normal ventricular function.
ICD implantation is reasonable for patients with HCM who have 1 or more major† risk factors for SCD.
ICD implantation is reasonable for the prevention of SCD in patients with arrhythmogenic right ventricular dysplasia/cardiomyopathy (ARVD/C) who have 1 or more risk factors for SCD.
ICD implantation is reasonable to reduce SCD in patients with long-QT syndrome who are experiencing syncope and/or VT while receiving beta blockers.
I IIaIIbIII
All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.
† See Section 3.2.4, “Hypertrophic Cardiomyopathy,” in the full-text guidelines for definition of major risk factors.
I IIaIIbIII
I IIaIIbIII
I IIaIIbIII
I IIaIIbIII
Implantable Cardioverter-Defibrillators
ICD implantation is reasonable for nonhospitalized patients awaiting transplantation.
ICD implantation is reasonable for patients with Brugada syndrome who have had syncope.
ICD implantation is reasonable for patients with Brugada syndrome who have documented VT that has not resulted in cardiac arrest.
ICD implantation is reasonable for patients with catecholaminergic polymorphic VT who have syncope and/or documented sustained VT while receiving beta blockers.
ICD implantation is reasonable for patients with cardiac sarcoidosis, giant cell myocarditis, or Chagas disease.
I IIaIIbIII
I IIaIIbIII
I IIaIIbIII
I IIaIIbIII
I IIaIIbIII
All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.
Implantable Cardioverter-Defibrillators
ICD therapy may be considered in patients with nonischemic heart disease who have an LVEF of less than or equal to 35% and who are in NYHA functional Class I.
ICD therapy may be considered for patients with long-QT syndrome and risk factors for SCD.
ICD therapy may be considered in patients with syncope and advanced structural heart disease in whom thorough invasive and noninvasive investigations have failed to define a cause.
ICD therapy may be considered in patients with a familial cardiomyopathy associated with sudden death.
ICD therapy may be considered in patients with LV noncompaction.
III IIaIIaIIa IIbIIbIIbIIIIIIIIIIII IIaIIaIIa IIbIIbIIbIIIIIIIIIIII IIaIIaIIa IIbIIbIIbIIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
I IIaIIbIII
I IIaIIbIII
I IIaIIbIII
I IIaIIbIII
All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.
Implantable Cardioverter-Defibrillators
ICD therapy is not indicated for patients who do not have a reasonable expectation of survival with an acceptable functional status for at least 1 year, even if they meet ICD implantation criteria specified in the Class I, IIa, and IIb recommendations above.
ICD therapy is not indicated for patients with incessant VT or VF.
ICD therapy is not indicated in patients with significant psychiatric illnesses that may be aggravated by device implantation or that may preclude systematic follow-up.
ICD therapy is not indicated for NYHA Class IV patients with drug-refractory congestive heart failure who are not candidates for cardiac transplantation or cardiac resynchronization therapy defibrillators (CRT-D).
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.
Implantable Cardioverter-Defibrillators
ICD therapy is not indicated for syncope of undetermined cause in a patient without inducible ventricular tachyarrhythmias and without structural heart disease.
ICD therapy is not indicated when VF or VT is amenable to surgical or catheter ablation (e.g., atrial arrhythmias associated with the Wolff-Parkinson-White syndrome, RV or LV outflow tract VT, idiopathic VT, or fascicular VT in the absence of structural heart disease).
ICD therapy is not indicated for patients with ventricular tachyarrhythmias due to a completely reversible disorder in the absence of structural heart disease (e.g., electrolyte imbalance, drugs, or trauma).
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.
Implantable Cardioverter-Defibrillators
ICDs in Pediatric Patients and Patients With Congenital Heart Disease
ICD implantation is indicated in the survivor of cardiac arrest after evaluation to define the cause of the event and exclusion of any reversible causes.
