Icd basic 042411(dr chang)

210
Basic ICD Treatment 張坤正 醫師 中國醫藥大學 內科副教授 中國醫藥大學附設醫院 心臟內科主任 Advance EP Training April/24/2011

Transcript of Icd basic 042411(dr chang)

Page 1: Icd basic 042411(dr chang)

Basic ICD Treatment

張坤正 醫師

中國醫藥大學 內科副教授

中國醫藥大學附設醫院 心臟內科主任

Advance EP Training

April/24/2011

Page 2: Icd basic 042411(dr chang)

ICD Concepts

ICD Automated Functions

Troubleshooting

Page 3: Icd basic 042411(dr chang)

ICD ConceptsOverview

•Evolution of ICDs•The ICD System•Components

– Device• Battery, Capacitors & Voltage• Circuitry• Connector Blocks

– Leads• Transvenous & Epicardial• Electrical Design• Connectors: IS-1 & DF-1

– Programmer

Page 4: Icd basic 042411(dr chang)

The Evolution of ICDs

Page 5: Icd basic 042411(dr chang)

1970

• Patent granted for first totally implantable defibrillator

• System used an intracardiac catheter and SQ patch with detection via RV pressure transducer

ICD Evolution

Michael Mirowski (1924-1990)

Page 6: Icd basic 042411(dr chang)

ICD Evolution

•1947 First human internal defibrillation •1956 First human external defibrillation•1969 First external canine prototype tested•1970 First implantable prototype (895 g)•1975 First implantable Defib in canines (250 g)•1980 First human implant @ Johns Hopkins•1985 ICD market released (350 units)

Page 7: Icd basic 042411(dr chang)

•1991 Non thoracotomy lead systems•1995 Pectoral ICD systems•1997 ICD & DDD•1998 ICD & DR•1999 ICD & Atrial Defibrillation•2001 ICD & Resynchronization Therapy

ICD Evolution

Page 8: Icd basic 042411(dr chang)

THE ICD SYSTEM

Page 9: Icd basic 042411(dr chang)

How it Works

The ICD System

*Animation

Page 10: Icd basic 042411(dr chang)

The ICD SystemHow it Works

Atrium & Ventricle• Bradycardia sensing • Bradycardia pacing• Antitachycardia

pacing

Ventricle• VT prevention• Antitachycardia pacing• Cardioversion• Defibrillation

Page 11: Icd basic 042411(dr chang)

Device Components

ICD

Page 12: Icd basic 042411(dr chang)

ICDDevice Components

Page 13: Icd basic 042411(dr chang)

c

c

c

Battery – Provides low voltage energy

High Voltage Shock – Delivered on demand

Capacitor – Stores high energy for use

Transformer – Multiplies Voltage

ICDHow High Voltage Is Created

Page 14: Icd basic 042411(dr chang)

How High Voltage Is Created

VoltageMultiplier

andTransformer High

VoltageCapacitor

High EnergyShocks(Up to 800 Volts)

Battery

Pacing Voltages

For Pacing

For Defibrillation

3.2 Volts

ICD

Page 15: Icd basic 042411(dr chang)

Battery Depletion

ICD

Page 16: Icd basic 042411(dr chang)

Battery DOWN

Charge Time UP

Both can Indicate Elective Replacement

(ERI)

Battery & Charge Time Relationship

Page 17: Icd basic 042411(dr chang)

Battery & Charge Time Relationship

Page 18: Icd basic 042411(dr chang)

Battery Voltage & Charge Time in an optimized ICD battery

Battery & Charge Time Relationship

Page 19: Icd basic 042411(dr chang)

The Importance of Charge Time

10 Seconds of VF

The longer the charge time, the longer this rhythm continues.

Page 20: Icd basic 042411(dr chang)

ICDConnector Block

Page 21: Icd basic 042411(dr chang)

ICDConnector Block

Page 22: Icd basic 042411(dr chang)

ICD Connector Blocks

Components

Page 23: Icd basic 042411(dr chang)

23

New DF4 DR Header (2 set screws)

Today’s Current® DR RF Header (6 set screws)

Page 24: Icd basic 042411(dr chang)

COMPONENTS

High Voltage Leads

Page 25: Icd basic 042411(dr chang)

High Voltage Leads

Transvenous

Introduced into the heart through a vein.

Epicardial

Placed directly on the epicardium of the heart.

Page 26: Icd basic 042411(dr chang)

High Voltage Leads

Page 27: Icd basic 042411(dr chang)

Passive

Active

No Fixation

High Voltage LeadsTypes of Transvenous Leads

Transvenous

Page 28: Icd basic 042411(dr chang)

Endocardial Lead Systems

Passive Fixation (Tines)

Active Fixation (Screw)

Page 29: Icd basic 042411(dr chang)

Transvenous LeadsPlacement

Single Coil Dual Coil

Page 30: Icd basic 042411(dr chang)

Distal Coil

Proximal Coil

ConnectorPins

Page 31: Icd basic 042411(dr chang)

Typical Transvenous Lead

High Voltage Leads

Page 32: Icd basic 042411(dr chang)

Steroid Tip

Transvenous Lead

*Steroid not available in all leads

Page 33: Icd basic 042411(dr chang)

Benefits of Steroid:• Reduces acute Trauma and

Inflammation

• Rapid Tissue Damage Repair

• Reduced Fibrosis

Steroid Tip

Transvenous Lead

*Steroid not available in all leads

Page 34: Icd basic 042411(dr chang)

Lead Design

Optimal Tissue Contact

Transvenous Lead

*Steroid not available in all leads

Page 35: Icd basic 042411(dr chang)

Electrical Design

Page 36: Icd basic 042411(dr chang)

Lead Electrical Design

• High Voltage Leads can have 2 Circuits:

PACING & SENSING

SHOCKING

Circuits

Page 37: Icd basic 042411(dr chang)

Lead Electrical Design

Bipolar

Circuits

PACING & SENSING

Page 38: Icd basic 042411(dr chang)

