HIGHLIGHTS OF MECHANICALVENTILATION UNIT 4
Modes and initiation of ventilationBy
Elizabeth Kelley Buzbee AAS, RRT-NPS
-
8/12/2019 Bipap y Aprv
2/77
The modes of ventilation: A spontaneous breathis one that the patient
triggers and cycles the breath, and he controlsthe VT. This breath could be assisted by theapplication of positive pressure.
A mandatory breathis defined as one that istriggered and cycled by the machine. All
mandatory breaths are assisted breaths.
-
8/12/2019 Bipap y Aprv
3/77
The modes of ventilation: full
support modes CMV: continuous mandatory ventilationin which
all breaths are mandatory.
VC-CMVvolume control also called
Assist/Control mode Set VT, f to get VE; guaranteed VT
Default ventilatory mode for full support with adults
PC-CMVpressure control mode. Patient can
trigger breaths just like with A/C Set PIP, f and TI no guaranteed VT
Default ventilator mode for full support for infants
-
8/12/2019 Bipap y Aprv
4/77
Indications for PC: the RCP selects
pressure ventilation when:
The adult patient who cannot be managed withVC In this case, we keep the PIP less than 30cmH20.
PC results in better distribution of ventilation inpersons with unequal RAW, but consistentcompliances.
There is such an airway leak so that the VTareunstable [most common with infants and smallchildren with uncuffed ET or tracheostomy tubes]
-
8/12/2019 Bipap y Aprv
5/77
Compare PC to VC
In PC, the airway pressures; mPAW and PIPwill stay the same, but the VE and VT can varybased on patients time constants
In VC the VE and VT are basically stable[patient can increase f so VE could vary] thePIP and the mPAW can be altered by patient
time constants
-
8/12/2019 Bipap y Aprv
6/77
Compare control mode to
Assist/control We control patients by giving them sedation
and paralytic agents so that the VE we set onVC-CMV is exactly the same
We can control their PaC02 thus their acidbase balance
In A/C, the patient can trigger breaths thatwill increase the VE, so that the VE based onset VT and f could be lower than the actualmeasured VE
-
8/12/2019 Bipap y Aprv
7/77
Controlling the chronic
hypercapnic patient If your patient has a hypoxic drive,
administrate enough Fi02to get his Pa02between 80-100 mmHg.
This will result in apnea and works as a formof sedation in the first 24 hours.
Must wean the Fi02 to get Pa02 between 55-65 mmHg before weaning
-
8/12/2019 Bipap y Aprv
8/77
Problems with A/C
Excessively high PAWcan cause problems withhemodynamics once patient starts to breath.
Another problem with A/C mode is the risk ofauto-PEEP and air trapping.
-
8/12/2019 Bipap y Aprv
9/77
Inverse Ratio Ventilation [with
PC or with VC] This is a form of full support that uses
increased Ti to raise the mPAW whenpatients compliance is so bad that PIP andPplateau are excessive
In IRV, the expiratory time is so short that thepatient never completely exhales. This works
like PEEP to recruit alveoli
-
8/12/2019 Bipap y Aprv
10/77
Raising mPAW with IRV
mPAW = PIP [I] + PEEP [E]
I + E
Because we raise the inspiratory time somuch we can decrease the PIP
Because we create auto-PEEP with the shortTE, we can decrease the PEEP
-
8/12/2019 Bipap y Aprv
11/77
-
8/12/2019 Bipap y Aprv
12/77
Problems with Negative pressure
Ventilation: patient must be able to:
protect airway
Handle being supine all the time
hemodynamically stable
be comfortable in one position all the time
handle being disconnected from vacuum for
short time spans
-
8/12/2019 Bipap y Aprv
13/77
More problems with NPV
Patient can get skin lesions from movementof body inside the device
Patient can get cold from wind Best 02 device is nasal cannula because 02
can be sucked into the neck opening
-
8/12/2019 Bipap y Aprv
14/77
NPV
Classified as controllers, but newer modelscan be A/C if there is a flow sensor placed onthe patients nose
Old metal iron lungs have a constant I:E of I:I;newer fiberglass devices can have altered I:Eratios
-
8/12/2019 Bipap y Aprv
15/77
Setting parameters on NPV:
Change level of the vacuum to increase the VT[he could use a Wrights spirometer attachedto an IPPB mask to measure exhaled V
T
]
Change the respiratory rate.
