Calorimeters for Precision Timing Measurements in High Energy Physics
Calor 2014
Gieβen 10.04.2014
Adi Bornheim, Cristian Pena, Artur Apresyan, Maria Spiropulu, Javier Duarte, Si Xie, Anatoly Ronzhin
Caltech, FNAL
Outline
Introduction Timing measurements of high energy particles
with scintillating crystals. Cosmic Test and Test Beam Setup Results Summary
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 2
Timing in high energy experiment
10.04.2014
High energy physics collider experiments (eg at LHC) : Size of the detector few meters, timing for final state particles of few GeV up to few 100 GeV.
Goal : Associate calorimeter measurements with primary vertex location. From MC simulations : Need single channel resolution of a few 10 ps. See also presentation «Timing Performance of the CMS Electromagnetic
Calorimeter and Prospects for the Future»
∆tvertex (0.2 – 1.0 ns)
∆tflight (4 – 12 ns, calo front face, > 1 ns for low pT charge particles)
∆tdetection (1.0 ns, photons in dense detector)
Adi Bornheim, Precision Timing Calorimeter 3
Electromagnetic Type Detection in Crystals
Photo Detector Crystal
γ
tTOF tC Conversion
Depth (Photons)
tS Scintillation
process
tT Transit time
jitter
tP Photo
detector jitter
x
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 4
tD DAQ
The contribution of the DAQ system – for our tests DRS4 based - to the
overall timing resolution will be discussed by A. Apresyan in the following talk.
Photon Traces in LYSO Crystal
t0 t1 t2
t3 t4 t6
For high energy showers in high light yield crystals, number of scintillation light yield is very large (>105 / GeV).
Photon detection at one location in the crystal will be an averaged transit time spectrum
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 5
Shower Shape and Size Size of the shower given by radiation length X0. We use 1.7 cm, 10 cm and
20 cm LYSO crystals as well as 1.5 mm thick LYSO plates. In dense scintillators X0 is of the order 1 cm. LYSO crystals : 1.2 cm. From simulation studies : Shower fluctuations in 100 GeV photon showers
cause fluctuation of the mean shower time of the order of few 10 ps, dominated by the conversion depth.
Mean shower depth varies by several X0 as a function of energy. ⇒ Shower propagation takes 100s of ps.
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 6
Scintillation Light Time Spectrum Scintillating crystals get often classified in fast and slow by their light
output decay constants. This is often 10s of ns – PWO, LYSO : ~40 ns. Timing information is extracted from the leading edge of the signal – the
rise time of the light output is important. LYSO :
Scintillation light output rise time tR = 75 ps. 35000 photons/MeV, tD = 33 ns. See : S Seifert, J H L Steenbergen, H T van Dam and D R Schaart, 2012
JINST 7 P09004. doi:10.1088/1748-0221/7/09/P09004
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 7
Photo Detector Timing Performance Typical timing performance parameters of photo detectors are the rise
time, single photon timing jitter, n-photon timing jitter. As we measure signals with many photons there may be additional
factors typically not quoted by manufactures – like the 100000-photon timing jitter.
Part of our program is to characterize the timing performance of various photo detectors.
We are considering PMTs, SiPMs, MCPs, HAPDs. Rise times of faster devices may be smaller than transit time spread.
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 8
Hamamatsu MCP-PMT
Optical Transit Time Spread The effect of the scintillation photon arrival at the photo detector we refer
to as Optical Transit Time Spread.
From previous slides follows : For a large crystals (eg. 20 cm) and showers from high energy particles (we consider 1 to 1000 GeV), the optical transit time spread may be the dominating factor in the timing resolution.
Experimental program to explore ultimate timing resolution, in particular the impact of the optical transit time spread.
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 9
γ x
γ x
t1
t2
EM shower propagation
snapshot
Scintillation light propagation
cS < c
Precision Timing Test Setup
We used two setups so far : • Cosmic Test Stand at Lauritsen Laboratory in
Caltech. • Beam Test Setup in Fermilab Test Beam
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 10
Cosmic Timing Test Stand
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 11
Cosmic Timing Test Stand @ Caltech
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 12
Light Propagation in Crystals The effective propagation speed of an large ensemble of photons from
scintillation depends on the geometry, surface structure, transparency ( … ) of the crystal.