Lead Electrical Design

•The electrical circuit for pacing and sensing includes:•Cathode – negative electrode (Tip)

•Anode: positive electrode (Ring or Coil)

Bipolar Configuration

Pacing/Sensing Circuit

Page 39: Icd basic 042411(dr chang)

Lead Electrical Design

True vs. Integrated

Circuits

Bipolar

PACING & SENSING

Page 40: Icd basic 042411(dr chang)

Polarity

Bipolaruses dedicated ring can as anode

Integrated Bipolaruses RV Coil can as anode

Pacing/Sensing Circuit

Page 41: Icd basic 042411(dr chang)

PolarityPacing/Sensing Circuit

Page 42: Icd basic 042411(dr chang)

Lead Electrical Design

High Voltage Coils

Circuits

True vs. IntegratedBipolar

PACING & SENSING

SHOCKING

Page 43: Icd basic 042411(dr chang)

Polarity

Single Coil• 1 high voltage coil per lead:

- RV (Right Ventricle)- Other extra HV Coils/Patches

• 1 or 2 Connector Pins - (HV only or P/S + HV)

Dual Coil• 2 High Voltage coils per lead:

- RV + SVC

• 3 Connector Pins (P/S + 2 HV)

Shocking Circuit

Page 44: Icd basic 042411(dr chang)

High Voltage Leads

Pace / Sense Connection = IS-1(International Standard)

High Voltage Defibrillation Connections = DF-1(Defibrillation Standard)

Connector Pins

Page 45: Icd basic 042411(dr chang)

Shocking Circuit

•The pathway in which high energy is delivered•Determined by electrodes used:

– HVA = active device– HVB = RV Coil– HVX = optional HV electrode (such as an SVC Coil)

Example: A > B

Vectors

A(Active Can)

B(RV HV Coil)

Page 46: Icd basic 042411(dr chang)

Shocking Circuit

Example: A > B(Active Can > RV HV Coil)

Vectors

Page 47: Icd basic 042411(dr chang)

Shocking Circuit

Example of Dual Coil: AX > B(Active Can + SVC HV Coil > RV HV Coil)

Vectors

Page 48: Icd basic 042411(dr chang)

Shocking Circuit

A > B AX > B

Vectors

Page 49: Icd basic 042411(dr chang)

COMPONENTS

Programmer

Page 50: Icd basic 042411(dr chang)

ICD Automated Functions

Page 51: Icd basic 042411(dr chang)

ICD Automated FunctionsOverview

• Sensing–The EGM Signal–The Sensing Circuit–Marker Channels

• Detection–Suspension (Magnet)–Rate & Duration (NID)–Detection Zones–SVT Discrimination

• Single Chamber• Dual Chamber

–Redetection

• Therapy –Tachyarrhythmia

• Anti-tachycardia Pacing (ATP)

• Cardioversion (Non-committed)

• Defibrillation (Committed)

–Bradyarrhythmia• Pacing Modes• Parameters• Other Brady Therapies

Low Power

Page 52: Icd basic 042411(dr chang)

ICD Automated Functions

Sensing

Page 53: Icd basic 042411(dr chang)

Sensing

• Sensing - what the device “sees”

• Electrical Activity - what the device is looking for

• Lead – contains the ‘eyeball’ of the device

Page 54: Icd basic 042411(dr chang)

SensingTrue vs. Integrated Bipolar

•Range of ‘eyeball’ is determined by polarityTrue Bipolar Sensing

Tip-to-ringSmaller surface areaMore “localized” sensing

Integrated Bipolar Sensing

Tip-to-coilLarger surface areaBroader sensing area

Page 55: Icd basic 042411(dr chang)

Sensing

•The ‘eye’ of the device– Constantly watches for electrical activity

– Sees or Senses all electrical signals that pass

– Signal produced by a passing electrical wave is called an IntracardiacElectrogram (EGM)

Page 56: Icd basic 042411(dr chang)

SENSING

The Intracardiac Electrogram EGM

Page 57: Icd basic 042411(dr chang)

Sensing

• Sensing is:

–The process of identifying cardiac depolarizations from an intracardiac electrogram

Page 58: Icd basic 042411(dr chang)

SensingThe EGM Signal

• The signal from a depolarization wave passing between sensing electrodes

Depolarization WaveProcessed by

Device

Page 59: Icd basic 042411(dr chang)

The EGM Signal

• Measured by:– Amplitude

• Peak-to-peak measurement (height) of deflection

• Measured in Millivolts (mV)

– Slew Rate• Speed of deflection change over time• Measured in volts per second (V/s)

Page 60: Icd basic 042411(dr chang)

The EGM Signal

• Measured Peak-to-Peak

Typical Ventricular EGM

• >5 mV for optimal sensing

Amplitude

Page 61: Icd basic 042411(dr chang)

The EGM Signal

• Measures Peak-to-Peak change in voltage over time

• R >.75 volts/sec for optimal sensing

Typical Ventricular EGM

Slew Rate = [V/s]

dVdt

Slew Rate

Page 62: Icd basic 042411(dr chang)

The EGM Signal

• Nearfield– Electrodes are close in range

• EGM Source = Tip-to-Ring / Tip-to-Coil

– Commonly a narrow signal (less myocardium in range)

– Used for arrhythmia detection • Devices are hardwired to nearfield for detection

Fields

Page 63: Icd basic 042411(dr chang)

The EGM Signal

EGM Source = Tip-to-Ring / Tip-to-Coil

Tip-to-Ring

(Bipolar)

Tip-to-Coil

(Integrated Bipolar)

Nearfield

Page 64: Icd basic 042411(dr chang)

The EGM SignalNearfield

Morphology Comparison

SINUS RHYTHM VT

EGM Source = Tip-to-Ring / Tip-to-Coil

Page 65: Icd basic 042411(dr chang)

The EGM Signal

•Nearfield– Electrodes are close in range

•EGM Source = Tip-to-Ring / Tip-to-Coil– Narrow signal – less myocardium in range– Used for arrhythmia detection