-
8/12/2019 Bipap y Aprv
16/77
CSV
continuous spontaneous ventilationin whichall breaths are spontaneous.
patient who can completely control his VE and only needs a little help such as with
increased baseline pressures [CPAP]
or some application of assisted breaths suchas pressure support [PS]
or who might require monitoring of VE
-
8/12/2019 Bipap y Aprv
17/77
Pressure support ventilation
PSV is the most common form of pressurecycled CSV.
Although this does raise the airway pressure sothat we have a higher and lower pressure, wecall this PS rather than PIP because of thespecific characteristics of PS
Flow triggered and flow cycled Patient controls his VT, f and inspiratory time
-
8/12/2019 Bipap y Aprv
18/77
Indications for PS:
When used with SIMV to reduce the WOB byincreasing the spontaneous VT. We generallyselect the PS that will deliver a reasonable VT
[watch the spontaneous RR] Can be used alone during weaning. Once a
patient is on a PS of 5-10 cmH20, he is
considered at a level that only compensatesfor RAW of the tubing, so is consideredconsistent with spontaneous breathing.
-
8/12/2019 Bipap y Aprv
19/77
PSV flow patterns
The flow pattern is descending till it reaches5 LPM [or 25% of the peak flow] in which theflow stops abruptly.
The flow slows down as the device attemptsto keep the PS at the preset pressure.
-
8/12/2019 Bipap y Aprv
20/77
VT on PS
There is no guaranteed VT, nor VE, but we canincrease the VT by increasing the PS pressure
We need to set VE & high f alarms closely towarn us of problems
The patient sends more air to Zone IIIbecause he is using his diaphragm more with
PS
-
8/12/2019 Bipap y Aprv
21/77
To choose the correct level of PSV
there are three methods: get an appropriate VT [10-15 ml/kg] and titrate the PS level toachieve this VT
increase the PS level till the respiratory rate is normalized [25
bpm or less]
increase the PSV until you decrease the work of breathingthrough the ET tube
To select the appropriate level of PSV to overcome the R
AWuse this
formula
PSV= (PIP - Pplateau) x spont insp. Flow rate [l/sec]
Ventilator flow rate [l/sec]
-
8/12/2019 Bipap y Aprv
22/77
PSmax
or straight pressure support or stand alone PS [PS without SIMV.] In this case, the PS is not usedas a weaning modality but for initial ofmechanical ventilation.
We generally select a PS level that will deliver 10-12 ml kg IBW.
The RCP must remember that this mode is anassist only and the patients VTand VE will vary
base on lung dynamics. There is no guaranteedVT.
Patient must have an intact ventilatory drivefor this to work
-
8/12/2019 Bipap y Aprv
23/77
CPAP mode
spontaneous mode
application of PEEP without any positivepressure breathes.
CPAP is merely a raised baseline with a flow rate
with adjustable Fi02 recruits alveoli which will improve diffusion of 02 CPAP can help return a low compliant lung back
to normal once atelectasis has been resolved.The FRC should rise.
should decrease WOB.
proper application of CPAP should decreaseWOB- watch respiratory rates on this
-
8/12/2019 Bipap y Aprv
24/77
CPAP interfaces
CPAP via the ET tube or a trach tube is calledCPAP
CPAP via a nose mask, face mask or full facemask is called nasal-CPAP [n-CPAP]
Obviously we select the interface based onthe patients ability to protect his airway
-
8/12/2019 Bipap y Aprv
25/77
n-CPAP indications
The successful candidate for n-CPAP would
be the patient who is oriented, has good ventilatory drive without excessive
WOB
and who has the ability to protect his airway.
-
8/12/2019 Bipap y Aprv
26/77
n-CPAP contraindications
Persons at risk for vomiting and aspiration
persons with skin necrosis,
claustrophobia.
-
8/12/2019 Bipap y Aprv
27/77
CPAP indications
Management of the person who is in hypoxemia respiratory failure.This patient will have refractory hypoxemia without respiratoryacidosis..