For our 20 cm LYSO crystal we measure a mean propagation speed for light from a cosmic muon of about 11 cm/ns (~1/3 x c). Refractive index of LYSO is ~1.8.
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 13
Geometric Timing Spread Location of the muon trespassing the crystal is given by the size of the trigger
counters. Plots below are from a test setup with PMTs. Observed signal signal rise time
of this setup : 2 ns. Reducing size from 3 cm to 1 cm reduced the differential timing distribution
from 300 ps to 160 ps. Clear evidence that the mean time tracks the geometry of the light produced in
the crystal.
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 14
Narrow Trigger Counter Wide Trigger Counter
∆tmax= ∆t1 -∆t2
∆t [ns] ∆t [ns]
Test Setup – Beam Test @ FNAL
LYSO crystal, 20x2.5x2.5 cm3
Photo detectors (here : PMTs)
Light tight box
Mounting base plate
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 15
Test Setup – Beam Test @ FNAL
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 16
Fermilab Test Beam March 2014 All results are preliminary
Test Beam – Perpendicular Setup Wire chamber used to define 16x16 mm area of the beam and
as trigger. Beam size ~2 cm. Reference Photek MCP 40 mm diameter active area, crystal
readout with Hamamatsu MCP, 8 mm diameter active are . Time resolution will be discussed in A. Apresyan presentation.
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 17
Beam
MCP Reference
Lead 2.6 cm
MCP Crystal Readout
MCP Crystal Readout
LYSO Crystal (10 cm)
Wire Chamber
2 cm
Time Of Flight Resolution – Perpendicular Setup
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 18
Measurement with and without lead absorber, 8 GeV electron beam. Beam located 2 cm away from one end of the crystal. Time resolution (difference between one crystal readout MCP and
reference MCP) : typically better than 60 ps. Geometric spread of the impact position (1.6 cm) should contribute
significantly to this. Reference MCP and readout resolution not unfolded.
With lead Without lead
Test Beam – Longitudinal Setup
Wire chamber used to define 16x16 mm area of the beam and as trigger. Beam size ~2 cm.
Reference MCP : Photek, 40 mm active area. Crystal readout MCP : Hamamatsu, 8mm active area.
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 19
Beam
MCP Reference
MCP Crystal Readout
farside
MCP Crystal Readout
nearside
LYSO Crystal (20 cm) Wire Chamber
Longitudinal Setup Electron beam, 8 GeV. Differential time (near side – far side crystal readout) resolution :
53 ps, core width 30 ps. To be compared with the 160 ps from the cosmic muon test.
In the longitudinal setup, transverse geometric spread negligible. Longitudinal shower fluctuation may impact timing measurement.
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 20
σt = 30 ps
Time Of Flight Resolution - Longitudinal Setup
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 21
Time of flight resolution with respect to reference MCP : ~50 ps. Near side MCP readout shows non-gaussian distribution.
Near Side MCP Far Side MCP σt = 47 ps σt = 49 ps
Simulation : Conversion Depth vs Time GEANT simulation of 100 GeV photon : Correlation between conversion
depth and scintillation photon arrival at crystal near and far side face. Stronger correlation at the near side face.
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 22
Z γ
16 cm
Conversion depth pos Z [cm]
phot
on a
rriv
al ti
me
[ps]
Near side crystal face Far side crystal face
Shower Depth Effects in Data Near side correlation is visible in data. On this data set, using correlation to correct shower depth effects
improves resolution significantly.
10.04.2014 23 Adi Bornheim, Precision Timing Calorimeter
∆tN -∆tF ∆tN -∆tF
∆t re
f -∆
t N
∆t re
f -∆
t F
∆tN -∆tF
LYSO Crystal Plates Small LYSO plates (1.4x1.4x0.15 cm) as foreseen for LYSO
Shashlik prototype. Measure time resolution around 50 ps for 8 GeV electrons. Geometry and alignment of the setup more difficult to control.
Further data analysis ongoing.
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 24
Summary
We study the timing measurement performance with LYSO crystals for particles in the GeV energy range.
We achieve timing resolutions of around 50 ps without unfolding reference measurement resolution.
We find evidence that : The timing measurement is driven by the mean time of the
shower evolution. Geometric effects of the shower evolution and light
propagation are dominating factors in the residual time resolution.
Single channel time resolution of a few 10 ps seems achievable.
10.04.2014 Adi Bornheim, Precision Timing Calorimeter 25