•Farfield–Electrodes are further in range

•EGM Source = Variable–More myocardium covered in range–Can resemble surface EKG

Fields

Page 66: Icd basic 042411(dr chang)

The EGM Signal

EGM Source = Variable

Can-to-RV Coil(HVA to HVB)

RV Coil-to-SVC Coil

(HVB to HVX)

Farfield

Page 67: Icd basic 042411(dr chang)

The EGM SignalFarfield

Morphology Comparison

EGM Source = Variable

SINUS RHYTHM VT

Page 68: Icd basic 042411(dr chang)

SENSING

The Sensing Circuit

Page 69: Icd basic 042411(dr chang)

The Sensing Circuit

• Upon being ‘seen,’ a signal is transported through the sensing circuit

Page 70: Icd basic 042411(dr chang)

The Sensing CircuitThe Amplifier

Page 71: Icd basic 042411(dr chang)

The Sensing CircuitThe Bandpass Filter

Page 72: Icd basic 042411(dr chang)

The Sensing CircuitRectifier

Page 73: Icd basic 042411(dr chang)

The Sensing CircuitLevel Detector

Page 74: Icd basic 042411(dr chang)

The Sensing CircuitLevel Detector

Page 75: Icd basic 042411(dr chang)

The Sensing Circuit

Page 76: Icd basic 042411(dr chang)

The Sensing Circuit

Page 77: Icd basic 042411(dr chang)

Sensitivity

•Minimum amplitude of electrical signal that registers as a sensed event

– Set in Millivolts (mV)– Programmable Setting

Page 78: Icd basic 042411(dr chang)

Sensitivity

If set at .15 mV

Programmed

Page 79: Icd basic 042411(dr chang)

SensitivityProgrammed

If set at .3 mVT-wave Not Sensed

Page 80: Icd basic 042411(dr chang)

SensitivityAuto-Adjusting

Page 81: Icd basic 042411(dr chang)

Sensitivity

• Allows sensing of fine VF waves

• Prevents sensing of T-waves, cross-chamberevents, and pacing artifacts

Auto-Adjusting

Page 82: Icd basic 042411(dr chang)

Auto-Adjusting SensitivityDecay Constant

Page 83: Icd basic 042411(dr chang)

Sensed Refractory

Max R-wave amplitude measured at 4 mV

Max R- wave amplitude measured at 7 mV

Threshold Start set to 50% of measured R-wave or 3 mV

2 mV 3 mV

R- wave is Sensed

Sensed Refractory

Maximum Sensitivity

Ventricular

Threshold Start

Page 84: Icd basic 042411(dr chang)

Decay Delay

• The Decay Delay holds the sensitivity threshold at the starting value for a programmable amount of time

60 ms

0 ms

R-wave

T-wave

Page 85: Icd basic 042411(dr chang)

Question?

What are the consequences of undersensing?

• Failure to sense VF, therefore, no therapy initiated

• Failure to treat VT that can accelerate to VF

• Can be fatal

Page 86: Icd basic 042411(dr chang)

Question?

What are the consequences of oversensing?

• Inappropriate therapies

• Potential for inducing fatal arrhythmias

• Patient suffering due to inappropriate therapy

Page 87: Icd basic 042411(dr chang)

SENSING

Marker Channels™

Page 88: Icd basic 042411(dr chang)

•Display:– Pace/Sense annotations

– ICD functional annotations:

• Sensing

• Detection

• Therapy

• Other information

Marker Channel™

* in Medtronic devices

Page 89: Icd basic 042411(dr chang)

ICD Automated Functions

Detection

Page 90: Icd basic 042411(dr chang)

Question?

What does “Detection” mean?

A device defined arrhythmia is present based on what is sensed.

Page 91: Icd basic 042411(dr chang)

Detection

• Confirms a sensed rhythm as an arrhythmia based on:– Rate – Duration

Page 92: Icd basic 042411(dr chang)

Detection

• Measured in:–Beat-to-beat intervals (milliseconds), or–Beats-per-minute (BPM)

Detection Rate

• Classifies rhythm by detection zone:–VT = Ventricular Tachycardia–VF = Ventricular Fibrillation

• Programmable in ranges of ratesExample: VT = 162 bpm – 188 bpm

VF = 188 bpm and faster

Page 93: Icd basic 042411(dr chang)

Detection

• Measured in:– Number of intervals to detect (NID), or– Length of time to detect

• Programmable by:– Beat or interval counters

• Consecutiveex: 16 beats within the detect zone

• Probabilistic (percentage or fraction) ex: 12 out of 16 beats within the detect zone

– Time in seconds

Detect Duration

Page 94: Icd basic 042411(dr chang)

Detection

Used for detection of VT

Consecutive Counter

Page 95: Icd basic 042411(dr chang)

Detection

NID = 12/16

Probabilistic Counter

Page 96: Icd basic 042411(dr chang)

Detection

Non-Sustained: Duration not met / No Detection

Sustained: Duration met / Detection Occurs

Detect Duration

Page 97: Icd basic 042411(dr chang)

Question?

What if the fast, sustained ventricular rhythm is the result of an SVT?

Page 98: Icd basic 042411(dr chang)

SVT Discriminators

•SVT Discriminators:– Prevent detection of tachyarrhythmias

caused by the presence of an SVT

– Prevent inappropriate, unnecessary therapy due to rapid SVT conduction

– Are present in some form in most ICDs

Page 99: Icd basic 042411(dr chang)

SVT Discriminators

•Discriminate based on:– Waveform morphology

• EGM Width (single chamber)

• Wavelet– Onset of arrhythmia– Stability of arrhythmia– Relationship between P- and R-waves

(dual chamber / requires an atrial lead)

Page 100: Icd basic 042411(dr chang)

SVT Discriminators

•Measures and stores the QRS characteristics of a normal sinus beat

•Identifies SVT vs. VT based on the QRS changes that occur in most VTs

SINUS RHYTHM

VT

Waveform Morphology

Page 101: Icd basic 042411(dr chang)

Electrical wavefronts generated by focus approach electrodes from different directions depending on the focus location.