Treatment of Congestive Heart Failure[CHF] in the patient whohas an intact ventilatory drive and can keep his PaC02down. CPAP of8-12 with Fi02 100% is suggested. [Egans pp, 1095]
A weaning modalityThis invasive CPAP may be the last step beforeextubation. Generally a patient can be extubated from a CPAP of 5-7cmH20 [or can be extubated at a stand-alone PSV of 5-7 cmH20.
Non-invasive management of persons with obstructive sleep apnea[OSA
-
8/12/2019 Bipap y Aprv
28/77
APRV
a spontaneous mode
airway pressure release ventilation
Patient is breathing on two different levels ofCPAP
-
8/12/2019 Bipap y Aprv
29/77
Initial settings for APRV for
ARDS: The higher CPAP is set with the Phigh, while the Plowsets the
lower pressure.
The RCP should also set the time interval [Thigh] for Phighand the time interval [Tlow]for Plow
To initial APRV, the RCP looks to the patients PplateauonPPV and uses that figure for the Phigh.
The Thighis started at 4 seconds for adults and can beprogressively increased to 10-15 seconds
Set the Plow at zero and use the release time [Tlow] to keep
the pressure from dropping to zero Set the Tlow at about.5 to .8 [one time constant] so that the
breath ends with the expiratory flow at 50-75% of peak flow
-
8/12/2019 Bipap y Aprv
30/77
What happens if the patient
goes apnic?
During APRV ventilation if the patient wasstop breathing, the time-cycling betweenhigh and low pressures would appear similarto PC-IRV.
So this is a spontaneous mode that happensto have a back up of sorts
-
8/12/2019 Bipap y Aprv
31/77
Contraindications to APRV
persons with COPD or other problemsassociated with air trapping.
persons with excessively high intracranialpressures [high ICP]
-
8/12/2019 Bipap y Aprv
32/77
Bilevel ventilation
An alternative to APRV is bilevel ventilation.The only difference between bilevelventilationand APRVis that the patientspends more time at the [Plow] lower airwaypressure than at the high airway pressure[Phigh].
-
8/12/2019 Bipap y Aprv
33/77
BiPap- NIPPV
Non-invasive positive pressure ventilation
These BiPap breathes tend to be flow or time
triggered, flow cycled off with the operator selecting PIP [called IPAP]
and PEEP [called EPAP] and bleeding insupplementary 02.
The newer Visioncan get a Fi02. http://emedicine.medscape.com/article/1417959-treatment
http://emedicine.medscape.com/article/1417959-treatmenthttp://emedicine.medscape.com/article/1417959-treatmenthttp://emedicine.medscape.com/article/1417959-treatmenthttp://emedicine.medscape.com/article/1417959-treatment -
8/12/2019 Bipap y Aprv
34/77
contraindications/hazards of NIPPV
do not put this device on an apnic patient because itis NOT a ventilatorit is a breath augmenter.
Persons who cannot protect their airways
Hemodynamically unstable patients Facial burns or trauma
Uncooperative patients
Persons at risk for aspiration: vomiting, nose bleeds,
unconscious, poor gag reflex
Copious secretions
Anatomical problems with gas exchange
-
8/12/2019 Bipap y Aprv
35/77
Indications for NIPPV: acute
care of:
congestive heart disease [n-CPAP or BiPap]
COPD patient who doesnt want to be
intubated recently extubated patient who is at risk of
failing.
immune-suppression for whom we may notwant to risk VAP
-
8/12/2019 Bipap y Aprv
36/77
Indications for long-term NIPPV
Long-term management of both obstructivesleep apnea and central sleep apnea
Long-term management of patients withskeletal or neuromuscular disorders
Long-term management of the COPD patientwho has s/s of chronic hypoventilation
[especially at night] and who is optimallytreated with drugs and other care.
-
8/12/2019 Bipap y Aprv
37/77
Initial settings for BiPap:
IPAP at 8 cmH20 and EPAP at 4 cmH20.
. Increase IPAP in increments of 2 cmH20 todeliver more VT.
To hypoxemia, increase the EPAP in incrementsof 2 cm H20.