ICD implantation is indicated for patients with symptomatic sustained VT in association with congenital heart disease who have undergone hemodynamic and electrophysiological evaluation. Catheter ablation or surgical repair may offer possible alternatives in carefully selected patients.
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.
ICDs in Pediatric Patients and Patients With Congenital Heart Disease
ICD implantation is reasonable for patients with congenital heart disease with recurrent syncope of undetermined origin in the presence of either ventricular dysfunction or inducible ventricular arrhythmias at electrophysiological study.
ICD implantation may be considered for patients with recurrent syncope associated with complex congenital heart disease and advanced systemic ventricular dysfunction when thorough invasive and noninvasive investigations have failed to define a cause.
All Class III recommendations found in Section 3 of the full-text guidelines, “Indications for Implantable Cardioverter-Defibrillator Therapy,” apply to pediatric patients and patients with congenital heart disease, and ICD implantation is not indicated in these patient populations.
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII
All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.
Major Implantable Cardioverter-Defibrillator Trials for Prevention of Sudden Cardiac Death
Trial Year Patients(n)
LVEF Additional StudyFeatures
HazardRatio*
95% CI p
MADIT I 1996 196 < 35% NSVT and EP+ 0.46 (0.26-0.82) p=0.009
MADIT II 2002 1232 < 30% Prior MI 0.69 (0.51-0.93) p=0.016
CABG-Patch 1997 900 < 36% +SAECG and CABG 1.07 (0.81-1.42) p=0.63
DEFINITE 2004 485 < 35% NICM, PVCs or NSVT 0.65 (0.40-1.06) p=0.08
DINAMIT 2004 674 < 35% 6-40 days post-MI and Impaired HRV
1.08 (0.76-1.55) p=0.66
SCD-HeFT 2006 1676 < 35% Prior MI of NICM 0.77 (0.62-0.96) p=0.007
AVID 1997 1016 Prior cardiac arrest
NA 0.62 (0.43-0.82) NS
CASH† 2000 191 Prior cardiac arrest
NA 0.766 ‡ 1-sided p=0.081
CIDS 2000 659 Prior cardiac arrest, syncope
NA 0.82 (0.60-1.1) NS
* Hazard ratios for death from any cause in the ICD group compared with the non-ICD group. Includes only ICD and amiodarone patients from CASH.‡CI Upper Bound 1.112 CI indicates Confidence Interval, NS = Not statistically significant, NSVT = nonsustained ventricular tachycardia, SAECG = signal-averaged electrocardiogram.
Epstein A, et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities. J Am Coll Cardiol 2008; 51:e1–62. Table 5.
Comparison of Medical Therapy, Pacing, and Defibrillation in Heart Failure (COMPANION) Trial
• 1520 patients with NYHA Class III or IV HF, ischemic cardiomyopathy (ICM) or nonischemic cardiomyopathy (NICM) and QRS ≥ 120 ms
• Randomized 1:2:2 to optimal pharmacological therapy (OPT) alone or in combination with cardiac resynchronization therapy with either a pacemaker (CRT-P) or pacemaker-defibrillator (CRT-D)
• Both device arms significantly ↓ combined risk of all-cause hospitalization and all-cause mortality by ~20% compared with OPT
• CRT-D ↓ mortality by 36% compared with OPT (p=0.003)• Insufficient evidence to conclude that CRT-P inferior to
CRT-D
Bristow MR, Saxon LA, Boehmer J, et al. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med 2004;350:2140-50.
Implantable Cardioverter-Defibrillators and Prevention of Sudden Cardiac Death in Hypertrophic Cardiomyopathy
• Multicenter registry study of implanted ICDs in 506 unrelated patients with HCM @ high risk for SCD (family hx of SCD, [septal thickness ≥ 30 mm], NSVT, syncope)
• Mean patient age 42 years (SD=17) and 87% had no or only mildly limiting symptoms
• Appropriate ICD discharge rates were 11% per year for 2o prevention and 4% per year for 1o prevention
• For 1o prevention, 35% of patients with appropriate ICD interventions had undergone implantation for only 1 risk factor
Maron BJ, Spirito P, Shen WK, et al. Implantable cardioverter-defibrillators and prevention of sudden cardiac death in hypertrophic cardiomyopathy. JAMA 2007;298:405-12.