Sinus

Supra-Ventricular

Ventricular

Classification: MD Morphology Discrimination

Page 102: Icd basic 042411(dr chang)

SVT Discriminators

•Match Threshold Programming– Lower %

• More likely to withhold appropriately; less likely to detect true VT

– Higher %• Less likely to appropriately withhold; more likely to detect true VT

Wavelet

Page 103: Icd basic 042411(dr chang)

SVT Discriminators

•Based on the premise that most VTs are characterized by a sudden onset

•Evaluates the acceleration of the ventricular rate

•Discriminates between:–Gradual rate increase–Abrupt rate increase

•Determines VT present if rateincrease is abrupt

Onset

Page 104: Icd basic 042411(dr chang)

SVT Discriminators

• When On:– Averages 4 beats and compares with an

average of previous 4 (multiplied by programmed Onset Percentage)

• Onset met if recent average is less than previous

Onset

Page 105: Icd basic 042411(dr chang)

SVT Discriminators

• Onset Percentage = 81%

530ms X 81% = 430ms

430ms ≠ 460ms = Onset Not Met

Onset

* in Medtronic devices

Page 106: Icd basic 042411(dr chang)

SVT Discriminators

•Based on the premise that AF conducts irregularly to the ventricles (and VT is a stable, regular rhythm)

•Discriminates regular from irregular intervals within a detect zone

Stability

Page 107: Icd basic 042411(dr chang)

SVT Discriminators

UnstableVaries >50 ms from previous 3

Stability = 50 msStability

Page 108: Icd basic 042411(dr chang)

SVT Discriminators

• Considers P and R relationship to discriminate SVT from VT

• Dual Chamber (requires an atrial and ventricular lead)

• Can be used in conjunction with other discriminators

Dual Chamber

Page 109: Icd basic 042411(dr chang)

Stability40 ms

V > AV = A

Compare atrial & ventricular frequency

V < A

Ventricular frequency > VT detection

Morphology45%

5 out of 8 AVA80 ms

Any

SVTOnVTOff

Bigeminal Avoidance

Morphology45%, 5/8ATU On

OnsetPassive

Optimized DR Discriminators

Page 110: Icd basic 042411(dr chang)

SVT Discriminators

•What it looks for:–Atrial Fibrillation / Flutter–Sinus Tachycardia–Other 1:1 SVTs (such as AVNRT)

•How it works:–Analyzes:

• Pattern (P:R wave relationship)• AV Association• Rate• Regularity (R waves)

PR Logic™

Page 111: Icd basic 042411(dr chang)

PR Logic™

•Distinguishes SVTs by analyzing P and R-wave:– Pattern: number and position of atrial events relative to ventricular events

Pattern

Page 112: Icd basic 042411(dr chang)

PR Logic™Example: Sinus Tachycardia

Page 113: Icd basic 042411(dr chang)

PR Logic™Example: Atrial Fibrillation

Page 114: Icd basic 042411(dr chang)

PR Logic™Example: AVNRT (Other 1:1 SVTs)

Junctional Zone

Page 115: Icd basic 042411(dr chang)

Question?

What about far-field R-wave sensing in the atrial channel?

Page 116: Icd basic 042411(dr chang)

PR Logic™

•Algorithm identifies and withholds therapy for:–Consistent, short-long pattern

Far-field R-wave Sensing

Page 117: Icd basic 042411(dr chang)

Question?

What happens when a magnet is applied over an ICD?

Page 118: Icd basic 042411(dr chang)

Tachy Detection

• ICDs of different manufacturers respond to magnets differently

•Use when:– EMI is present (surgery, TENS, etc.)– Temporary suspension is preferred over permanent programming

– Therapy is temporarily not desired

Magnet Mode

Page 119: Icd basic 042411(dr chang)

ICD Automated Functions

Therapy

Page 120: Icd basic 042411(dr chang)

Question?

Can you name some therapies delivered by an ICD?

Page 121: Icd basic 042411(dr chang)

ICD Therapies

• ICD Therapy– Low Power (Pacing Therapies)

• Anti-tachycardia Pacing (ATP)• Bradyarrhythmia Pacing

– High Power (Shock Therapies)• Cardioversion• Defibrillation

Page 122: Icd basic 042411(dr chang)

Anti-Tachycardia Pacing

Page 123: Icd basic 042411(dr chang)

Anti-Tachycardia Pacing

Re-entry initiated

ATP delivered at a rate faster than tachyarrhythmia.Wavefronts collide.

Subsequent Pulse: Wavefronts collide closer to re-entry circuit

Subsequent Pulses: Wavefronts collide even closer to re-entry circuit

Arrhythmia terminated

Page 124: Icd basic 042411(dr chang)

Anti-Tachycardia Pacing

•Has programmable:–Sequences – the number of times ATP will be applied upon re-detection

Page 125: Icd basic 042411(dr chang)

Anti-Tachycardia Pacing

•Has programmable:

–Pulses – the number of pulses per sequence

Page 126: Icd basic 042411(dr chang)

Anti-Tachycardia Pacing

•Has programmable:– Sequences – the number of times ATP will be applied upon re-detection (max = 10 in most)

–Pulses – the number of pulses per sequence (1-15)

–Rate of pulses delivered (percent or ms)

Page 127: Icd basic 042411(dr chang)

Anti-Tachycardia Pacing

• Types:–Burst

•A series of pacing pulses delivered at equal intervals

•Interval decrement per sequence

Page 128: Icd basic 042411(dr chang)

Anti-Tachycardia PacingBurst

Page 129: Icd basic 042411(dr chang)

Anti-Tachycardia Pacing

•Types:–Burst

•A series of pacing pulses delivered at equal intervals

•Interval decrement per sequence–Ramp

• A series of pacing pulses delivered at ever decreasing intervals

• Adds a pulse per sequence

Page 130: Icd basic 042411(dr chang)

Anti-Tachycardia PacingRamp

Page 131: Icd basic 042411(dr chang)

Anti-Tachycardia PacingBurst and Ramp Comparison

Programmed Values:

Number of S1 Pulses = 4Number of Sequences =

4R- S1% = 91%Decrement* = 10 ms

* Decrement between sequences

Programmed Values:

Number of S1 Pulses = 4Number of Sequences = 4R-S1% = 91%Decrement* = 10 ms

* Decrement between pulses

* Adds a pulse per sequence

Page 132: Icd basic 042411(dr chang)

ICD Therapies

•Tachyarrhythmia Therapy

–Cardioversion (CV)•Non-committed shock therapy (must synchronize to an R-wave to be delivered)

•Designed to treat re-entrant tachycardias

–Anti-Tachycardia Pacing (ATP)•Pacing pulses delivered at a rate faster than the rhythm detected

•Can successfully terminate re-entrant tachycardias

Low Power

High Power

Page 133: Icd basic 042411(dr chang)

Cardioversion

•Delivers shock on an R-wave•Aborts if synchronization cannot be obtained due to arrhythmia termination

Page 134: Icd basic 042411(dr chang)

CardioversionProgramming

Page 135: Icd basic 042411(dr chang)

ICD Therapies

•Tachyarrhythmia Therapy

–Cardioversion (CV)•A non-committed shock (must synchronize to an R-wave to be delivered)

•Designed to treat organized tachyarrhythmias

–Anti-Tachycardia Pacing (ATP)•Pacing pulses delivered at a rate faster than the rhythm detected

•Can successfully terminate re-entrant tachycardias

Low Power

High Power

–Defibrillation Shock•A shock delivered to the heart to terminate a tachyarrhythmia

High Power

Page 136: Icd basic 042411(dr chang)

ICD Therapy

VT

FVT

VF

Benefits of Tiered Therapy

Page 137: Icd basic 042411(dr chang)

Defibrillation Programming

*Medtronic Programming Screen

Page 138: Icd basic 042411(dr chang)

Question?

When does the device call a therapy a success?

Page 139: Icd basic 042411(dr chang)

ICD Therapies

•“8 to terminate” rule–8 sinus and/or paced beats outside the slowed programmed detection zone

Termination/Redetection

* in Medtronic devices

Page 140: Icd basic 042411(dr chang)

ICD System Troubleshooting

Page 141: Icd basic 042411(dr chang)

Troubleshooting

•ICD System Issues–Lead System & Connections–Device–Other Issues (Patient, EMI)

Overview

Page 142: Icd basic 042411(dr chang)

ICD SYSTEM ISSUES

Lead Systems & Connections

Page 143: Icd basic 042411(dr chang)

Question?

In an ICD System, what component is generally known to be the source of most issues?a. device

b. lead

Page 144: Icd basic 042411(dr chang)

ICD System Issues

• Points of Failure– Acute (common suspects)

• Connector• Lead Dislodgement/Perforation

Leads

Page 145: Icd basic 042411(dr chang)

ICD System Issues

Points of Failure– Chronic (common suspects: High points of stress/pressure)

Lead fracture (commonly exhibits HIGH impedance/resistance)

Lead insulation break (commonly exhibits LOW impedance/resistance)

Leads

Page 146: Icd basic 042411(dr chang)

ICD System Issues

Points of Failure– Acute (common suspects)

Connector

What kinds of things can go wrong here?

Page 147: Icd basic 042411(dr chang)

Can you identify a problem?

Connector

Set screw is obstructing the bore

ICD System Issues

Page 148: Icd basic 042411(dr chang)

Connector

ICD System Issues

What might happen if a lead is connected?

Page 149: Icd basic 042411(dr chang)

ICD System Issues

What is going on here?

Set Screw Noise

Connector

Page 150: Icd basic 042411(dr chang)

ICD System Issues

• Points of Failure– Acute (common suspects)

• ConnectorLead dislodgement/perforation

How do we know when this occurs?

Page 151: Icd basic 042411(dr chang)

Common signs– Intermittent or loss of capture– Intermittent or loss or sensing– Inappropriate therapy during SVT

(in this example due to atrial lead dislodgement)

Lead Dislodgement

Atrial Dislodgement

ICD System Issues

Page 152: Icd basic 042411(dr chang)

Can you identify the dislodged lead(s)?

Lead Dislodgement

ICD System Issues

Page 153: Icd basic 042411(dr chang)

Lead Dislodgement

ICD System Issues

• Avoiding dislodgements– Ensure sufficient slack in lead– Use suture sleeves– Check lead tip stability during implant

Page 154: Icd basic 042411(dr chang)

ICD System Issues

• Occurs when:– Lead tip exits the heart, through the heart

Lead Perforation

Page 155: Icd basic 042411(dr chang)

ICD System Issues

•Diagnosing–Can be seen on x-ray–Definitive by Echo–Threshold and lead impedance may remain unchanged (due to possible continuous contact with tissue outside the heart)

•Possible patient symptoms–Change in pressure–Cardiac tamponade–Dyspnea

Lead Perforation

Page 156: Icd basic 042411(dr chang)

ICD System Issues

• Points of Failure– Acute (common suspects)

• Connector• Lead dislodgement/perforation

– Chronic (common suspects: High points of stress/pressure)

Lead fracture (commonly exhibits HIGH impedance)

Page 157: Icd basic 042411(dr chang)

Can you identify a problem?

1st Rib-Clavicle Crush (lead fracture)

ICD System Issues

Page 158: Icd basic 042411(dr chang)

Lead Crush

1st Rib & Clavicle Crush

Lead Fracture

ICD System Issues

Page 159: Icd basic 042411(dr chang)

Lead Fracture

ICD System Issues

Can you identify the fracture?