Oddly enough, if the EPAP is raised withoutraising the IPAP, the VT might decrease becausethe VTis a function of the change in pressure orthe delta P [ P]
-
8/12/2019 Bipap y Aprv
38/77
The BiPap ST/D
EPAP/CPAP: in this mode, all you get is CPAP
IPAP: in this mode, again, all you get is CPAP.
Spontaneous modethis is a form of PSV in which
you select the PS with the IPAPand the PEEP withthe EPAP. All breaths are patient triggered
Spontaneous/timed: is their version of A/C PCwith each breath patient or time triggered. In this
mode you select the bpm Timed mode: their version of control ventilation
in which you now select the rate and theinspiratory time
-
8/12/2019 Bipap y Aprv
39/77
What is so strange about the
BiPAP ST/D circuit? only a single, large-bore tubing going from
the compressor to the patients mask.
constant leak at the Whisper swivel thiswill leak a minimal amount of gas out of thecircuit and between the very high flow rates
and the leak, the patient doesnt rebreathehis C02. Never plug up this hole!
-
8/12/2019 Bipap y Aprv
40/77
Adding extra 02to the BiPap STD
without starting a fire
add 02at the mask,
start machine first before adding 02so gaswill not leak back into machine
never exceed 15 LPM
-
8/12/2019 Bipap y Aprv
41/77
Compare the BiPap STD to the Vision
BiPap machine
The BiPAP ST/D has no 02inlet
The Respironics Visionplugs into 50 psig 02&can get 21% to 100% Fi0
2
The BiPAP ST/D has no internal alarm, youmust buy a separate alarm
The Respironics Visioncan be used forinvasive ventilation with A/C, SIMV +PSV andCPAP as well as NIPPV [CPAP and S/T]
-
8/12/2019 Bipap y Aprv
42/77
Use of critical care ventilators such as
BiPap machines in the ICU.
As a rule, we would operate these machinesin the PSV mode with PEEP to mimic theBiPap.
It is important to understand that the alarmson these machines may have to be adjusted
out of range
-
8/12/2019 Bipap y Aprv
43/77
Dual modes
combinemandatory ventilation withspontaneous ventilation
IMV:intermittent mandatory ventilation: inwhich some breaths are mandatory and
others are spontaneous. In this type of breath, the ventilator will give
a PPV usually based on VC at timed intervals.The patient can breathe off a constant flowrate or from a demand valve at a VT and flowrate determined by his muscle strength,ventilatory drive and lung mechanics.
-
8/12/2019 Bipap y Aprv
44/77
Advantages of IMV/ SIMV
patient comfort
maintains muscle coordination & musclestrength
reduces V/Q mismatch;Zone III is beingutilized,
[4] lower PAW and is an excellent weaning
modality less likely to cause air-trapping
-
8/12/2019 Bipap y Aprv
45/77
Disadvantage of IMV/ SIMV:
If the patients PPV support is removed tooquickly the patient can suffer increased WOB
We need to monitor the spontaneous VE , RRand VT, we may need to increase support by:
increasing the SIMV rate
adding PS
-
8/12/2019 Bipap y Aprv
46/77
Indications for IMV/SIMV:
IMV is a partial mode of ventilation thatusually includes dual modes.
weaning from CMV when the patients
ventilator muscles are weakened an initial ventilatorsetting when the patient is
at risk for air trapping and is breathing on hisown,
or if the patient who is able to breathe partiallyfor himself is at risk for decreased CO.
-
8/12/2019 Bipap y Aprv
47/77
The difference between SIMV and IMV:
SIMV stands for synchronized intermittentmandatory ventilation.
The mandatory breath can come in sooner ifpatient triggers within the synchronizationwindowof fractions of seconds.
-
8/12/2019 Bipap y Aprv
48/77
In apressure regulated volume controlmode,we are attempting to deliver the VT[becausewe are in VC mode] but we want to keep the
airway pressures low.
ventilator will attempt to deliver the VTat 5
cmH20 below a preset pressure setting.
Special modes: PRVC
PRVC
-
8/12/2019 Bipap y Aprv
49/77
Special modes: VAPS
volume assured, pressure support,the ventilatorwill be attempting to deliver a stable VTwith PSbreaths so that the patient has the advantage of
stable VEas well as the advantages of If a PS breath fails to reach the pre-set VT, the
breath will continue at a constant flow until thevolume is reached. If the patient got the pre-set VT
with the PS breath, it stays PS. Unlike normal PS, these breaths arent just flow
triggered, but can be time triggered.