Multicenter Automatic Defibrillator Implantation Trial II (MADIT II)
• 1232 patients ≥ 1 month post-MI and LVEF ≤ 30%• Randomized to ICD (n=742) or medical therapy (n=490)• No spontaneous or induced arrhythmia required for
enrollment• 6% absolute and 31% relative risk ↓ in all-cause mortality
with ICD therapy (p=0.016)
Moss AJ, Zareba W, Hall WJ, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med 2002;346:877-83.
Sudden Death in Heart Failure (SCD-HeFT) Trial
• 2521 patients with NYHA Class II or III HF, ICM, or NICM and LVEF ≤ 35%
• Randomized to 1) conventional rx for HF + placebo; 2) conventional rx + amiodarone; or 3) conventional rx + conservatively programmed shock-only single lead ICD
• No survival benefit for amiodarone• 23% ↓ in overall mortality with ICD therapy• Absolute ↓ in mortality of 7.2% after 5 y in the overall
populationBardy GH, Lee KL, Mark DB, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 2005;352:225-37.
Defibrillator in Acute Myocardial Infarction (DINAMIT) Trial
• 674 patients 6 to 40 days post-MI with LVEF ≤ 35% and impaired cardiac autonomic function
• Randomized to ICD therapy (n=332) or no ICD therapy (n=342)
• Arrhythmic death ↓ in ICD group, but ↑ in nonarrhythmic death (6.1% per year vs. 3.5% per year, HR 1.75 (95% CI 1.11 to 2.76; p=0.016)
• No difference in total mortality
Hohnloser SH, Kuck KH, Dorian P, et al. Prophylactic use of an implantable cardioverter-defibrillator after acute myocardial infarction. N Engl J Med 2004;351:2481-8.
Defibrillators in Nonischemic Cardiomyopathy Treatment Evaluation (DEFINITE) Trial
• 458 patients with NYHA Class I to III, NICM, LVEF ≤ 35% and premature ventricular contractions (> 10/h) or NSVT
• Randomized to standard medical rx alone or in combination with single-chamber ICD
• Strong trend toward ↓ all-cause mortality with ICD therapy, although not statistically significant (p=0.08)
Kadish A, Dyer A, Daubert JP, et al. Prophylactic defibrillator implantation in patients with nonischemic dilated cardiomyopathy. N Engl J Med 2004;350:2151-8.
Notable Changes in 2008 ACC/AHA/HRS Guidelines
1. ICD recommendations are combined into a single list because of overlap between primary and secondary indications.
2. Primary prevention ICD indications in nonischemic cardiomyopathy are clarified using data from SCD-HeFT (i.e., ischemic and nonischemic cardiomyopathies and LVEF ≤35%, NYHA II-III) for support.
3. Indications for ICD therapy in inherited arrhythmia syndromes and selected nonischemic cardiomyopathies are listed.
4. MADIT II indication (i.e., ischemic cardiomypathy and LVEF ≤30%, NYHA I) is now Class I, elevated from Class IIa.
5. EF criteria for primary prevention ICD indications are based on entry criteria for trials on which the recommendations are based.
6. Emphasized primary SCD prevention ICD recommendations apply only to patients receiving optimal medical therapy and reasonable expectation of survival with good functional capacity for >1 year.
7. Independent risk assessment preceding ICD implantation is emphasized, including consideration of patient preference.
8. Optimization of pacemaker programming to minimize unneeded RV pacing is encouraged.9. Pacemaker insertion is discouraged for asymptomatic bradycardia, particularly at night.10. A section has been added that addresses ICD and pacemaker programming at end of life.