Page 160: Icd basic 042411(dr chang)

• Common behavior– Erratic sensing– Intermittent or loss of capture– High lead impedance

Lead Fracture

ICD System Issues

Page 161: Icd basic 042411(dr chang)

Lead Fracture

ICD System Issues

Example

Page 162: Icd basic 042411(dr chang)

ICD System Issues

• Points of Failure–Acute (common suspects)

• Connector• Lead dislodgement/perforation

– Chronic (common suspects: high points of stress/pressure)

Lead fracture (commonly exhibits HIGH impedance)

Lead insulation break (commonly exhibits LOW impedance)

Page 163: Icd basic 042411(dr chang)

Coaxial Inner Insulation Breach

Insulation Break

ICD System Issues

• Common Behavior–Lead impedance

• Low• Can be intermittent

–Capture threshold• Sudden rise or loss• Can be intermittent

–Sensing• Over/undersensing• Can be intermittent

Page 164: Icd basic 042411(dr chang)

Insulation Break

ICD System Issues

Example

Page 165: Icd basic 042411(dr chang)

ICD SYSTEM ISSUES

Device

Page 166: Icd basic 042411(dr chang)

ICD System Issues

• Common device issues:– Long charge times– Battery depletion

• Elective Replacement (ERI)• End of Life (EOL)

Device

– Inappropriate programmingAccelerationOver/UndersensingOutput

Page 167: Icd basic 042411(dr chang)

T-wave oversensing

Device

ICD System Issues

Can you identify a problem?

Page 168: Icd basic 042411(dr chang)

Device

ICD System Issues

Undersensing*Medtronic Programming Screen

Can you identify the problem?

Page 169: Icd basic 042411(dr chang)

Device

ICD System Issues

Undersensing

Can you identify the problem?

Page 170: Icd basic 042411(dr chang)

Infection/Erosion/Allergy

Page 171: Icd basic 042411(dr chang)

Allergy/Infection/Allergy

Page 172: Icd basic 042411(dr chang)

Allergy Test Kit Contents:

Polyurethane

Silicone

Titanium

Platinum Iridium

Polysulfone

EpoxyGold-sputtered devices

Infection/Erosion/AllergyAllergy Testing

Allergy Test Metals(included in kits)

Page 173: Icd basic 042411(dr chang)

Abandoned Leads

• Chatter between active and abandoned leads may cause inappropriate therapy

Page 174: Icd basic 042411(dr chang)

Electromagnetic Interference(EMI)

60 Hz (oversensing)

• The sensing of electrical signals other than those produced by the heart

Page 175: Icd basic 042411(dr chang)

Test Your Skills

What might you suspect from this EGM?

Page 176: Icd basic 042411(dr chang)

Case Study (Noise)

• Male 62 years• ICD Implant for VT/VF• Patient got 11 shocks

Page 177: Icd basic 042411(dr chang)

Fast Path Summary

Page 178: Icd basic 042411(dr chang)

Trends: Signal Amplitude

Page 179: Icd basic 042411(dr chang)

Trends: Lead Impedance

Impedance Decrease

Page 180: Icd basic 042411(dr chang)

Appropriate or Inappropriate Therapy?

Page 181: Icd basic 042411(dr chang)

Episode 52A/V lead 皆有noise

Page 182: Icd basic 042411(dr chang)

Case study (Morphology Change)

• Female 72 years old• ICD for VT episodes• Standard follow-up patient didn’t feel

anything.

Page 183: Icd basic 042411(dr chang)

FastPath Summary

Page 184: Icd basic 042411(dr chang)

Episode Directory

Page 185: Icd basic 042411(dr chang)

Episode 1: Appropriate or Inappropriate Therapy?

Morphology score low

Page 186: Icd basic 042411(dr chang)

Diagnostics

Page 187: Icd basic 042411(dr chang)

Therapy

Page 188: Icd basic 042411(dr chang)

MD Auto Update

Morphology auto update score increase to 100

Page 189: Icd basic 042411(dr chang)

Episode 3: Correct Diagnosis?

SVT Diagnosis

Page 190: Icd basic 042411(dr chang)

Diagnostics

After morphology update Correct Discreamation is SVT

Page 191: Icd basic 042411(dr chang)

CMUH

Page 192: Icd basic 042411(dr chang)

CMUH

Page 193: Icd basic 042411(dr chang)

Indications for ICD Therapy

Page 194: Icd basic 042411(dr chang)

Implantable Cardioverter-Defibrillators

ICD therapy is indicated in patients who are survivors of cardiac arrest due to ventricular fibrillation or hemodynamically unstable sustained VT after evaluation to define the cause of the event and to exclude any completely reversible causes.

ICD therapy is indicated in patients with structural heart disease and spontaneous sustained VT, whether hemodynamically stable or unstable.

ICD therapy is indicated in patients with syncope of undetermined origin with clinically relevant, hemodynamically significant sustained VT or VF induced at electrophysiological study.

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.

Page 195: Icd basic 042411(dr chang)

ICD therapy is indicated in patients with LVEF less than or equal to 35% due to prior MI who are at least 40 days post-MI and are in NYHA functional Class II or III.

ICD therapy is indicated in patients with nonischemic DCM who have an LVEF less than or equal to 35% and who are in NYHA functional Class II or III.

ICD therapy is indicated in patients with LV dysfunction due to prior MI who are at least 40 days post-MI, have an LVEF less than or equal to 30%, and are in NYHA functional Class I.

ICD therapy is indicated in patients with nonsustained VT due to prior MI, LVEF less than or equal to 40%, and inducible VF or sustained VT at electrophysiological study.

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.

Implantable Cardioverter-Defibrillators

Page 196: Icd basic 042411(dr chang)

ICD implantation is reasonable for patients with unexplained syncope, significant LV dysfunction, and nonischemic DCM.

ICD implantation is reasonable for patients with sustained VT and normal or near-normal ventricular function.

ICD implantation is reasonable for patients with HCM who have 1 or more major† risk factors for SCD.

ICD implantation is reasonable for the prevention of SCD in patients with arrhythmogenic right ventricular dysplasia/cardiomyopathy (ARVD/C) who have 1 or more risk factors for SCD.

ICD implantation is reasonable to reduce SCD in patients with long-QT syndrome who are experiencing syncope and/or VT while receiving beta blockers.

I IIaIIbIII

All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.

† See Section 3.2.4, “Hypertrophic Cardiomyopathy,” in the full-text guidelines for definition of major risk factors.