-
8/12/2019 Bipap y Aprv
50/77
Special modes: MMV
Mandatory minute ventilation gives the patient extra breaths or extra PS
pressure to keep a predetermined minimal VE.
This differs from apnea parameters in thatthe patient doesnt have to actually go apneic
for 20 seconds or more for this to activate. Hemerely needs to have hypoventilation.
-
8/12/2019 Bipap y Aprv
51/77
One problem with MMV
when the patient starts the rapid, shallowbreathing associated with respiratorydistress.
If a patient keeps the VEup with rate only, hecan be in a lot of distress
It is suggested to keep the maximal high
respiratory rate 10 BPM above the average
-
8/12/2019 Bipap y Aprv
52/77
Special modes: ASV adaptive support ventilation: the RCP inputs
the patients IBW and a percentage of the VE.
The ventilator will deliver a VEbased on thepatients IBW.
As the patient takes over more of the breathingthe VE is maintained with PS breaths.
The level of PS changes to give the VT calculated
by the machine, The VT will be determined bythe patients IBW and VD ventilation.
-
8/12/2019 Bipap y Aprv
53/77
Special modes: PAV
In proportional assist ventilation modesimilar to ASV in that the ventilator willcollect data about patients elasticity and
resistance and flow or volume demands inorder to arrive at PS levels that varies.
-
8/12/2019 Bipap y Aprv
54/77
High frequency ventilation
controlled ventilation- the patient is sedatedand paralyzed
VT of less or equal to the VDanatomical
respiratory frequencies of 60 BPM-3600 bpm
All HFV counts on the gas stream going downthe ET tube (inside) AT THE SAME TIME andthe gas flow existing (outside stream).
-
8/12/2019 Bipap y Aprv
55/77
How does HFV work
Penduluft action due to various timeconstants of different portions of the lungs,the gas moves from one lobe to another ,
there is some bulk transfer
-
8/12/2019 Bipap y Aprv
56/77
What are the types of HFV
high frequency jet ventilation
high frequency positive pressure
ventilation
high frequency oscillation
combination of HFJ with CMV
-
8/12/2019 Bipap y Aprv
57/77
Special modes: PRVC
Inpressure regulated volume control, aneffort is made to maintain both a safe levelof airway pressure and delivered V
T
.
In PRVC, the RCP selects a PIP that willnot be exceeded.
To keep the VT, at this safe PIP, the
inspiratory time and the flow rate mustvary.
-
8/12/2019 Bipap y Aprv
58/77
Special modes:Auto-mode:
in some ventilators selection of the auto-mode will allow the ventilator to decreasesupport as a patient starts to take over the
WOB.
The ventilator reverts between a CMV modeand a spontaneous mode based on breath by
breath assessment of the patient
-
8/12/2019 Bipap y Aprv
59/77
Special modes: ATC
Automatic tubing compensation, in thismode the ventilator will compensate for theRAW of the ET tube.
-
8/12/2019 Bipap y Aprv
60/77
INITIAL VENTILATOR SETTINGS
-
8/12/2019 Bipap y Aprv
61/77
VT, set f and VE
Full support A/C or SIMV rate 12-16 BPM
Partial support SIMV below 10 BPM
8-10 ml/Kg IBW normal lungs
6-8 ml/kg IBW asthma
5-8 ml/ kg IBW for ARDS & COPD
VE needs to be 80-100 ml/KgIBW
-
8/12/2019 Bipap y Aprv
62/77
TI and Flow rates
Inspiratory flow ratesof 60-80 LPM for most
If air hungry raise above 80
COPD- 60-100 LPM
Inspiratory times .80-1.2 seconds
-
8/12/2019 Bipap y Aprv
63/77
Flow wave pattern
Constant flowswill decrease inspiratory timeand help with I:E ratios, but can raise the PIP.
Descendingflow curve has the advantage ofbetter distribution of gas into the lung, butwill increase the TIand increase the mPAW
Sine wave: while considered more
physiological, a classic sine wave may nothave enough initial flow to satisfy a patient.Like the descending flow pattern it will raise
the TIand change the I:E ratio
-
8/12/2019 Bipap y Aprv
64/77
Rise time or Ramp
: in an effort to fine-tune flow patterns, theconstant flow can be damped by a rise timeadjustments. When set high, this almost
mimics an ascending flow pattern.
-
8/12/2019 Bipap y Aprv
65/77
Inspiratory pause
The temporary use of the inspiratory pauseat about .5 to 1 second is generally reservedfor gathering Pplateau
-
8/12/2019 Bipap y Aprv
66/77
Fi02
100% is a good place to
Weaning rapidly to 40-50% after ABG
Fi02 needs to be weaned about 20% at a step.
-
8/12/2019 Bipap y Aprv
67/77
PEEP
may be started at zero, PEEP at 5 or lesscmH20 is considered physiologicaland shouldnot result in CV problems-but-- remember
any PEEP that causes hemodynamicproblems is excessive.
Increase or decrease by units of 2
-
8/12/2019 Bipap y Aprv
68/77
Humidification by HME
is limited to persons with good fluid balances,normal secretions and VEless than 10 LPMand normal body temperatures.
If the patient has a gross leak so that 30% ofthe delivered VTis lost, the HME will notwork.
-
8/12/2019 Bipap y Aprv
69/77
Humidification by heated
humidifier can be used with everyone but are necessaryfor patients with secretions. Keep thetemperatures close to 330C +/- 2
-
8/12/2019 Bipap y Aprv
70/77
Sighs:
multiple sighs every hour or so. These sighvolumes were about 1.5 x the VT.
important if VT is less than 7 ml/kg
-
8/12/2019 Bipap y Aprv
71/77
s/p lung resection or lung
transplants
need lower VTand faster rates to protect the
torn lung from rupture. Keep the Pplateauat or below 30 cmH20 [old
Egans 1011]
-
8/12/2019 Bipap y Aprv
72/77
lobar pneumonia:
place patient on the good lung sideso gas goesto the bad lung
avoid PEEP in lobar pneumonia if possible
Try to prolong the Ti
Consider double lumen ET tube so we can settwo ventilators on the patient
-
8/12/2019 Bipap y Aprv
73/77
long-term neuromuscular
patients more comfortable at higher than usual VT[decrease the RR] of 10-12 ml/kg. Thesepatients also tend to want higher flow rates.
They can be managed with low Fi02-even .21as long as Sp02 is above 90-92%
low PEEP of 3-5 to prevent atelectasis are ok
-
8/12/2019 Bipap y Aprv
74/77
Persons with Congestive
Heart failure We can start with normal settings, but if thePIP and Pplateauare excessive, we need todecrease the VT
PEEP at 10 cmH20 and wean the
Once the patients compliance gets better, wemust wean the PEEP
If the patient has an intact ventilatory drive, &good VE, he could be maintained on CPAP
-
8/12/2019 Bipap y Aprv
75/77
Initial parameters when High RAW
is an issue?
start with SIMV because this mode is lesslikely to cause air trapping.
minimize air trapping and auto-PEEP
-
8/12/2019 Bipap y Aprv
76/77
COPD SIMV rate between 10-12 BPM : decrease this to 6-8 to
allow time to exhale start at 60 and raise to100 LPM].
A COPD patient can be started at 40-50% Fi02
Use of PEEP with COPD is dangerous, but if the set PEEP
and the auto-PEEP are kept about the same, the gas ismore likely to leave the lung
, keep Sp02 at 90-92% and keep the PaC02and pH closeto baseline so the patient will not sufferpost-hypercapnic
alkalosis
A th ti [AHI 2005 CPR CPG IV 141]
-
8/12/2019 Bipap y Aprv
77/77
Asthmatic [AHI 2005 CPR CPG pp IV 141]
Alert? may do well on BiPap machine
SIMV rate 6-10 BPM
VTof 6-8 ml/ kg IBW
80-100 LPM with a descending flow pattern to
get 1:4 or 1:5 Start Fi02 at 100%.
Use of PEEP with asthmatics is dangerous, but if
the set PEEP and the auto-PEEP are kept thesame, the gas is more likely to leave the lung.
permissive hypercapnia,