I IIaIIbIII

I IIaIIbIII

I IIaIIbIII

I IIaIIbIII

Implantable Cardioverter-Defibrillators

Page 197: Icd basic 042411(dr chang)

ICD implantation is reasonable for nonhospitalized patients awaiting transplantation.

ICD implantation is reasonable for patients with Brugada syndrome who have had syncope.

ICD implantation is reasonable for patients with Brugada syndrome who have documented VT that has not resulted in cardiac arrest.

ICD implantation is reasonable for patients with catecholaminergic polymorphic VT who have syncope and/or documented sustained VT while receiving beta blockers.

ICD implantation is reasonable for patients with cardiac sarcoidosis, giant cell myocarditis, or Chagas disease.

I IIaIIbIII

I IIaIIbIII

I IIaIIbIII

I IIaIIbIII

I IIaIIbIII

All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.

Implantable Cardioverter-Defibrillators

Page 198: Icd basic 042411(dr chang)

ICD therapy may be considered in patients with nonischemic heart disease who have an LVEF of less than or equal to 35% and who are in NYHA functional Class I.

ICD therapy may be considered for patients with long-QT syndrome and risk factors for SCD.

ICD therapy may be considered in patients with syncope and advanced structural heart disease in whom thorough invasive and noninvasive investigations have failed to define a cause.

ICD therapy may be considered in patients with a familial cardiomyopathy associated with sudden death.

ICD therapy may be considered in patients with LV noncompaction.

III IIaIIaIIa IIbIIbIIbIIIIIIIIIIII IIaIIaIIa IIbIIbIIbIIIIIIIIIIII IIaIIaIIa IIbIIbIIbIIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

I IIaIIbIII

I IIaIIbIII

I IIaIIbIII

I IIaIIbIII

All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.

Implantable Cardioverter-Defibrillators

Page 199: Icd basic 042411(dr chang)

ICD therapy is not indicated for patients who do not have a reasonable expectation of survival with an acceptable functional status for at least 1 year, even if they meet ICD implantation criteria specified in the Class I, IIa, and IIb recommendations above.

ICD therapy is not indicated for patients with incessant VT or VF.

ICD therapy is not indicated in patients with significant psychiatric illnesses that may be aggravated by device implantation or that may preclude systematic follow-up.

ICD therapy is not indicated for NYHA Class IV patients with drug-refractory congestive heart failure who are not candidates for cardiac transplantation or cardiac resynchronization therapy defibrillators (CRT-D).

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.

Implantable Cardioverter-Defibrillators

Page 200: Icd basic 042411(dr chang)

ICD therapy is not indicated for syncope of undetermined cause in a patient without inducible ventricular tachyarrhythmias and without structural heart disease.

ICD therapy is not indicated when VF or VT is amenable to surgical or catheter ablation (e.g., atrial arrhythmias associated with the Wolff-Parkinson-White syndrome, RV or LV outflow tract VT, idiopathic VT, or fascicular VT in the absence of structural heart disease).

ICD therapy is not indicated for patients with ventricular tachyarrhythmias due to a completely reversible disorder in the absence of structural heart disease (e.g., electrolyte imbalance, drugs, or trauma).

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.

Implantable Cardioverter-Defibrillators

Page 201: Icd basic 042411(dr chang)

ICDs in Pediatric Patients and Patients With Congenital Heart Disease

ICD implantation is indicated in the survivor of cardiac arrest after evaluation to define the cause of the event and exclusion of any reversible causes.

ICD implantation is indicated for patients with symptomatic sustained VT in association with congenital heart disease who have undergone hemodynamic and electrophysiological evaluation. Catheter ablation or surgical repair may offer possible alternatives in carefully selected patients.

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.

Page 202: Icd basic 042411(dr chang)

ICDs in Pediatric Patients and Patients With Congenital Heart Disease

ICD implantation is reasonable for patients with congenital heart disease with recurrent syncope of undetermined origin in the presence of either ventricular dysfunction or inducible ventricular arrhythmias at electrophysiological study.

ICD implantation may be considered for patients with recurrent syncope associated with complex congenital heart disease and advanced systemic ventricular dysfunction when thorough invasive and noninvasive investigations have failed to define a cause.

All Class III recommendations found in Section 3 of the full-text guidelines, “Indications for Implantable Cardioverter-Defibrillator Therapy,” apply to pediatric patients and patients with congenital heart disease, and ICD implantation is not indicated in these patient populations.

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

III IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIII IIaIIaIIa IIbIIbIIb IIIIIIIIIIIaIIaIIa IIbIIbIIb IIIIIIIII

All primary SCD prevention ICD recommendations apply only to patients who are receiving optimal medical therapy and have reasonable expectation of survival with good functional capacity for more than 1 year.

Page 203: Icd basic 042411(dr chang)

Major Implantable Cardioverter-Defibrillator Trials for Prevention of Sudden Cardiac Death

Trial Year Patients(n)

LVEF Additional StudyFeatures

HazardRatio*

95% CI p

MADIT I 1996 196 < 35% NSVT and EP+ 0.46 (0.26-0.82) p=0.009

MADIT II 2002 1232 < 30% Prior MI 0.69 (0.51-0.93) p=0.016

CABG-Patch 1997 900 < 36% +SAECG and CABG 1.07 (0.81-1.42) p=0.63

DEFINITE 2004 485 < 35% NICM, PVCs or NSVT 0.65 (0.40-1.06) p=0.08

DINAMIT 2004 674 < 35% 6-40 days post-MI and Impaired HRV

1.08 (0.76-1.55) p=0.66

SCD-HeFT 2006 1676 < 35% Prior MI of NICM 0.77 (0.62-0.96) p=0.007

AVID 1997 1016 Prior cardiac arrest

NA 0.62 (0.43-0.82) NS

CASH† 2000 191 Prior cardiac arrest

NA 0.766 ‡ 1-sided p=0.081

CIDS 2000 659 Prior cardiac arrest, syncope

NA 0.82 (0.60-1.1) NS

* Hazard ratios for death from any cause in the ICD group compared with the non-ICD group. Includes only ICD and amiodarone patients from CASH.‡CI Upper Bound 1.112 CI indicates Confidence Interval, NS = Not statistically significant, NSVT = nonsustained ventricular tachycardia, SAECG = signal-averaged electrocardiogram.

Epstein A, et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities. J Am Coll Cardiol 2008; 51:e1–62. Table 5.

Page 204: Icd basic 042411(dr chang)

Comparison of Medical Therapy, Pacing, and Defibrillation in Heart Failure (COMPANION) Trial

• 1520 patients with NYHA Class III or IV HF, ischemic cardiomyopathy (ICM) or nonischemic cardiomyopathy (NICM) and QRS ≥ 120 ms

• Randomized 1:2:2 to optimal pharmacological therapy (OPT) alone or in combination with cardiac resynchronization therapy with either a pacemaker (CRT-P) or pacemaker-defibrillator (CRT-D)

• Both device arms significantly ↓ combined risk of all-cause hospitalization and all-cause mortality by ~20% compared with OPT

• CRT-D ↓ mortality by 36% compared with OPT (p=0.003)• Insufficient evidence to conclude that CRT-P inferior to

CRT-D

Bristow MR, Saxon LA, Boehmer J, et al. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med 2004;350:2140-50.

Page 205: Icd basic 042411(dr chang)

Implantable Cardioverter-Defibrillators and Prevention of Sudden Cardiac Death in Hypertrophic Cardiomyopathy

• Multicenter registry study of implanted ICDs in 506 unrelated patients with HCM @ high risk for SCD (family hx of SCD, [septal thickness ≥ 30 mm], NSVT, syncope)

• Mean patient age 42 years (SD=17) and 87% had no or only mildly limiting symptoms

• Appropriate ICD discharge rates were 11% per year for 2o prevention and 4% per year for 1o prevention

• For 1o prevention, 35% of patients with appropriate ICD interventions had undergone implantation for only 1 risk factor

Maron BJ, Spirito P, Shen WK, et al. Implantable cardioverter-defibrillators and prevention of sudden cardiac death in hypertrophic cardiomyopathy. JAMA 2007;298:405-12.

Page 206: Icd basic 042411(dr chang)

Multicenter Automatic Defibrillator Implantation Trial II (MADIT II)

• 1232 patients ≥ 1 month post-MI and LVEF ≤ 30%• Randomized to ICD (n=742) or medical therapy (n=490)• No spontaneous or induced arrhythmia required for

enrollment• 6% absolute and 31% relative risk ↓ in all-cause mortality

with ICD therapy (p=0.016)

Moss AJ, Zareba W, Hall WJ, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med 2002;346:877-83.

Page 207: Icd basic 042411(dr chang)

Sudden Death in Heart Failure (SCD-HeFT) Trial

• 2521 patients with NYHA Class II or III HF, ICM, or NICM and LVEF ≤ 35%

• Randomized to 1) conventional rx for HF + placebo; 2) conventional rx + amiodarone; or 3) conventional rx + conservatively programmed shock-only single lead ICD

• No survival benefit for amiodarone• 23% ↓ in overall mortality with ICD therapy• Absolute ↓ in mortality of 7.2% after 5 y in the overall

populationBardy GH, Lee KL, Mark DB, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 2005;352:225-37.

Page 208: Icd basic 042411(dr chang)

Defibrillator in Acute Myocardial Infarction (DINAMIT) Trial

• 674 patients 6 to 40 days post-MI with LVEF ≤ 35% and impaired cardiac autonomic function

• Randomized to ICD therapy (n=332) or no ICD therapy (n=342)

• Arrhythmic death ↓ in ICD group, but ↑ in nonarrhythmic death (6.1% per year vs. 3.5% per year, HR 1.75 (95% CI 1.11 to 2.76; p=0.016)

• No difference in total mortality

Hohnloser SH, Kuck KH, Dorian P, et al. Prophylactic use of an implantable cardioverter-defibrillator after acute myocardial infarction. N Engl J Med 2004;351:2481-8.

Page 209: Icd basic 042411(dr chang)

Defibrillators in Nonischemic Cardiomyopathy Treatment Evaluation (DEFINITE) Trial

• 458 patients with NYHA Class I to III, NICM, LVEF ≤ 35% and premature ventricular contractions (> 10/h) or NSVT

• Randomized to standard medical rx alone or in combination with single-chamber ICD

• Strong trend toward ↓ all-cause mortality with ICD therapy, although not statistically significant (p=0.08)

Kadish A, Dyer A, Daubert JP, et al. Prophylactic defibrillator implantation in patients with nonischemic dilated cardiomyopathy. N Engl J Med 2004;350:2151-8.

Page 210: Icd basic 042411(dr chang)

Notable Changes in 2008 ACC/AHA/HRS Guidelines

1. ICD recommendations are combined into a single list because of overlap between primary and secondary indications.

2. Primary prevention ICD indications in nonischemic cardiomyopathy are clarified using data from SCD-HeFT (i.e., ischemic and nonischemic cardiomyopathies and LVEF ≤35%, NYHA II-III) for support.

3. Indications for ICD therapy in inherited arrhythmia syndromes and selected nonischemic cardiomyopathies are listed.

4. MADIT II indication (i.e., ischemic cardiomypathy and LVEF ≤30%, NYHA I) is now Class I, elevated from Class IIa.

5. EF criteria for primary prevention ICD indications are based on entry criteria for trials on which the recommendations are based.

6. Emphasized primary SCD prevention ICD recommendations apply only to patients receiving optimal medical therapy and reasonable expectation of survival with good functional capacity for >1 year.

7. Independent risk assessment preceding ICD implantation is emphasized, including consideration of patient preference.

8. Optimization of pacemaker programming to minimize unneeded RV pacing is encouraged.9. Pacemaker insertion is discouraged for asymptomatic bradycardia, particularly at night.10. A section has been added that addresses ICD and pacemaker programming at end of life.